Randomized Controlled Trial
. 2017 May:90:97-105.
doi: 10.1016/j.ejrad.2017.02.020. Epub 2017 Feb 17.

Prospectively ECG-triggered high-pitch coronary CT angiography at 70 kVp with 30mL contrast agent: An intraindividual comparison with sequential scanning at 120 kVp with 60mL contrast agent

Affiliations
Randomized Controlled Trial

Prospectively ECG-triggered high-pitch coronary CT angiography at 70 kVp with 30mL contrast agent: An intraindividual comparison with sequential scanning at 120 kVp with 60mL contrast agent

Wei Wang et al. Eur J Radiol. 2017 May.

Abstract

Purpose: To investigate image quality, radiation dose, and diagnostic efficiency of prospectively ECG-triggered high-pitch coronary CT angiography (CCTA) at 70 kVp with 30mL contrast agent intra-individually compared with routine CCTA protocol.

Materials and methods: One hundred and thirty eight patients with suspected coronary artery disease, body mass index (BMI)≤25kg/m2 and heart rate (HR)≤70 beats per minute (bpm) underwent prospectively ECG-triggered high-pitch CCTA at 70 kVp and 30mL contrast agent (protocol A) and prospectively ECG-triggered sequential scanning at 120 kVp and 60mL contrast medium (protocol B). Objective and subjective image quality, radiation doses, and diagnostic accuracy were evaluated and compared between the two protocols.

Results: Higher CT attenuation, higher noise, lower signal-to-noise ratios (SNRs) and lower contrast-to-noise ratios (CNRs) were found in protocol A than in protocol B (P<0.001). However, image quality of protocol A were diagnostic. In patients with BMI<23kg/m2 or HR<60bpm, subjective image quality scores of some coronary arteries in protocol A were not significantly different from protocol B (P>0.05). Effective dose in protocol A has reduced by 96.7% compared with protocol B (P<0.001). No significant differences were found for diagnostic accuracy between the two protocols on a per-segment (P=0.513), per-vessel (P=0.317) and per-patient (P=0.125) basis.

Conclusions: Prospectively ECG-triggered high-pitch CCTA at 70kVp with 30mL contrast agent can reduce radiation dose but maintain image quality and high diagnostic accuracy in a selected, non-obese population.

Keywords: Coronary CT angiography; Diagnostic accuracy; High-pitch acquisition; Low contrast medium; Low tube voltage; Radiation dose.