Poly-energetic and virtual mono-energetic images from a novel dual-layer spectral detector CT: optimization of window settings is crucial to improve subjective image quality in abdominal CT angiographies
- PMID: 28677003
- DOI: 10.1007/s00261-017-1241-1
Poly-energetic and virtual mono-energetic images from a novel dual-layer spectral detector CT: optimization of window settings is crucial to improve subjective image quality in abdominal CT angiographies
Abstract
Purpose: We aimed to determine optimal window settings for poly-energetic (PolyE) and virtual mono-energetic images (MonoE) derived from abdominal angiographic studies on a novel dual-layer spectral detector CT (SDCT) system.
Methods: From 50 patients, SDCT datasets PolyE and MonoE at 70 and 40 keV levels were reconstructed and best individual window width and level (BI-W/L) manually assessed. Through regression analysis, the so-called optimized individual (OI-W/L) values were obtained. Subjective image quality parameters and vessel diameters were measured to determine influences of different W/L settings.
Results: Image noise was lower and attenuation and contrast-to-noise ratio were higher in MonoE compared to PolyE (all p ≤ 0.002). Mean BI-W/L values for PolyE, 70, and 40 keV were 637/284, 647/291, and 1568/691, respectively. Mean OI-W/L values were 631/276, 628/286, and 1516/667, respectively. Compared to standard settings, all adjusted W/L settings varied significantly and yielded higher subjective scoring. No between-group differences were found between manually adjusted and mathematically calculated W/L settings.
Conclusion: PolyE and MonoE from abdominal angiographic SDCT studies require appropriate W/L settings especially at low energy reconstruction levels. Individual adjustment reaches the best image quality but is time consuming. From our data, predefined W/L settings of 640/280 (PolyE/MonoE 70 keV) and 1570/690 (MonoE 40 keV) as a non-individualized starting point for abdominal angiographic studies from the novel SDCT system are suggested.
Keywords: CT angiography; Dual-energy CT; Dual-layer spectral detector CT; Virtual mono-energetic imaging; Window setting.
Similar articles
-
Virtual Monoenergetic Images From a Novel Dual-Layer Spectral Detector Computed Tomography Scanner in Portal Venous Phase: Adjusted Window Settings Depending on Assessment Focus Are Essential for Image Interpretation.J Comput Assist Tomogr. 2018 May/Jun;42(3):350-356. doi: 10.1097/RCT.0000000000000711. J Comput Assist Tomogr. 2018. PMID: 29369944
-
Intra-individual comparison between abdominal virtual mono-energetic spectral and conventional images using a novel spectral detector CT.PLoS One. 2017 Aug 24;12(8):e0183759. doi: 10.1371/journal.pone.0183759. eCollection 2017. PLoS One. 2017. PMID: 28837641 Free PMC article.
-
Optimization of window settings for standard and advanced virtual monoenergetic imaging in abdominal dual-energy CT angiography.Abdom Radiol (NY). 2017 Mar;42(3):772-780. doi: 10.1007/s00261-016-0963-9. Abdom Radiol (NY). 2017. PMID: 27878637
-
Abdominal Applications of a Novel Detector-Based Spectral CT.Curr Probl Diagn Radiol. 2018 Mar-Apr;47(2):110-118. doi: 10.1067/j.cpradiol.2017.05.001. Epub 2017 May 17. Curr Probl Diagn Radiol. 2018. PMID: 28673603 Review.
-
Utility of Iodine Density Perfusion Maps From Dual-Energy Spectral Detector CT in Evaluating Cardiothoracic Conditions: A Primer for the Radiologist.AJR Am J Roentgenol. 2020 Apr;214(4):775-785. doi: 10.2214/AJR.19.21818. Epub 2020 Feb 11. AJR Am J Roentgenol. 2020. PMID: 32045305 Review.
Cited by
-
Dual-source photon-counting computed tomography for coronary in-stent observation: influence of heart rate and virtual monoenergetic image.Int J Cardiovasc Imaging. 2024 Oct;40(10):2117-2124. doi: 10.1007/s10554-024-03203-x. Epub 2024 Aug 12. Int J Cardiovasc Imaging. 2024. PMID: 39128972
-
A deep learning framework integrating MRI image preprocessing methods for brain tumor segmentation and classification.IBRO Neurosci Rep. 2022 Nov 7;13:523-532. doi: 10.1016/j.ibneur.2022.10.014. eCollection 2022 Dec. IBRO Neurosci Rep. 2022. PMID: 36590099 Free PMC article.
-
Coronary stent imaging in photon counting computed tomography: improved imaging of in-stent stenoses in a phantom with optimized reconstruction kernels.BJR Open. 2024 Oct 18;6(1):tzae030. doi: 10.1093/bjro/tzae030. eCollection 2024 Jan. BJR Open. 2024. PMID: 39444459 Free PMC article.
-
Dual-layer spectral computed tomography aortography using a seventy-five-percent-reduced iodine dose protocol and multiparameter spectral imaging: comparison with conventional computed tomography imaging.Quant Imaging Med Surg. 2023 Oct 1;13(10):6456-6467. doi: 10.21037/qims-23-101. Epub 2023 Sep 18. Quant Imaging Med Surg. 2023. PMID: 37869326 Free PMC article.
-
Computed tomography pulmonary angiograms using a novel dual-layer spectral detector: Adjusted window settings are essential for diagnostic image quality.Medicine (Baltimore). 2019 Aug;98(33):e16606. doi: 10.1097/MD.0000000000016606. Medicine (Baltimore). 2019. PMID: 31415352 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources