. 2017 Nov 27;56(1):51-58.
doi: 10.1515/cclm-2017-0175.

Determination of serum calcium levels by 42Ca isotope dilution inductively coupled plasma mass spectrometry

Affiliations

Determination of serum calcium levels by 42Ca isotope dilution inductively coupled plasma mass spectrometry

Bingqing Han et al. Clin Chem Lab Med. .

Abstract

Background: Serum calcium level is an important clinical index that reflects pathophysiological states. However, detection accuracy in laboratory tests is not ideal; as such, a high accuracy method is needed.

Methods: We developed a reference method for measuring serum calcium levels by isotope dilution inductively coupled plasma mass spectrometry (ID ICP-MS), using 42Ca as the enriched isotope. Serum was digested with 69% ultrapure nitric acid and diluted to a suitable concentration. The 44Ca/42Ca ratio was detected in H2 mode; spike concentration was calibrated by reverse IDMS using standard reference material (SRM) 3109a, and sample concentration was measured by a bracketing procedure. We compared the performance of ID ICP-MS with those of three other reference methods in China using the same serum and aqueous samples.

Results: The relative expanded uncertainty of the sample concentration was 0.414% (k=2). The range of repeatability (within-run imprecision), intermediate imprecision (between-run imprecision), and intra-laboratory imprecision were 0.12%-0.19%, 0.07%-0.09%, and 0.16%-0.17%, respectively, for two of the serum samples. SRM909bI, SRM909bII, SRM909c, and GBW09152 were found to be within the certified value interval, with mean relative bias values of 0.29%, -0.02%, 0.10%, and -0.19%, respectively. The range of recovery was 99.87%-100.37%. Results obtained by ID ICP-MS showed a better accuracy than and were highly correlated with those of other reference methods.

Conclusions: ID ICP-MS is a simple and accurate candidate reference method for serum calcium measurement and can be used to establish and improve serum calcium reference system in China.

Keywords: high accuracy; isotope dilution inductively coupled plasma mass spectrometry (ID ICP-MS); method comparison; serum calcium.