Abstract
Background: Valve-in-valve (VIV) transcatheter aortic valve replacement (TAVR) may be less effective in small surgical valves because of patient/prosthesis mismatch. Bioprosthetic valve fracture (BVF) using a high-pressure balloon can be performed to facilitate VIV TAVR.
Methods and results: We report data from 20 consecutive clinical cases in which BVF was successfully performed before or after VIV TAVR by inflation of a high-pressure balloon positioned across the valve ring during rapid ventricular pacing. Hemodynamic measurements and calculation of the valve effective orifice area were performed at baseline, immediately after VIV TAVR, and after BVF. BVF was successfully performed in 20 patients undergoing VIV TAVR with balloon-expandable (n=8) or self-expanding (n=12) transcatheter valves in Mitroflow, Carpentier-Edwards Perimount, Magna and Magna Ease, Biocor Epic and Biocor Epic Supra, and Mosaic surgical valves. Successful fracture was noted fluoroscopically when the waist of the balloon released and by a sudden drop in inflation pressure, often accompanied by an audible snap. BVF resulted in a reduction in the mean transvalvular gradient (from 20.5±7.4 to 6.7±3.7 mm Hg, P<0.001) and an increase in valve effective orifice area (from 1.0±0.4 to 1.8±0.6 cm2, P<0.001). No procedural complications were reported.
Conclusions: BVF can be performed safely in small surgical valves to facilitate VIV TAVR with either balloon-expandable or self-expanding transcatheter valves and results in reduced residual transvalvular gradients and increased valve effective orifice area.
Keywords: aortic stenosis; bioprosthesis; transcatheter aortic valve replacement.
© 2017 American Heart Association, Inc.
Comment in
-
When Device Failure Translates to Therapeutic Efficacy: Intentional Fracture of Bioprosthetic Valve Rings as an Adjunctive Technique in Transcatheter Valve-in-Valve Replacement.Circ Cardiovasc Interv. 2017 Jul;10(7):e005662. doi: 10.1161/CIRCINTERVENTIONS.117.005662. Circ Cardiovasc Interv. 2017. PMID: 28698294 No abstract available.
Similar articles
-
Bioprosthetic valve fracture: Technical insights from a multicenter study.J Thorac Cardiovasc Surg. 2019 Nov;158(5):1317-1328.e1. doi: 10.1016/j.jtcvs.2019.01.073. Epub 2019 Jan 31. J Thorac Cardiovasc Surg. 2019. PMID: 30857820
-
Is Bioprosthetic Valve Fracture Beneficial for Patients with Larger Surgical Valves Undergoing Valve in Valve TAVR?Curr Cardiol Rep. 2018 Aug 22;20(10):95. doi: 10.1007/s11886-018-1035-2. Curr Cardiol Rep. 2018. PMID: 30135999 Review.
-
Valve-in-Valve Transcatheter Aortic Valve Replacement and Bioprosthetic Valve Fracture Comparing Different Transcatheter Heart Valve Designs: An Ex Vivo Bench Study.JACC Cardiovasc Interv. 2019 Jan 14;12(1):65-75. doi: 10.1016/j.jcin.2018.10.043. JACC Cardiovasc Interv. 2019. PMID: 30621980
-
Transcatheter Replacement of Failed Bioprosthetic Valves: Large Multicenter Assessment of the Effect of Implantation Depth on Hemodynamics After Aortic Valve-in-Valve.Circ Cardiovasc Interv. 2016 Jun;9(6):e003651. doi: 10.1161/CIRCINTERVENTIONS.115.003651. Circ Cardiovasc Interv. 2016. PMID: 27301396
-
Transcatheter Aortic Valve-in-Valve Procedure in Patients with Bioprosthetic Structural Valve Deterioration.Methodist Debakey Cardiovasc J. 2017 Jul-Sep;13(3):132-141. doi: 10.14797/mdcj-13-3-132. Methodist Debakey Cardiovasc J. 2017. PMID: 29743998 Free PMC article. Review.
Cited by
-
Balloon Aortic Valvuloplasty in the Modern Era: A Review of Outcomes, Indications, and Technical Advances.J Soc Cardiovasc Angiogr Interv. 2023 Jul 27;2(4):101002. doi: 10.1016/j.jscai.2023.101002. eCollection 2023 Jul-Aug. J Soc Cardiovasc Angiogr Interv. 2023. PMID: 39131636 Free PMC article. Review.
-
Aortic Valve-in-Valve Procedures: Challenges and Future Directions.J Clin Med. 2024 Aug 12;13(16):4723. doi: 10.3390/jcm13164723. J Clin Med. 2024. PMID: 39200865 Free PMC article. Review.
-
Transcatheter valve-in-valve implantation for degenerated surgical bioprostheses.J Thorac Dis. 2018 Nov;10(Suppl 30):S3573-S3577. doi: 10.21037/jtd.2018.05.66. J Thorac Dis. 2018. PMID: 30505537 Free PMC article. Review.
-
Pre-dilation and Post-dilation in Transcatheter Aortic Valve Replacement: Indications, Benefits and Risks.Interv Cardiol. 2021 Oct 12;16:e28. doi: 10.15420/icr.2020.35. eCollection 2021 Apr. Interv Cardiol. 2021. PMID: 34721667 Free PMC article. Review.
-
5 Year Outcomes of Patients With Aortic Structural Valve Deterioration Treated With Transcatheter Valve in Valve - A Single Center Prospective Registry.Front Cardiovasc Med. 2021 Sep 9;8:713341. doi: 10.3389/fcvm.2021.713341. eCollection 2021. Front Cardiovasc Med. 2021. PMID: 34568456 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical