Efficacy and safety of the subcutaneous implantable cardioverter defibrillator: a systematic review
- PMID: 28687562
- DOI: 10.1136/heartjnl-2016-310852
Efficacy and safety of the subcutaneous implantable cardioverter defibrillator: a systematic review
Abstract
Background: Subcutaneous implantable cardioverter defibrillators (S-ICDs) are considered an alternative to conventional transvenous ICDs (TV-ICDs) in patients not requiring pacing.
Methods: We searched MEDLINE and EMBASE for studies evaluating efficacy and safety outcomes in S-ICD patients. Outcomes were pooled across studies.
Results: Sixteen studies were included with 5380 participants (mean age range 33-56 years). Short-term follow-up data were available for 1670 subjects. The most common complication was pocket infection, affecting 2.7%. Other complications included delayed wound healing (0.6%) and wound discomfort (0.8%). 3.8% of S-ICDs were explanted, most commonly for pocket infection. Mortality rates in hospital (0.4%) and during follow-up (3.4% from 12 studies reporting) were low. Incidence of ventricular arrhythmia varied from 0% to 12%. Overall shock efficacy exceeded 96%. Inappropriate shocks affected 4.3% and was most commonly caused by T-wave oversensing.
Conclusions: Although long-term randomised data are lacking, observational data suggest similar shock efficacy and short-term complication rates between the S-ICD and TV-ICD.
Keywords: implanted cardiac defibrillators; systematic review; ventricular fibrillation; ventricular tachycardia.
© Article author(s) (or their employer(s) unless otherwise stated in the text of the article) 2017. All rights reserved. No commercial use is permitted unless otherwise expressly granted.
Similar articles
-
A review of the evidence on the effects and costs of implantable cardioverter defibrillator therapy in different patient groups, and modelling of cost-effectiveness and cost-utility for these groups in a UK context.Health Technol Assess. 2006 Aug;10(27):iii-iv, ix-xi, 1-164. doi: 10.3310/hta10270. Health Technol Assess. 2006. PMID: 16904046
-
Transvenous versus subcutaneous implantable cardioverter defibrillators in young cardiac arrest survivors.Intern Med J. 2023 Nov;53(11):1956-1962. doi: 10.1111/imj.16259. Epub 2023 Nov 6. Intern Med J. 2023. PMID: 37929818 Review.
-
Subcutaneous Versus Transvenous Implantable Defibrillator Therapy: A Systematic Review and Meta-Analysis of Randomized Trials and Propensity Score-Matched Studies.J Am Heart Assoc. 2022 Jun 7;11(11):e024756. doi: 10.1161/JAHA.121.024756. Epub 2022 Jun 3. J Am Heart Assoc. 2022. PMID: 35656975 Free PMC article.
-
Implantable cardioverter-defibrillator harm in young patients with inherited arrhythmia syndromes: A systematic review and meta-analysis of inappropriate shocks and complications.Heart Rhythm. 2016 Feb;13(2):443-54. doi: 10.1016/j.hrthm.2015.09.010. Epub 2015 Sep 15. Heart Rhythm. 2016. PMID: 26385533
-
Performance and Safety of the Extravascular Implantable Cardioverter Defibrillator Through Long-Term Follow-Up: Final Results From the Pivotal Study.Circulation. 2025 Jan 28;151(4):322-332. doi: 10.1161/CIRCULATIONAHA.124.071795. Epub 2024 Sep 26. Circulation. 2025. PMID: 39327797 Free PMC article. Clinical Trial.
Cited by
-
The utility of routine clinical 12-lead ECG in assessing eligibility for subcutaneous implantable cardioverter defibrillator.Comput Biol Med. 2018 Nov 1;102:242-250. doi: 10.1016/j.compbiomed.2018.05.002. Epub 2018 May 8. Comput Biol Med. 2018. PMID: 29754992 Free PMC article.
-
Fibrous encapsulation of defibrillation electrode and elevated high-voltage impedance in patients with a subcutaneous implantable cardioverter-defibrillator.HeartRhythm Case Rep. 2019 Dec 16;6(3):148-152. doi: 10.1016/j.hrcr.2019.11.009. eCollection 2020 Mar. HeartRhythm Case Rep. 2019. PMID: 32181134 Free PMC article. No abstract available.
-
Systematic Review on S-ICD Lead Extraction.J Clin Med. 2023 May 27;12(11):3710. doi: 10.3390/jcm12113710. J Clin Med. 2023. PMID: 37297905 Free PMC article. Review.
-
Do all patients with implantable cardioverter defibrillator need a generator change? A health service evaluation of patients who underwent generator changes from a single tertiary center.Indian Pacing Electrophysiol J. 2020 Nov-Dec;20(6):257-260. doi: 10.1016/j.ipej.2020.08.003. Epub 2020 Aug 28. Indian Pacing Electrophysiol J. 2020. PMID: 32866597 Free PMC article.
-
Efficacy, safety, and in-hospital outcomes of subcutaneous versus transvenous implantable defibrillator therapy: A meta-analysis and systematic review.Medicine (Baltimore). 2019 May;98(19):e15490. doi: 10.1097/MD.0000000000015490. Medicine (Baltimore). 2019. PMID: 31083185 Free PMC article.