Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2018 Feb;79(2):730-740.
doi: 10.1002/mrm.26726. Epub 2017 Jun 11.

Model-based T1 mapping with sparsity constraints using single-shot inversion-recovery radial FLASH

Affiliations
Free article

Model-based T1 mapping with sparsity constraints using single-shot inversion-recovery radial FLASH

Xiaoqing Wang et al. Magn Reson Med. 2018 Feb.
Free article

Abstract

Purpose: To develop a model-based reconstruction technique for single-shot T1 mapping with high spatial resolution, accuracy, and precision using an inversion-recovery (IR) fast low-angle shot (FLASH) acquisition with radial encoding.

Methods: The proposed model-based reconstruction jointly estimates all model parameters, that is, the equilibrium magnetization, steady-state magnetization, 1/ T1*, and all coil sensitivities from the data of a single-shot IR FLASH acquisition with a small golden-angle radial trajectory. Joint sparsity constraints on the parameter maps are exploited to improve the performance of the iteratively regularized Gauss-Newton method chosen for solving the nonlinear inverse problem. Validations include both a numerical and experimental T1 phantom, as well as in vivo studies of the human brain and liver at 3 T.

Results: In comparison to previous reconstruction methods for single-shot T1 mapping, which are based on real-time MRI with pixel-wise fitting and a model-based approach with a predetermination of coil sensitivities, the proposed method presents with improved robustness against phase errors and numerical precision in both phantom and in vivo studies.

Conclusion: The comprehensive model-based reconstruction with L1 regularization offers rapid and robust T1 mapping with high accuracy and precision. The method warrants accelerated computing and online implementation for extended clinical trials. Magn Reson Med 79:730-740, 2018. © 2017 International Society for Magnetic Resonance in Medicine.

Keywords: Look-Locker; T1 mapping; parallel imaging; model-based reconstruction; sparsity constraint.