The Evaluation of a Noninvasive Respiratory Volume Monitor in Pediatric Patients Undergoing General Anesthesia
- PMID: 28759491
- DOI: 10.1213/ANE.0000000000002029
The Evaluation of a Noninvasive Respiratory Volume Monitor in Pediatric Patients Undergoing General Anesthesia
Abstract
Background: Pediatric patients following surgery are at risk for respiratory compromise such as hypoventilation and hypoxemia depending on their age, comorbidities, and type of surgery. Quantitative measurement of ventilation in nonintubated infants/children is a difficult and inexact undertaking. Current respiratory assessment in nonintubated patients relies on oximetry data, respiratory rate (RR) monitors, and subjective clinical assessment, but there is no objective measure of respiratory parameters that could be utilized to predict early respiratory compromise. New advances in technology and digital signal processing have led to the development of an impedance-based respiratory volume monitor (RVM, ExSpiron, Respiratory Motion, Inc, Waltham, MA). The RVM has been shown to provide accurate real-time, continuous, noninvasive measurements of tidal volume (TV), minute ventilation (MV), and RR in adult patients.In this prospective observational study, our primary aim was to determine whether the RVM accurately measures TV, RR, and MV in pediatric patients.
Methods: A total of 72 pediatric patients (27 females, 45 males), ASA I to III, undergoing general anesthesia with endotracheal intubation were enrolled. After endotracheal intubation, continuous data of MV, TV, and RR were recorded from the RVM and an in-line monitoring spirometer (NM3 monitor, Phillips Healthcare). RVM and NM3 measurements of MV, TV, and RR were compared during a 10-minute period prior to the incision ("Presurgery") and a 10-minute period after the end of surgery ("Postsurgery"). Relative errors were calculated over 1-minute segment within each 10-minute period. Bias, precision, and accuracy were calculated using Bland-Altman analyses and paired-difference equivalence tests were performed.
Results: Combined across the Presurgery and Postsurgery periods, the RVM's mean measurement bias (RVM - NM3 measurement) for MV was -3.8% (95% limits of agreement) (±1.96 SD): (-19.9% to 12.2%), for TV it was -4.9 (-21.0% to 11.3%), and for RR it was 1.1% (-4.1% to 6.2%). The mean measurement accuracies for MV, TV, and RR were 11.9%, 12.0%, and 4.2% (0.6 breaths/min), respectively. Note that lower accuracy numbers correspond to more accurate RVM measurements. The equivalence tests rejected the null hypothesis that the RVM and NM3 have different mean values and conclude with 90% power that the measurements of MV, TV, and RR from the RVM and NM3 are equivalent within ±10%.
Conclusions: Our data indicate acceptable agreement between RVM and NM3 measurements in pediatric mechanically-ventilated patients. Future studies assessing the capability of the RVM to detect respiratory compromise in other clinical settings are needed.
Similar articles
-
The Evaluation of a Noninvasive Respiratory Volume Monitor in Mechanically Ventilated Neonates and Infants.Anesth Analg. 2022 Jan 1;134(1):141-148. doi: 10.1213/ANE.0000000000005562. Anesth Analg. 2022. PMID: 33929346
-
Special article: evaluation of a novel noninvasive respiration monitor providing continuous measurement of minute ventilation in ambulatory subjects in a variety of clinical scenarios.Anesth Analg. 2013 Jul;117(1):91-100. doi: 10.1213/ANE.0b013e3182918098. Epub 2013 Jun 3. Anesth Analg. 2013. PMID: 23733842
-
A Comparison of Measurements of Change in Respiratory Status in Spontaneously Breathing Volunteers by the ExSpiron Noninvasive Respiratory Volume Monitor Versus the Capnostream Capnometer.Anesth Analg. 2017 Jan;124(1):120-126. doi: 10.1213/ANE.0000000000001395. Anesth Analg. 2017. PMID: 27384980
-
A Brief Review of Non-invasive Monitoring of Respiratory Condition for Extubated Patients with or at Risk for Obstructive Sleep Apnea after Surgery.Front Med (Lausanne). 2017 Mar 8;4:26. doi: 10.3389/fmed.2017.00026. eCollection 2017. Front Med (Lausanne). 2017. PMID: 28337439 Free PMC article. Review.
-
Respiratory rates in pediatric emergency patients.J Emerg Med. 1992 Jul-Aug;10(4):407-10. doi: 10.1016/0736-4679(92)90268-x. J Emerg Med. 1992. PMID: 1430976 Review.
Cited by
-
Breathing variability during propofol/remifentanil procedural sedation with a single additional dose of midazolam or s-ketamine: a prospective observational study.J Clin Monit Comput. 2022 Aug;36(4):1219-1225. doi: 10.1007/s10877-021-00773-2. Epub 2021 Nov 12. J Clin Monit Comput. 2022. PMID: 34767130 Free PMC article.
-
Variations in respiratory rate do not reflect changes in tidal volume or minute ventilation after major abdominal surgery.J Clin Monit Comput. 2021 Aug;35(4):787-796. doi: 10.1007/s10877-020-00538-3. Epub 2020 Jun 1. J Clin Monit Comput. 2021. PMID: 32488678 Free PMC article.
-
Smart Vest for Respiratory Rate Monitoring of COPD Patients Based on Non-Contact Capacitive Sensing.Sensors (Basel). 2018 Jul 3;18(7):2144. doi: 10.3390/s18072144. Sensors (Basel). 2018. PMID: 29970861 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical