. 2017 Dec;41(12):2503-2508.
doi: 10.1007/s00264-017-3528-9. Epub 2017 Jun 14.

Metallic ion release after knee prosthesis implantation: a prospective study

Affiliations

Metallic ion release after knee prosthesis implantation: a prospective study

Adrien Lons et al. Int Orthop. 2017 Dec.

Abstract

Introduction: Metal-on-metal (MoM) hip replacement bearings produce metallic ions that can cause health complications. Metallic release also occurs with other materials, but data on metallic ion levels after knee arthroplasty are sparse. We postulate that knee replacement generates elevating metallic ions (chromium (Cr), cobalt (Co) and titanium (Ti)) during the first year after implantation.

Patients and methods: This ongoing prospective study included all patients who underwent the same type of knee arthroplasty between May and December 2013. Cr, Co and Ti levels were measured in whole blood at pre-operation and one-year follow-up (6 and 12 months). Clinical and radiographic data (range of motion, Oxford, International Knee Society (IKS) and satisfaction scores) were recorded.

Results: In 90 patients, preoperative Cr, Co and Ti metallic ion levels were respectively 0.45 μg/l, 0.22 μg/l, 2.94 μg/l and increased to 1.27 μg/l, 1.41 μg/l, 4.08 μg/l (p < 0.0001) at last one-year follow-up. Mean Oxford and IKS scores rose, respectively, from 45.9 (30-58) and 24.9 (12-52) to 88.3 (0-168) and 160.8 (93-200) (p < 0.001).

Conclusion: After the implantation of knee arthroplasty, we found significant blood elevation of Cr, Co and Ti levels one year after implantation exceeding the normal values. This metallic ion release could lead to numerous effects: allergy, hypersensitivity, etc.

Keywords: Chromium; Cobalt; Knee prosthesis; Metallic ions; Titanium.

PubMed Disclaimer

References

    1. J Arthroplasty. 2000 Oct;15(7):938-43 - PubMed
    1. Acta Orthop Belg. 2010 Aug;76(4):513-20 - PubMed
    1. J Bone Joint Surg Am. 1998 Feb;80(2):268-82 - PubMed
    1. Biomed Res Int. 2014;2014:817257 - PubMed
    1. J Arthroplasty. 2011 Feb;26(2):282-8 - PubMed

LinkOut - more resources