. 2017 May 16:11:907-913.
doi: 10.2147/OPTH.S133840. eCollection 2017.

The use of WaveLight® Contoura to create a uniform cornea: the LYRA Protocol. Part 2: the consequences of treating astigmatism on an incorrect axis via excimer laser

Affiliations

The use of WaveLight® Contoura to create a uniform cornea: the LYRA Protocol. Part 2: the consequences of treating astigmatism on an incorrect axis via excimer laser

Manoj Motwani. Clin Ophthalmol. .

Abstract

Purpose: To show how an incorrect manifest astigmatism axis can cause an abnormal induced astigmatism on a new axis.

Patients and methods: Four eyes of three patients were treated primarily with WaveLight® EX500 wavefront optimized (WFO) treatments for astigmatism. All four eyes needed enhancements and were analyzed retrospectively via WaveLight® Contoura to determine the reason for the incorrect astigmatism treatment. Two of the eyes were retreated with topographic-guided ablation, and two were treated with WFO correction.

Results: The eyes that had an incorrect manifest axis resulted in a new abnormal induced astigmatism on a wholly new axis. Treatment with topographic-guided ablation completely eliminated the newly induced astigmatism. Treatment with WFO of an abnormal induced astigmatism corrected the refraction but still left topographic evidence of the abnormal astigmatism behind. One eye was incorrectly treated for astigmatism due to coma affecting refraction when the patient was dilated. This eye had a normal induced astigmatism on a perpendicular axis and was corrected using WFO. The use of manifest refraction with WaveLight® Contoura topographic-guided ablation will lead to incorrect astigmatism correction when the manifest astigmatism and axis differ from the WaveLight® Contoura measured.

Conclusion: Correction of an incorrect manifest refraction astigmatic axis does not simply create undercorrection of the astigmatism but induces an entirely new abnormal astigmatism on a different axis. Manifest refraction is less accurate and can lead to abnormal astigmatism when laser ablation is performed.

Keywords: Contoura with LYRA Protocol; LASIK; PRK; WaveLight Contoura; astigmatism treatment; asymmetric astigmatism; higher order aberrations; topographic guide ablation.

PubMed Disclaimer

Conflict of interest statement

Disclosure Dr Motwani has received a grant from Alcon for a previous unrelated study. The Contoura system using the LYRA Protocol was developed by Dr Motwani in conjunction with Sissimos Lemonis of WaveLight® Lasers. The author reports no other conflicts of interest in this work.

Figures

Figure 1
Figure 1
Patient 1: preoperative OD. Abbreviation: OD, oculus dexter.
Figure 2
Figure 2
Patient 1: postoperative WFO OD. Abbreviations: OD, oculus dexter; WFO, wavefront optimized.
Figure 3
Figure 3
Patient 1: postoperative T-CAT (WaveLight, Erlangen, Germany) OD. Abbreviation: OD, oculus dexter.
Figure 4
Figure 4
Patient 2: preoperative OS. Abbreviation: OS, oculus sinister.
Figure 5
Figure 5
Patient 2: postoperative WFO OS. Abbreviations: OS, oculus sinister; WFO, wavefront optimized.
Figure 6
Figure 6
Patient 2: postoperative WFO OS. Abbreviations: OS, oculus sinister; WFO, wavefront optimized.
Figure 7
Figure 7
Patient 3: preoperative OD. Abbreviation: OD, oculus dexter.
Figure 8
Figure 8
Patient 3: postoperative WFO OD. Abbreviations: OD, oculus dexter; WFO, wavefront optimized.
Figure 9
Figure 9
Patient 4: postoperative WFO OS. Abbreviations: OS, oculus sinister; WFO, wavefront optimized.
Figure 10
Figure 10
Patient 3: post T-CAT (WaveLight, Erlangen, Germany) OD. Abbreviation: OD, oculus dexter.
Figure 11
Figure 11
Patient 4: preoperative OS. Abbreviation: OS, oculus sinister.
Figure 12
Figure 12
Patient 4: postoperative WFO OS. Abbreviations: OS, oculus sinister; WFO, wavefront optimized.

References

    1. Bueeler M, Mrochen M, Seiler T. Maximum permissible torsional misalignment in aberration-sensing and wavefront-guided corneal ablation. J Cataract Refract Surg. 2004;30(1):17–25. - PubMed
    1. Novis C. Astigmatism and toric intraocular lenses. Curr Opin Ophthalmol. 2000;11(1):47–50. - PubMed
    1. Motwani M. The use of WaveLight® Contoura-measured astigmatism to create a uniform cornea: the LYRA Protocol part 3: results from 50 eyes. Clin Ophthalmol. 2017 In press. - PMC - PubMed
    1. Motwani M. The use of WaveLight® Contoura to create a uniform cornea: the LYRAProtocol. Part 1: the effect of higher-order corneal aberrations on refractive astigmatism. Clin Ophthalmol. 2017 In press. - PMC - PubMed