Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2017 Jul 7;7(1):4920.
doi: 10.1038/s41598-017-05249-5.

John's Equation-based Consistency Condition and Corrupted Projection Restoration in Circular Trajectory Cone Beam CT

Affiliations

John's Equation-based Consistency Condition and Corrupted Projection Restoration in Circular Trajectory Cone Beam CT

Jianhui Ma et al. Sci Rep. .

Abstract

In transmitted X-ray tomography imaging, the acquired projections may be corrupted for various reasons, such as defective detector cells and beam-stop array scatter correction problems. In this study, we derive a consistency condition for cone-beam projections and propose a method to restore lost data in corrupted projections. In particular, the relationship of the geometry parameters in circular trajectory cone-beam computed tomography (CBCT) is utilized to convert an ultra-hyperbolic partial differential equation (PDE) into a second-order PDE. The second-order PDE is then transformed into a first-order ordinary differential equation in the frequency domain. The left side of the equation for the newly derived consistency condition is the projection derivative of the current and adjacent views, whereas the right side is the projection derivative of the geometry parameters. A projection restoration method is established based on the newly derived equation to restore corrupted data in projections in circular trajectory CBCT. The proposed method is tested in beam-stop array scatter correction, metal artifact reduction, and abnormal pixel correction cases to evaluate the performance of the consistency condition and corrupted projection restoration method. Qualitative and quantitative results demonstrate that the present method has considerable potential in restoring lost data in corrupted projections.

PubMed Disclaimer

Conflict of interest statement

The authors declare that they have no competing interests.

Figures

Figure 1
Figure 1
Geometry configuration of circular trajectory CBCT.
Figure 2
Figure 2
Illustration of the corrupted projection restoration capabilities of JECC and SI in the frequency domain.
Figure 3
Figure 3
Workflow of the JECC method in circular trajectory CBCT.
Figure 4
Figure 4
Configuration of moving BSA scatter correction case. (a) BSA in odd-numbered views (mode I); (b) BSA in even-numbered views (mode II); (c) projections in a circular trajectory by moving the BSA scatter correction protocol. The odd-numbered views use mode I in (a), and the even-numbered views use mode II in (b).
Figure 5
Figure 5
In-house bench-top CBCT system for raw projection acquisition.
Figure 6
Figure 6
JECC performance on different slices. (a) Reference image on the 100th slice; (bd) images on the 100th slice reconstructed using the SI method, JECC 1st iteration, and JECC 4th iteration, respectively; (e) reference image on the 80th slice; (f–h) images on the 80th slice reconstructed using the SI method, JECC 1st iteration, and JECC 4th iteration, respectively.
Figure 7
Figure 7
Image profiles of the results in Fig. 6. (a) Profiles across the 95th to 145th columns in the 255th row of the results indicated by the blue line in Fig. 6(a); (b) profiles across the 225th to 285th rows in the 401th column of the results indicated by the blue line in Fig. 6(e).
Figure 8
Figure 8
Comparison of the different methods in terms of UQI.
Figure 9
Figure 9
Image without correction and images reconstructed using the SI method and the JECC method. The first row presents the images on the 190th slice, and the third row presents the images on the 220th. The second and fourth rows present the magnified ROIs marked with yellow squares in the images in the first and third rows.
Figure 10
Figure 10
Images corrected with SI method and the proposed method. The first and third rows present two representative slices. The second and fourth rows present the zoomed ROIs marked with blue dashed squares in the first and third rows.
Figure 11
Figure 11
Image profiles of the results in Fig. 9. (a) Profiles across the 315th to 375th rows in the 145th column of the results indicated by a blue line in the top row in Fig. 9; (b) profiles across the 345th to 385th rows in the 170th column of the results indicated by a blue line in the third row in Fig. 9.
Figure 12
Figure 12
Comparison of the different methods in terms of UQI.
Figure 13
Figure 13
Bilateral hip prostheses simulation. The transverse, sagittal, and coronal views are displayed in the first, second, and third rows, respectively.
Figure 14
Figure 14
Image Profiles of the results in Fig. 13. (a) Profiles indicated by the blue dashed line in transverse view; (b) profiles indicated by the blue dashed line in coronal view.
Figure 15
Figure 15
Two clinical cases are displayed here. The first row is the head case with a brain stimulator and the second row presents the corresponding magnified ROIs which are the blue dashed squares containing the metal implant, the bottom two rows present the dental case with three dental fillings on the back teeth and one dental filling on the front tooth.
Figure 16
Figure 16
Frequency domain illustrations of the same image. (a) Frequency domain of the image obtained using the SI method; (b) frequency domain of the image obtained using the JECC method; (c) frequency domain of the difference between the images in (a) and (b).

Similar articles

Cited by

References

    1. Tang X, Ning R, Yu R, Conover D. Cone beam volume CT image artifacts caused by defective cells in x-ray flat panel imagers and the artifact removal using a wavelet-analysis-based algorithm. Med. Phys. 2001;28:812–825. doi: 10.1118/1.1368878. - DOI - PubMed
    1. Prell D, Kyriakou Y, Kalender WA. Comparison of ring artifact correction methods for flat-detector CT. Phys. Med. Biol. 2009;54:3881–3895. doi: 10.1088/0031-9155/54/12/018. - DOI - PubMed
    1. Anas EMA, Kim JG, Lee SY, Hasan MK. Comparison of ring artifact removal methods using flat panel detector based CT images. Biomed. Eng. Online. 2011;10:1–25. doi: 10.1186/1475-925X-10-1. - DOI - PMC - PubMed
    1. Anas EMA, Lee SY, Hasan MK. Classification of ring artifacts for their effective removal using type adaptive correction schemes. Comput. Biol. Med. 2011;41:390–401. doi: 10.1016/j.compbiomed.2011.03.018. - DOI - PubMed
    1. Anas EMA, Lee SY, Hasan MK. Removal of ring artifacts in CT imaging through detection and correction of stripes in the sinogram. Phys. Med. Biol. 2010;55:6911–6930. doi: 10.1088/0031-9155/55/22/020. - DOI - PubMed

Publication types

LinkOut - more resources