Minimally invasive cardiac output technologies in the ICU: putting it all together
- PMID: 28538248
- DOI: 10.1097/MCC.0000000000000417
Minimally invasive cardiac output technologies in the ICU: putting it all together
Abstract
Purpose of review: Haemodynamic monitoring is a cornerstone in the diagnosis and evaluation of treatment in critically ill patients in circulatory distress. The interest in using minimally invasive cardiac output monitors is growing. The purpose of this review is to discuss the currently available devices to provide an overview of their validation studies in order to answer the question whether these devices are ready for implementation in clinical practice.
Recent findings: Current evidence shows that minimally invasive cardiac output monitoring devices are not yet interchangeable with (trans)pulmonary thermodilution in measuring cardiac output. However, validation studies are generally single centre, are based on small sample sizes in heterogeneous groups, and differ in the statistical methods used.
Summary: Minimally and noninvasive monitoring devices may not be sufficiently accurate to replace (trans)pulmonary thermodilution in estimating cardiac output. The current paradigm shift to explore trending ability rather than investigating agreement of absolute values alone is to be applauded. Future research should focus on the effectiveness of these devices in the context of (functional) haemodynamic monitoring before adoption into clinical practice can be recommended.
Similar articles
-
Cardiac output monitoring: a contemporary assessment and review.Crit Care Med. 2015 Jan;43(1):177-85. doi: 10.1097/CCM.0000000000000608. Crit Care Med. 2015. PMID: 25251758 Review.
-
Cardiac output monitoring - invasive and noninvasive.Curr Opin Crit Care. 2022 Jun 1;28(3):340-347. doi: 10.1097/MCC.0000000000000937. Epub 2022 Mar 11. Curr Opin Crit Care. 2022. PMID: 35275876 Review.
-
[Hemodynamic monitoring in the intensive care unit: pulmonary artery catheter versus PiCCO].Dtsch Med Wochenschr. 2011 Feb;136(8):376-80. doi: 10.1055/s-0031-1272539. Epub 2011 Feb 17. Dtsch Med Wochenschr. 2011. PMID: 21332037 Review. German.
-
The PiCCO monitor: a review.Anaesth Intensive Care. 2012 May;40(3):393-409. doi: 10.1177/0310057X1204000304. Anaesth Intensive Care. 2012. PMID: 22577904 Review.
-
Minimally invasive cardiac output monitoring.Curr Opin Crit Care. 2008 Jun;14(3):322-6. doi: 10.1097/MCC.0b013e3282fd6e4a. Curr Opin Crit Care. 2008. PMID: 18467894 Review.
Cited by
-
Accuracy and reliability of noninvasive stroke volume monitoring via ECG-gated 3D electrical impedance tomography in healthy volunteers.PLoS One. 2018 Jan 26;13(1):e0191870. doi: 10.1371/journal.pone.0191870. eCollection 2018. PLoS One. 2018. PMID: 29373611 Free PMC article.
-
Perioperative fluid management in kidney transplantation: a black box.Crit Care. 2018 Jan 25;22(1):14. doi: 10.1186/s13054-017-1928-2. Crit Care. 2018. PMID: 29368625 Free PMC article. Review.
-
Goal-directed therapy: hit early and personalize!J Clin Monit Comput. 2018 Jun;32(3):375-377. doi: 10.1007/s10877-017-0043-x. Epub 2017 Jun 26. J Clin Monit Comput. 2018. PMID: 28653134 No abstract available.
-
Disagreement in cardiac output measurements between fourth-generation FloTrac and critical care ultrasonography in patients with circulatory shock: a prospective observational study.J Intensive Care. 2019 Apr 11;7:21. doi: 10.1186/s40560-019-0373-5. eCollection 2019. J Intensive Care. 2019. PMID: 31011425 Free PMC article.
-
Leveraging Hypotension Prediction Index to Forecast LPS-Induced Acute Lung Injury and Inflammation in a Porcine Model: Exploring the Role of Hypoxia-Inducible Factor in Circulatory Shock.Biomedicines. 2024 Jul 25;12(8):1665. doi: 10.3390/biomedicines12081665. Biomedicines. 2024. PMID: 39200130 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials