The insertion torque-depth curve integral as a measure of implant primary stability: An in vitro study on polyurethane foam blocks
- PMID: 28689908
- DOI: 10.1016/j.prosdent.2017.04.012
The insertion torque-depth curve integral as a measure of implant primary stability: An in vitro study on polyurethane foam blocks
Abstract
Statement of problem: Recent research has shown that dynamic parameters correlate with insertion energy-that is, the total work needed to place an implant into its site-might convey more reliable information concerning immediate implant primary stability at insertion than the commonly used insertion torque (IT), the reverse torque (RT), or the implant stability quotient (ISQ). Yet knowledge on these dynamic parameters is still limited.
Purpose: The purpose of this in vitro study was to evaluate whether an energy-related parameter, the torque-depth curve integral (I), could be a reliable measure of primary stability. This was done by assessing if (I) measurement was operator-independent, by investigating its correlation with other known primary stability parameters (IT, RT, or ISQ) by quantifying the (I) average error and correlating (I), IT, RT, and ISQ variations with bone density.
Material and methods: Five operators placed 200 implants in polyurethane foam blocks of different densities using a micromotor that calculated the (I) during implant placement. Primary implant stability was assessed by measuring the ISQ, IT, and RT. ANOVA tests were used to evaluate whether measurements were operator independent (P>.05 in all cases). A correlation analysis was performed between (I) and IT, ISQ, and RT. The (I) average error was calculated and compared with that of the other parameters by ANOVA. (I)-density, IT-density, ISQ-density, and RT-density plots were drawn, and their slopes were compared by ANCOVA.
Results: The (I) measurements were operator independent and correlated with IT, ISQ, and RT. The average error of these parameters was not significantly different (P>.05 in all cases). The (I)-density, IT-density, ISQ-density, and RT-density curves were linear in the 0.16 to 0.49 g/cm³ range, with the (I)-density curves having a significantly greater slope than those regarding the other parameters (P≤.001 in all cases).
Conclusions: The torque-depth curve integral (I) provides a reliable assessment of primary stability and shows a greater sensitivity to density variations than other known primary stability parameters.
Copyright © 2017 Editorial Council for the Journal of Prosthetic Dentistry. Published by Elsevier Inc. All rights reserved.
Similar articles
-
Cortical Thickness, Bone Density, and the Insertion Torque/Depth Integral: A Study Using Polyurethane Foam Blocks.Int J Oral Maxillofac Implants. 2021 May-Jun;36(3):423-431. doi: 10.11607/jomi.8419. Int J Oral Maxillofac Implants. 2021. PMID: 34115054
-
Correlation Between Bone Density and Instantaneous Torque at Implant Site Preparation: A Validation on Polyurethane Foam Blocks of a Device Assessing Density of Jawbones.Int J Oral Maxillofac Implants. 2016 Sep-Oct;31(5):e128-35. doi: 10.11607/jomi.4475. Int J Oral Maxillofac Implants. 2016. PMID: 27632279
-
Influence of surgical technique and surface roughness on the primary stability of an implant in artificial bone with different cortical thickness: a laboratory study.Clin Oral Implants Res. 2010 Feb;21(2):213-20. doi: 10.1111/j.1600-0501.2009.01823.x. Clin Oral Implants Res. 2010. PMID: 20070754
-
Relationship Between Primary/Mechanical and Secondary/Biological Implant Stability.Int J Oral Maxillofac Implants. 2019 Suppl;34:s7-s23. doi: 10.11607/jomi.19suppl.g1. Int J Oral Maxillofac Implants. 2019. PMID: 31116830 Review.
-
The Interrelation between Cortical Bone Thickness and Primary and Secondary Dental Implant Stability: a Systematic Review.J Oral Maxillofac Res. 2024 Dec 31;15(4):e2. doi: 10.5037/jomr.2024.15402. eCollection 2024 Oct-Dec. J Oral Maxillofac Res. 2024. PMID: 40017687 Free PMC article. Review.
Cited by
-
Primary Implant Stability Analysis of Different Dental Implant Connections and Designs-An In Vitro Comparative Study.Materials (Basel). 2022 Apr 23;15(9):3072. doi: 10.3390/ma15093072. Materials (Basel). 2022. PMID: 35591407 Free PMC article.
-
The effect of undersizing and tapping on bone to implant contact and implant primary stability: A histomorphometric study on bovine ribs.J Adv Prosthodont. 2018 Jun;10(3):227-235. doi: 10.4047/jap.2018.10.3.227. Epub 2018 Jun 12. J Adv Prosthodont. 2018. PMID: 29930793 Free PMC article.
-
Effects of implant thread design on primary stability-a comparison between single- and double-threaded implants in an artificial bone model.Int J Implant Dent. 2020 Aug 20;6(1):42. doi: 10.1186/s40729-020-00239-1. Int J Implant Dent. 2020. PMID: 32815046 Free PMC article.
-
Bone Density and Implant Primary Stability. A Study on Equine Bone Blocks.Dent J (Basel). 2019 Jul 1;7(3):73. doi: 10.3390/dj7030073. Dent J (Basel). 2019. PMID: 31266214 Free PMC article.
-
A New Site Preparation Protocol That Supports Bone Quality Evaluation and Provides Predictable Implant Insertion Torque.J Clin Med. 2020 Feb 11;9(2):494. doi: 10.3390/jcm9020494. J Clin Med. 2020. PMID: 32054122 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials