Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2017 Aug 14;10(15):1578-1587.
doi: 10.1016/j.jcin.2017.05.031. Epub 2017 Jul 19.

CT-Defined Prosthesis-Patient Mismatch Downgrades Frequency and Severity, and Demonstrates No Association With Adverse Outcomes After Transcatheter Aortic Valve Replacement

Affiliations
Free article
Comparative Study

CT-Defined Prosthesis-Patient Mismatch Downgrades Frequency and Severity, and Demonstrates No Association With Adverse Outcomes After Transcatheter Aortic Valve Replacement

John Mooney et al. JACC Cardiovasc Interv. .
Free article

Abstract

Objectives: This study sought to determine if indexed effective orifice area (EOAi), using left ventricular outflow tract measured from computed tomography (EOAiCT), reclassified prosthesis-patient mismatch (PPM) compared with conventional echocardiogram-defined measurements (EOAiTTE).

Background: PPM does not predict mortality following transcatheter aortic valve replacement (TAVR). However, it is unknown if the EOAiCT of the left ventricular outflow tract improves risk stratification.

Methods: A total of 765 TAVR patients from the PARTNER II (Placement of Aortic Transcatheter Valves II) trial S3i cohort were evaluated. EOAi was calculated using the continuity equation, and the left ventricular outflow tract area was derived from baseline computed tomography. Traditional echocardiographic categories defined PPM: absent (>0.85 cm2/m2), moderate (≥0.65 and ≤0.85 cm2/m2), or severe (≤0.65 cm2/m2). Correlation of EOAiCT and EOAiTTE to 1-year outcomes was performed.

Results: The incidence of PPM was 24% with EOACT compared with 45% with EOAiTTE. Only 6% of PPM was graded severe by EOAiCT compared with 9% by EOAiTTE. EOAiTTE, but not EOAiCT, defined PPM showed association with reduced left ventricular mass regression (p = 0.03 vs. p = 0.52). There was no association between PPM and death or rehospitalization at 1 year with either modality. EOACT was associated with minor stroke at 1 year (log-rank p = 0.04), and EOAiTTE with stroke/transient ischemic attack (log-rank p = 0.030). Furthermore, when subjects with mild or greater paravalvular regurgitation were excluded, the presence of PPM did not show association with any outcome.

Conclusions: EOAiCT downgrades frequency and severity of PPM in patients after TAVR, and was not associated with mortality 1 year after TAVR. EOAiTTE, but not EOAiCT, was associated with less left ventricular mass regression.

Keywords: CT; echocardiography; prosthesis–patient mismatch; transcatheter aortic valve replacement.