AAPM Medical Physics Practice Guideline 8.a.: Linear accelerator performance tests
- PMID: 28548315
- PMCID: PMC5874895
- DOI: 10.1002/acm2.12080
AAPM Medical Physics Practice Guideline 8.a.: Linear accelerator performance tests
Abstract
Purpose: The purpose of this guideline is to provide a list of critical performance tests in order to assist the Qualified Medical Physicist (QMP) in establishing and maintaining a safe and effective quality assurance (QA) program. The performance tests on a linear accelerator (linac) should be selected to fit the clinical patterns of use of the accelerator and care should be given to perform tests which are relevant to detecting errors related to the specific use of the accelerator.
Methods: A risk assessment was performed on tests from current task group reports on linac QA to highlight those tests that are most effective at maintaining safety and quality for the patient. Recommendations are made on the acquisition of reference or baseline data, the establishment of machine isocenter on a routine basis, basing performance tests on clinical use of the linac, working with vendors to establish QA tests and performing tests after maintenance.
Results: The recommended tests proposed in this guideline were chosen based on the results from the risk analysis and the consensus of the guideline's committee. The tests are grouped together by class of test (e.g., dosimetry, mechanical, etc.) and clinical parameter tested. Implementation notes are included for each test so that the QMP can understand the overall goal of each test.
Conclusion: This guideline will assist the QMP in developing a comprehensive QA program for linacs in the external beam radiation therapy setting. The committee sought to prioritize tests by their implication on quality and patient safety. The QMP is ultimately responsible for implementing appropriate tests. In the spirit of the report from American Association of Physicists in Medicine Task Group 100, individual institutions are encouraged to analyze the risks involved in their own clinical practice and determine which performance tests are relevant in their own radiotherapy clinics.
Keywords: quality assurance.
© 2017 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.
Similar articles
-
AAPM Medical Physics Practice Guideline 8.b: Linear accelerator performance tests.J Appl Clin Med Phys. 2023 Nov;24(11):e14160. doi: 10.1002/acm2.14160. Epub 2023 Oct 4. J Appl Clin Med Phys. 2023. PMID: 37793084 Free PMC article.
-
Automating linear accelerator quality assurance.Med Phys. 2015 Oct;42(10):6074-83. doi: 10.1118/1.4931415. Med Phys. 2015. PMID: 26429282
-
AAPM task group 224: Comprehensive proton therapy machine quality assurance.Med Phys. 2019 Aug;46(8):e678-e705. doi: 10.1002/mp.13622. Epub 2019 Jun 14. Med Phys. 2019. PMID: 31125441
-
Quality assurance needs for modern image-based radiotherapy: recommendations from 2007 interorganizational symposium on "quality assurance of radiation therapy: challenges of advanced technology".Int J Radiat Oncol Biol Phys. 2008;71(1 Suppl):S2-12. doi: 10.1016/j.ijrobp.2007.08.080. Int J Radiat Oncol Biol Phys. 2008. PMID: 18406928
-
American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53: quality assurance for clinical radiotherapy treatment planning.Med Phys. 1998 Oct;25(10):1773-829. doi: 10.1118/1.598373. Med Phys. 1998. PMID: 9800687 Review.
Cited by
-
Improving Management and Compliance of Radiation Therapy Linear Accelerator Quality Assurance Program With Automated Tracking Tools.Adv Radiat Oncol. 2024 Feb 9;9(5):101469. doi: 10.1016/j.adro.2024.101469. eCollection 2024 May. Adv Radiat Oncol. 2024. PMID: 38550367 Free PMC article.
-
Implementation of an ionization chamber array for linear accelerator monthly dosimetry QA.J Appl Clin Med Phys. 2024 Sep;25(9):e14433. doi: 10.1002/acm2.14433. Epub 2024 Jun 23. J Appl Clin Med Phys. 2024. PMID: 38923344 Free PMC article.
-
Clinical commissioning of an adaptive radiotherapy platform: Results and recommendations.J Appl Clin Med Phys. 2022 Dec;23(12):e13801. doi: 10.1002/acm2.13801. Epub 2022 Oct 31. J Appl Clin Med Phys. 2022. PMID: 36316805 Free PMC article.
-
Scintillation imaging as a high-resolution, remote, versatile 2D detection system for MR-linac quality assurance.Med Phys. 2020 Sep;47(9):3861-3869. doi: 10.1002/mp.14353. Epub 2020 Jul 18. Med Phys. 2020. PMID: 32583484 Free PMC article.
-
Characterization and longitudinal assessment of daily quality assurance for an MR-guided radiotherapy (MRgRT) linac.J Appl Clin Med Phys. 2019 Nov;20(11):27-36. doi: 10.1002/acm2.12735. Epub 2019 Oct 21. J Appl Clin Med Phys. 2019. PMID: 31633882 Free PMC article.
References
-
- Svensson GK, Baily NA, Loevinger R, et al. Physical Aspects of Quality Assurance in Radiation Therapy, AAPM Task Group 24. American Institute of Physics; 1984.
-
- Kutcher GJ, Coia L, Gillin M, et al. Comprehensive QA for radiation oncology: report of AAPM Radiation Therapy Committee Task Group 40. Med Phys. 1994;21:581–618. - PubMed
-
- Klein EE, Hanley J, Bayouth J, et al. Task Group 142 report: quality assurance of medical accelerators. Med Phys. 2009;36:4197–4212. - PubMed
-
- Bissonnette J, Balter PA, Dong L, et al. Quality assurance for image‐guided radiation therapy utilizing CT‐based technologies: a report of the AAPM TG‐179. Med Phys. 2012;39:1946–1963. - PubMed