Background: To compare and correlate the retinal sensitivity measurements obtained with Nidek Microperimetry-3 (MP-3) and the CenterVue Macular Integrity Assessment (MAIA) microperimeters among healthy subjects.
Methods: In this prospective comparative study, 31 eyes of 23 subjects underwent complete ophthalmological examination including retinal sensitivity assessments using two microperimeters, the MP-3 (Nidek Technologies) and the MAIA (CenterVue). The mean retinal sensitivity (dB) and its corresponding luminance (asb) and contrast (log units) were analysed between the two instruments. The interdevice reproducibility and level of agreement between the sensitivity values of the devices were assessed.
Results: The mean retinal sensitivity (dB) measured by the MP-3 (25.02±1.06 dB, range: 20.90-26.70) was significantly (p<0.0001) lower compared with the MAIA (30.68±0.74 dB, range: 28-31.84). The luminosity levels were significantly (p<0.0001) higher with the MP3 (7.75±1.31 asb, range: 6.44-9.06) compared with the MAIA (0.92±0.14 asb, range: 0.78-1.06). The contrast sensitivity was significantly higher for the MP-3 (0.94±0.33 log units, range: 0.61-1.27) compared with the MAIA (0.23±0.03 log units, range: 0.20-0.26). Despite these absolute differences, the intraclass coefficient was 0.85 (95% CI 0.70 to 0.92) between the two devices after applying a standard correction factor to each data point (MAIA sensitivity=MP-3 sensitivity+5.65) with a mean difference between MAIA and MP-3 of 0.01.
Conclusion: Retinal sensitivity measures higher, but luminance and contrast sensitivity measure lower for MAIA-generated values compared with the MP-3. The relationships, however, appeared fairly consistent, and application of a standard correction factor allowed the data to be inter-related, at least for normal eyes.
Keywords: imaging; retina.
© Article author(s) (or their employer(s) unless otherwise stated in the text of the article) 2018. All rights reserved. No commercial use is permitted unless otherwise expressly granted.
Conflict of interest statement
Competing interests: None declared.
Similar articles
-
Inter-device comparison of retinal sensitivity measurements: the CenterVue MAIA and the Nidek MP-1.Clin Exp Ophthalmol. 2016 Jan-Feb;44(1):15-23. doi: 10.1111/ceo.12629. Clin Exp Ophthalmol. 2016. PMID: 26285157 Clinical Trial.
-
MP1 AND MAIA FUNDUS PERIMETRY IN HEALTHY SUBJECTS AND PATIENTS AFFECTED BY RETINAL DYSTROPHIES.Retina. 2015 Aug;35(8):1662-9. doi: 10.1097/IAE.0000000000000504. Retina. 2015. PMID: 26214316
-
Investigating the discrepancy between MAIA and MP-1 microperimetry results.Ophthalmic Physiol Opt. 2021 Nov;41(6):1231-1240. doi: 10.1111/opo.12877. Epub 2021 Aug 29. Ophthalmic Physiol Opt. 2021. PMID: 34459022
-
Scotopic microperimetry: evolution, applications and future directions.Clin Exp Optom. 2022 Nov;105(8):793-800. doi: 10.1080/08164622.2021.2023477. Epub 2022 Jan 13. Clin Exp Optom. 2022. PMID: 35025727 Review.
-
Current Clinical Application of Microperimetry: A Review.Semin Ophthalmol. 2018;33(5):620-628. doi: 10.1080/08820538.2017.1375125. Epub 2017 Oct 9. Semin Ophthalmol. 2018. PMID: 28991503 Review.
Cited by
-
Measuring of retina function using microperimetry in diabetic retinopathy.Graefes Arch Clin Exp Ophthalmol. 2022 Mar;260(3):1035. doi: 10.1007/s00417-021-05500-w. Epub 2021 Nov 27. Graefes Arch Clin Exp Ophthalmol. 2022. PMID: 34837508 No abstract available.
-
Microperimetry differences in macular sensitivity threshold between first and second tested eyes.Arq Bras Oftalmol. 2021 May-Jun;84(3):203-208. doi: 10.5935/0004-2749.20210034. Arq Bras Oftalmol. 2021. PMID: 33567023 Free PMC article. Clinical Trial.
-
Topographic and quantitative correlation of structure and function using deep learning in subclinical biomarkers of intermediate age-related macular degeneration.Sci Rep. 2024 Nov 15;14(1):28165. doi: 10.1038/s41598-024-72522-9. Sci Rep. 2024. PMID: 39548108 Free PMC article.
-
Comparison Between MAIA and MP-3 In Healthy Subjects and Patients With Diabetes, Diabetic Retinopathy, and Age-Related Macular Degeneration.Invest Ophthalmol Vis Sci. 2025 Mar 3;66(3):59. doi: 10.1167/iovs.66.3.59. Invest Ophthalmol Vis Sci. 2025. PMID: 40152761 Free PMC article.
-
Correlation of optical coherence tomography angiography and microperimetry (MP3) features in wet age-related macular degeneration.Indian J Ophthalmol. 2018 Dec;66(12):1790-1795. doi: 10.4103/ijo.IJO_866_18. Indian J Ophthalmol. 2018. PMID: 30451180 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous