Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2017 Aug:215:197-205.
doi: 10.1016/j.ejogrb.2017.05.027. Epub 2017 Jun 1.

Electrohysterography for uterine monitoring during term labour compared to external tocodynamometry and intra-uterine pressure catheter

Affiliations

Electrohysterography for uterine monitoring during term labour compared to external tocodynamometry and intra-uterine pressure catheter

Marion W C Vlemminx et al. Eur J Obstet Gynecol Reprod Biol. 2017 Aug.

Abstract

Objective: Current uterine monitoring techniques have major drawbacks that could be avoided when using electrohysterography for uterine monitoring. Recently, a new electrohysterography method has been developed, providing a real-time tocogram on standard cardiotocography monitors. The diagnostic characteristics of this novel method need to be determined and compared to conventional methods We hypothesised that electrohysterography can perform better than external tocodynamometry due to the adhesive properties of the contact electrodes (less motion sensitive), and the improved signal acquisition through subcutaneous tissue (less obesity sensitive).

Study design: In this prospective diagnostic accuracy study, uterine contractions of labouring women were simultaneously monitored by three different monitoring techniques: electrohysterography, external tocodynamometry, and intra-uterine pressure catheter as method of reference. We performed a two-hour measurement during first and/or second stage of term labour. The contractions of each method were automatically detected by a computer-based algorithm. As the applied method had not been described in literature before, an interim analysis was performed to minimise exposure to the invasive pressure catheter. The main outcome parameter was the sensitivity of electrohysterography in comparison to external tocodynamometry for uterine contraction detection, tested by the Wilcoxon signed rank test.

Results: Uterine contractions of 48 term labouring women were simultaneously monitored by electrohysterography, external tocodynamometry, and intra-uterine pressure catheter. The study was terminated after the interim analysis as the sensitivity of electrohysterography was significantly higher compared to external tocodynamometry: median 89.5% (interquartile range (IQR); 82-93) and 65.3% (IQR; 53-81) respectively, p<0.001. In a subgroup analysis of obese women (n=15), the sensitivity of electrohysterography was significantly higher than external tocodynamometry (median 88.4% (IQR; 79-95) and 45.8% (IQR; 38-61) respectively, p<0.001). Whereas in a subanalysis of second stage of labour (n=8), electrohysterography did not perform better than external tocodynamometry (median 72.8% (IQR; 61-87) and 66.4% (IQR; 46-75) respectively, p=0.225). Electrohysterography registered 0.4 more contractions per 10min than the intra-uterine pressure measurement (p<0.001) and 0.5 more contractions per 10min than external tocodynamometry (p<0.001).

Conclusion: Electrohysterography has a higher sensitivity for uterine contraction detection than external tocodynamometry during first stage of labour, in non-obese and obese women. Electrohysterography identifies more contractions than conventional techniques.