Carbon Dioxide Flush of an Integrated Minimized Perfusion Circuit Prior to Priming Prevents Spontaneous Air Release Into the Arterial Line During Clinical Use
- PMID: 28741663
- DOI: 10.1111/aor.12909
Carbon Dioxide Flush of an Integrated Minimized Perfusion Circuit Prior to Priming Prevents Spontaneous Air Release Into the Arterial Line During Clinical Use
Abstract
Recently, an oxygenator with an integrated centrifugal blood pump (IP) was designed to minimize priming volume and to reduce blood foreign surface contact even further. The use of this oxygenator with or without integrated arterial filter was compared with a conventional oxygenator and nonintegrated centrifugal pump. To compare the air removal characteristics 60 patients undergoing coronary artery bypass grafting were alternately assigned into one of three groups to be perfused with a minimized extracorporeal circuit either with the conventional oxygenator, the oxygenator with IP, or the oxygenator with IP plus integrated arterial filter (IAF). Air entering and leaving the three devices was measured accurately with a bubble counter during cardiopulmonary bypass. No significant differences between all groups were detected, considering air entering the devices. Our major finding was that in both integrated devices groups incidental spontaneous release of air into the arterial line in approximately 40% of the patients was observed. Here, detectable bolus air (>500 µm) was shown in the arterial line, whereas in the minimal extracorporeal circulation circuit (MECC) group this phenomenon was not present. We decided to conduct an amendment of the initial design with METC-approval. Ten patients were assigned to be perfused with an oxygenator with IP and IAF. Importantly, the integrated perfusion systems used in these patients were flushed with carbon dioxide (CO2 ) prior to priming of the systems. In the group with CO2 flush no spontaneous air release was observed in all cases and this was significantly different from the initial study with the group with the integrated device and IAF. This suggests that air spilling may be caused by residual air in the integrated device. In conclusion, integration of a blood pump may cause spontaneous release of large air bubbles (>500 µm) into the arterial line, despite the presence of an integrated arterial filter. CO2 flushing of an integrated cardiopulmonary bypass system prior to priming may prevent spontaneous air release and is strongly recommended to secure patient safety.
Keywords: -Air removal; -Cardiopulmonary bypass; -Gaseous microemboli; -Integration; -Minimized circuit; -Oxygenator; Carbon dioxide flush.
© 2017 International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.
Similar articles
-
Clinical evaluation of the air removal characteristics of an oxygenator with integrated arterial filter in a minimized extracorporeal circuit.Int J Artif Organs. 2011 Apr;34(4):374-82. doi: 10.5301/IJAO.2011.7749. Int J Artif Organs. 2011. PMID: 21534248 Clinical Trial.
-
Pulsatile flow decreases gaseous micro-bubble filtering properties of oxygenators without integrated arterial filters during cardiopulmonary bypass.Interact Cardiovasc Thorac Surg. 2013 Nov;17(5):811-7. doi: 10.1093/icvts/ivt264. Epub 2013 Jul 9. Interact Cardiovasc Thorac Surg. 2013. PMID: 23842758 Free PMC article. Clinical Trial.
-
In vitro air removal characteristics of two neonatal cardiopulmonary bypass systems: filtration may lead to fractionation of bubbles.Int J Artif Organs. 2014 Sep;37(9):688-96. doi: 10.5301/ijao.5000348. Epub 2014 Sep 20. Int J Artif Organs. 2014. PMID: 25262633
-
Impact of oxygenator characteristics on its capability to remove gaseous microemboli.J Extra Corpor Technol. 2007 Dec;39(4):271-3. J Extra Corpor Technol. 2007. PMID: 18293817 Review.
-
Pumping O2 with no N2: An Overview of Hollow Fiber Membrane Oxygenators with Integrated Arterial Filters.Curr Top Med Chem. 2020;20(1):78-85. doi: 10.2174/1568026619666191210161013. Curr Top Med Chem. 2020. PMID: 31820691 Review.
Cited by
-
2024 EACTS/EACTAIC/EBCP Guidelines on cardiopulmonary bypass in adult cardiac surgery.Br J Anaesth. 2025 Apr;134(4):917-1008. doi: 10.1016/j.bja.2025.01.015. Epub 2025 Feb 14. Br J Anaesth. 2025. PMID: 39955230 Free PMC article.
-
2019 EACTS/EACTA/EBCP guidelines on cardiopulmonary bypass in adult cardiac surgery.Interact Cardiovasc Thorac Surg. 2020 Feb 1;30(2):161-202. doi: 10.1093/icvts/ivz251. Interact Cardiovasc Thorac Surg. 2020. PMID: 31576402 Free PMC article. No abstract available.
-
Recent innovations in perfusion and cardiopulmonary bypass for neonatal and infant cardiac surgery.Transl Pediatr. 2018 Apr;7(2):139-150. doi: 10.21037/tp.2018.03.05. Transl Pediatr. 2018. PMID: 29770295 Free PMC article. Review.
-
2024 EACTS/EACTAIC/EBCP Guidelines on cardiopulmonary bypass in adult cardiac surgery.Interdiscip Cardiovasc Thorac Surg. 2025 Feb 5;40(2):ivaf002. doi: 10.1093/icvts/ivaf002. Interdiscip Cardiovasc Thorac Surg. 2025. PMID: 39949317 Free PMC article. No abstract available.
-
2024 EACTS/EACTAIC/EBCP Guidelines on cardiopulmonary bypass in adult cardiac surgery.Eur J Cardiothorac Surg. 2025 Feb 4;67(2):ezae354. doi: 10.1093/ejcts/ezae354. Eur J Cardiothorac Surg. 2025. PMID: 39949326 Free PMC article. No abstract available.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical