1-Year Results in Patients Undergoing Transcatheter Aortic Valve Replacement With Failed Surgical Bioprostheses
- PMID: 28521921
- DOI: 10.1016/j.jcin.2017.03.018
1-Year Results in Patients Undergoing Transcatheter Aortic Valve Replacement With Failed Surgical Bioprostheses
Abstract
Objectives: This study evaluated the safety and effectiveness of self-expanding transcatheter aortic valve replacement (TAVR) in patients with surgical valve failure (SVF).
Background: Self-expanding TAVR is superior to medical therapy for patients with severe native aortic valve stenosis at increased surgical risk.
Methods: The CoreValve U.S. Expanded Use Study was a prospective, nonrandomized study that enrolled 233 patients with symptomatic SVF who were deemed unsuitable for reoperation. Patients were treated with self-expanding TAVR and evaluated for 30-day and 1-year outcomes after the procedure. An independent core laboratory was used to evaluate serial echocardiograms for valve hemodynamics and aortic regurgitation.
Results: SVF occurred through stenosis (56.4%), regurgitation (22.0%), or a combination (21.6%). A total of 227 patients underwent attempted TAVR and successful TAVR was achieved in 225 (99.1%) patients. Patients were elderly (76.7 ± 10.8 years), had a Society of Thoracic Surgeons Predicted Risk of Mortality score of 9.0 ± 6.7%, and were severely symptomatic (86.8% New York Heart Association functional class III or IV). The all-cause mortality rate was 2.2% at 30 days and 14.6% at 1 year; major stroke rate was 0.4% at 30 days and 1.8% at 1 year. Moderate aortic regurgitation occurred in 3.5% of patients at 30 days and 7.4% of patients at 1 year, with no severe aortic regurgitation. The rate of new permanent pacemaker implantation was 8.1% at 30 days and 11.0% at 1 year. The mean valve gradient was 17.0 ± 8.8 mm Hg at 30 days and 16.6 ± 8.9 mm Hg at 1 year. Factors significantly associated with higher discharge mean aortic gradients were surgical valve size, stenosis as modality of SVF, and presence of surgical valve prosthesis patient mismatch (all p < 0.001).
Conclusions: Self-expanding TAVR in patients with SVF at increased risk for surgery was associated with a low 1-year mortality and major stroke rate, significantly improved aortic valve hemodynamics, and low rates of moderate and no severe residual aortic regurgitation, with improved quality of life.
Keywords: TAV in SAV; TAVR; prospective trial; surgical valve failure.
Copyright © 2017 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.
Comment in
-
A Big Step Forward in the Validation of the Transcatheter Valve-in-Valve Procedure for the Treatment of Failed Surgical Bioprostheses.JACC Cardiovasc Interv. 2017 May 22;10(10):1045-1047. doi: 10.1016/j.jcin.2017.04.002. JACC Cardiovasc Interv. 2017. PMID: 28521922 No abstract available.
Similar articles
-
Self-Expanding Transcatheter Aortic Valve Replacement Versus Surgical Valve Replacement in Patients at High Risk for Surgery: A Study of Echocardiographic Change and Risk Prediction.Circ Cardiovasc Interv. 2016 Jun;9(6):e003426. doi: 10.1161/CIRCINTERVENTIONS.115.003426. Circ Cardiovasc Interv. 2016. PMID: 27313280 Clinical Trial.
-
Durability and Clinical Outcomes of Transcatheter Aortic Valve Replacement for Failed Surgical Bioprostheses.Circ Cardiovasc Interv. 2019 Oct;12(10):e008155. doi: 10.1161/CIRCINTERVENTIONS.119.008155. Epub 2019 Oct 14. Circ Cardiovasc Interv. 2019. PMID: 31607151
-
Long-Term Outcomes After Transcatheter Aortic Valve-in-Valve Replacement.Circ Cardiovasc Interv. 2018 Sep;11(9):e007038. doi: 10.1161/CIRCINTERVENTIONS.118.007038. Circ Cardiovasc Interv. 2018. PMID: 30354588
-
Transcatheter versus surgical aortic valve replacement in patients at low surgical risk: A meta-analysis of randomized trials and propensity score matched observational studies.Catheter Cardiovasc Interv. 2018 Aug 1;92(2):408-416. doi: 10.1002/ccd.27518. Epub 2018 Feb 1. Catheter Cardiovasc Interv. 2018. PMID: 29388308
-
Bailout surgical explantation of a transcatheter valve-in-valve for subacute thrombosis: When there is no time for anticoagulation: Case report and literature review.Cardiovasc Revasc Med. 2018 Jul;19(5 Pt A):536-539. doi: 10.1016/j.carrev.2017.12.005. Epub 2017 Dec 13. Cardiovasc Revasc Med. 2018. PMID: 29352701 Review.
Cited by
-
Balloon-expanding transcatheter aortic valve implantation for degenerated Mitroflow bioprostheses: clinical and echocardiographic long-term outcomes.Interact Cardiovasc Thorac Surg. 2021 Jul 26;33(2):173-180. doi: 10.1093/icvts/ivab065. Interact Cardiovasc Thorac Surg. 2021. PMID: 33782703 Free PMC article.
-
The perfect prosthesis/patient match: pursuit of the Holy Grail.Ann Cardiothorac Surg. 2024 May 31;13(3):224-235. doi: 10.21037/acs-2023-aae-0181. Epub 2024 May 22. Ann Cardiothorac Surg. 2024. PMID: 38841081 Free PMC article. Review.
-
Allegra Transcatheter Heart Valve inside a Degenerated Sutureless Aortic Bioprosthesis.Thorac Cardiovasc Surg Rep. 2021 Jan;10(1):e1-e5. doi: 10.1055/s-0040-1721483. Epub 2021 Jan 19. Thorac Cardiovasc Surg Rep. 2021. PMID: 33489712 Free PMC article.
-
Transcatheter Aortic Valve-in-Valve Procedure in Patients with Bioprosthetic Structural Valve Deterioration.Methodist Debakey Cardiovasc J. 2017 Jul-Sep;13(3):132-141. doi: 10.14797/mdcj-13-3-132. Methodist Debakey Cardiovasc J. 2017. PMID: 29743998 Free PMC article. Review.
-
Transcatheter Aortic Valve Replacement With Self-Expandable Supra-Annular Valves for Degenerated Surgical Bioprostheses: Insights From Transcatheter Valve Therapy Registry.J Am Heart Assoc. 2021 Sep 21;10(18):e021871. doi: 10.1161/JAHA.121.021871. Epub 2021 Sep 13. J Am Heart Assoc. 2021. PMID: 34514840 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical