. 2017 Nov;11(6):1187-1195.
doi: 10.1177/1932296817710474. Epub 2017 Jun 1.

Clinical Impact of Blood Glucose Monitoring Accuracy: An In-Silico Study

Affiliations

Clinical Impact of Blood Glucose Monitoring Accuracy: An In-Silico Study

Enrique Campos-Náñez et al. J Diabetes Sci Technol. 2017 Nov.

Abstract

Background: Patients with diabetes rely on blood glucose (BG) monitoring devices to manage their condition. As some self-monitoring devices are becoming more and more accurate, it becomes critical to understand the relationship between system accuracy and clinical outcomes, and the potential benefits of analytical accuracy.

Methods: We conducted a 30-day in-silico study in type 1 diabetes mellitus (T1DM) patients using continuous subcutaneous insulin infusion (CSII) therapy and a variety of BG meters, using the FDA-approved University of Virginia (UVA)/Padova Type 1 Simulator. We used simulated meter models derived from the published characteristics of 43 commercial meters. By controlling random events in each parallel run, we isolated the differences in clinical performance that are directly associated with the meter characteristics.

Results: A meter's systematic bias has a significant and inverse effect on HbA1c ( P < .01), while also affecting the number of severe hypoglycemia events. On the other hand, error, defined as the fraction of measurements beyond 5% of the true value, is a predictor of severe hypoglycemia events ( P < .01), but in the absence of bias has a nonsignificant effect on average glycemia (HbA1c). Both bias and error have significant effects on total daily insulin (TDI) and the number of necessary glucose measurements per day ( P < .01). Furthermore, these relationships can be accurately modeled using linear regression on meter bias and error.

Conclusions: Two components of meter accuracy, bias and error, clearly affect clinical outcomes. While error has little effect on HbA1c, it tends to increase episodes of severe hypoglycemia. Meter bias has significant effects on all considered metrics: a positive systemic bias will reduce HbA1c, but increase the number of severe hypoglycemia attacks, TDI use, and number of fingersticks per day.

Keywords: accuracy; blood glucose meters; clinical outcomes.

PubMed Disclaimer

Conflict of interest statement

Declaration of Conflicting Interests: The author(s) declared the following potential conflicts of interest with respect to the research, authorship, and/or publication of this article: KF is a full-time employee of Roche Diabetes Care.

Figures

Figure 1.
Figure 1.
Project approach to estimating clinical impacts of meter accuracy.
Figure 2.
Figure 2.
Example simulated glucose readings using our modeling approach.
Figure 3.
Figure 3.
Description of the elements of the basic behavioral model.
Figure 4.
Figure 4.
Relationship between consecutive meal times and sizes, using values of δt=15(min),andδa=20(g).
Figure 5.
Figure 5.
Behavioral model of bolus and fingerstick behavior.
Figure 6.
Figure 6.
Relationship between error, bias, HbA1c, and severe hypoglycemia. Each meter is represented by a colored dot. The x-coordinate represents error, while the y-coordinate represents the resulting HbA1c. The size of the dot is proportional to the number of severe hypoglycemia events in 6 months, while the dot’s color shows the meter’s systematic bias.

Similar articles

Cited by

References

    1. Freckmann G, Schmid C, Baumstark A, Pleus S, Link M, Haug C. System accuracy evaluation of 43 blood glucose monitoring systems for self-monitoring of blood glucose according to DIN EN ISO 15197. J Diabetes Sci Technol. 2012;6(5):1060-1075. - PMC - PubMed
    1. Cobelli C, Dalla Man C, Pedersen MG, Bertoldo A, Toffolo G. Advancing our understanding of the glucose systems via modeling. IEEE Trans Biomed Eng. 2014;61(5):1577-1592. - PubMed
    1. Kovatchev BP, Breton M, Dalla Man CD, Cobelli C. In-silico preclinical trials: a proof of concept in closed-loop control of type 1 diabetes. J Diabetes Sci Technol. 2009;3(1):44-55. - PMC - PubMed
    1. Wilinska ME, Chassin LJ, Acerini CL, Allen JM, Dunger DB, Hovorka R. Simulation environment to evaluate closed-loop insulin delivery systems in type 1 diabetes. J Diabetes Sci Technol. 2010;4(1):132-144. - PMC - PubMed
    1. Lee I, Sokolsky O, Chen S, et al. Challenges and research directions in medical cyber-physical systems. Proc IEEE. 2012;100(1):75-90.

MeSH terms