. 2017 Apr;11(4):OC05-OC08.
doi: 10.7860/JCDR/2017/23926.9613. Epub 2017 Apr 1.

Accuracy and Precision of Measured Blood Sugar Values by Three Glucometers Compared to the Standard Technique

Affiliations

Accuracy and Precision of Measured Blood Sugar Values by Three Glucometers Compared to the Standard Technique

Shadi Khakpour Kermani et al. J Clin Diagn Res. 2017 Apr.

Abstract

Introduction: The prevalence of diabetes mellitus has magnified, globally. Being cost effective and user friendly, using the glucometer systems has become more widespread. However, despite their advantages, there are controversies regarding their accuracy and precision.

Aim: To determine the accuracy and precision of the common glucometers of Bionime, Accu-check, and Beurer compared with the laboratory technique.

Materials and methods: In a cross-sectional study, 257 patients of a general internal unit of Imam Reza hospital of Kermanshah were recruited. The blood sample of the subjects was transferred to the laboratory, then, two samples of capillary blood were tested by the three glucometers. The data was entered into a checklist, then entered to SPSS-16.0 software, and analyzed by descriptive and inferential statistics.

Results: In this study, there was a positive correlation between the three blood glucose measurement methods with the standard approach, significantly, which indicates acceptable accuracy. However, this correlation was higher in the Beurer Glucometer (r=0.976, p<0.001). There were no significant differences in repeated measurements of blood glucose in the glucometers of Beurer (t=0.96, p=0.92), and Bionime (t=0.50, p=0.61) by paired t-test, but this test was significant for Accu-chek system (t=2.7, p=0.006).

Conclusion: With respect to the acceptable accuracy and precision of Beurer and Bionime glucometers, it is suggested to use these methods for diabetic patients at homes as well as in the hospital units.

Keywords: Blood glucose self-monitoring; Diabetes complications; Diabetes mellitus.

PubMed Disclaimer

Figures

[Table/Fig-3]:
[Table/Fig-3]:
The area under curve by using the ROC curve.

References

    1. Zhang P, Zhang X, Brown J, Vistisen D, Sicree R, Shaw J, et al. Global healthcare expenditure on diabetes for 2010 and 2030. Diabetes Research and Clinical Practice. 2010;87(3):293–301. - PubMed
    1. Khatoni A, Abdi A, Hoseini E, Fattahpour T, Feizi H. Assessment of risk factors and complications of type2 diabetes in patients of Ghasreshirin diabetes center in 2011. Nursing Journal of the Vulnerable. 2015;2(2):1–14.
    1. Esteghamati A, Gouya MM, Abbasi M, Delavari A, Alikhani S, Alaedini F, et al. Prevalence of diabetes and impaired fasting glucose in the adult population of Iran national survey of risk factors for non-communicable diseases of Iran. Diabetes Care. 2008;31(1):96–98. - PubMed
    1. American Diabetes Association. Economic costs of diabetes in the US in 2007. Diabetes Care. 2008;31(3):596–615. - PubMed
    1. Shrestha N, Lohani SP, Angdembe MR, Bhattarai K, Bhattarai J. Cost of diabetes mellitus care among patients attending selected outpatient clinics. Journal of Nepal Medical Association. 2013;52(190):343–38. - PubMed

LinkOut - more resources