Clinical Impact of Blood Glucose Monitoring Accuracy: An In-Silico Study
- PMID: 28569076
- PMCID: PMC5951046
- DOI: 10.1177/1932296817710474
Clinical Impact of Blood Glucose Monitoring Accuracy: An In-Silico Study
Abstract
Background: Patients with diabetes rely on blood glucose (BG) monitoring devices to manage their condition. As some self-monitoring devices are becoming more and more accurate, it becomes critical to understand the relationship between system accuracy and clinical outcomes, and the potential benefits of analytical accuracy.
Methods: We conducted a 30-day in-silico study in type 1 diabetes mellitus (T1DM) patients using continuous subcutaneous insulin infusion (CSII) therapy and a variety of BG meters, using the FDA-approved University of Virginia (UVA)/Padova Type 1 Simulator. We used simulated meter models derived from the published characteristics of 43 commercial meters. By controlling random events in each parallel run, we isolated the differences in clinical performance that are directly associated with the meter characteristics.
Results: A meter's systematic bias has a significant and inverse effect on HbA1c ( P < .01), while also affecting the number of severe hypoglycemia events. On the other hand, error, defined as the fraction of measurements beyond 5% of the true value, is a predictor of severe hypoglycemia events ( P < .01), but in the absence of bias has a nonsignificant effect on average glycemia (HbA1c). Both bias and error have significant effects on total daily insulin (TDI) and the number of necessary glucose measurements per day ( P < .01). Furthermore, these relationships can be accurately modeled using linear regression on meter bias and error.
Conclusions: Two components of meter accuracy, bias and error, clearly affect clinical outcomes. While error has little effect on HbA1c, it tends to increase episodes of severe hypoglycemia. Meter bias has significant effects on all considered metrics: a positive systemic bias will reduce HbA1c, but increase the number of severe hypoglycemia attacks, TDI use, and number of fingersticks per day.
Keywords: accuracy; blood glucose meters; clinical outcomes.
Conflict of interest statement
Figures






Similar articles
-
Effect of BGM Accuracy on the Clinical Performance of CGM: An In-Silico Study.J Diabetes Sci Technol. 2017 Nov;11(6):1196-1206. doi: 10.1177/1932296817710476. Epub 2017 May 31. J Diabetes Sci Technol. 2017. PMID: 28560900 Free PMC article.
-
Cost-effectiveness of sensor-augmented pump therapy in two different patient populations with type 1 diabetes in Italy.Nutr Metab Cardiovasc Dis. 2018 Jul;28(7):707-715. doi: 10.1016/j.numecd.2018.03.011. Epub 2018 Apr 5. Nutr Metab Cardiovasc Dis. 2018. PMID: 29753586
-
Overnight closed-loop insulin delivery with model predictive control: assessment of hypoglycemia and hyperglycemia risk using simulation studies.J Diabetes Sci Technol. 2009 Sep 1;3(5):1109-20. doi: 10.1177/193229680900300514. J Diabetes Sci Technol. 2009. PMID: 20144424 Free PMC article.
-
Clinical requirements for closed-loop control systems.J Diabetes Sci Technol. 2012 Mar 1;6(2):444-52. doi: 10.1177/193229681200600233. J Diabetes Sci Technol. 2012. PMID: 22538159 Free PMC article. Review.
-
Clinical use of U-500 regular insulin: review and meta-analysis.J Diabetes Sci Technol. 2012 Mar 1;6(2):412-20. doi: 10.1177/193229681200600229. J Diabetes Sci Technol. 2012. PMID: 22538155 Free PMC article. Review.
Cited by
-
Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus.Diabetes Care. 2023 Oct 1;46(10):e151-e199. doi: 10.2337/dci23-0036. Diabetes Care. 2023. PMID: 37471273 Free PMC article. Review.
-
Analysis of "Use of Blood Glucose Meters Featuring Color Range Indicators Improves Glycemic Control and Patients With Diabetes in Comparison to Blood Glucose Meters Without Color (ACCENTS Study)".J Diabetes Sci Technol. 2018 Nov;12(6):1220-1222. doi: 10.1177/1932296818793115. Epub 2018 Aug 10. J Diabetes Sci Technol. 2018. PMID: 30095006 Free PMC article.
-
Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus.Clin Chem. 2023 Aug 2;69(8):808-868. doi: 10.1093/clinchem/hvad080. Clin Chem. 2023. PMID: 37473453 Free PMC article.
-
Continuous Glucose Monitoring (CGM) or Blood Glucose Monitoring (BGM): Interactions and Implications.J Diabetes Sci Technol. 2018 Jul;12(4):873-879. doi: 10.1177/1932296818768834. Epub 2018 Apr 12. J Diabetes Sci Technol. 2018. PMID: 29648465 Free PMC article.
-
User Performance Evaluation and System Accuracy Assessment of Four Blood Glucose Monitoring Systems With Color Coding of Measurement Results.J Diabetes Sci Technol. 2024 May;18(3):644-652. doi: 10.1177/19322968221141926. Epub 2022 Nov 26. J Diabetes Sci Technol. 2024. PMID: 36433806 Free PMC article.
References
-
- Cobelli C, Dalla Man C, Pedersen MG, Bertoldo A, Toffolo G. Advancing our understanding of the glucose systems via modeling. IEEE Trans Biomed Eng. 2014;61(5):1577-1592. - PubMed
-
- Lee I, Sokolsky O, Chen S, et al. Challenges and research directions in medical cyber-physical systems. Proc IEEE. 2012;100(1):75-90.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Research Materials