Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2017 Jul 31;12(7):e0182167.
doi: 10.1371/journal.pone.0182167. eCollection 2017.

Contamination with HIV antibody may be responsible for false positive results in specimens tested on automated platforms running HIV 4th generation assays in a region of high HIV prevalence

Affiliations

Contamination with HIV antibody may be responsible for false positive results in specimens tested on automated platforms running HIV 4th generation assays in a region of high HIV prevalence

Diana Ruth Hardie et al. PLoS One. .

Abstract

Introduction: In South Africa where the prevalence of HIV infection is very high, 4th generation HIV antibody/p24 antigen combo immunoassays are the tests of choice for laboratory based screening. Testing is usually performed in clinical pathology laboratories on automated analysers. To investigate the cause of false positive results on 4th generation HIV testing platforms in public sector laboratories, the performance of two automated platforms was compared in a clinical pathology setting, firstly on routine diagnostic specimens and secondly on known sero-negative samples.

Methods: Firstly, 1181 routine diagnostic specimens were sequentially tested on Siemens and Roche automated 4th generation platforms. HIV viral load, western blot and follow up testing were used to determine the true status of inconclusive specimens. Subsequently, known HIV seronegative samples from a single donor were repeatedly tested on both platforms and an analyser was tested for surface contamination with HIV positive serum to identify how suspected specimen contamination could be occurring.

Results: Serial testing of diagnostic specimens yielded 163 weakly positive or discordant results. Only 3 of 163 were conclusively shown to indicate true HIV infection. Specimen contamination with HIV antibody was suspected, based on the following evidence: the proportion of positive specimens increased on repeated passage through the analysers; viral loads were low or undetectable and western blots negative or indeterminate on problem specimens; screen negative, 2nd test positive specimens tested positive when reanalysed on the screening assay; follow up specimens (where available) were negative. Similarly, an increasing number of known negative specimens became (repeatedly) sero-positive on serial passage through one of the analysers. Internal and external analyser surfaces were contaminated with HIV serum, evidence that sample splashes occur during testing.

Conclusions: Due to the extreme sensitivity of these assays, contamination with minute amounts of HIV antibody can cause a negative sample to test positive. Better contamination control measures are needed on analysers used in clinical pathology environments, especially in regions where HIV sero-prevalence is high.

PubMed Disclaimer

Conflict of interest statement

Competing Interests: The authors have declared that no competing interests exist.

Figures

Fig 1
Fig 1. Study plan.
(a) 1181 consecutive routine diagnostic samples (serum or plasma) tested on the Siemens HIV 4th generation platform at the primary testing laboratories were re-tested on the Roche platform at the referral laboratory UN (using the same primary tube). Inconclusive specimens were further investigated with HIV viral load and western blot where specimen volume permitted. A subset of 26 specimens which were screen negative, second test positive was retested on the Siemens (initial screening assay) at the JE laboratory. (b) 210 known HIV antibody negative, NAT negative specimens from the same donor were tested on the Siemens and Roche platforms at JE and UN respectively. 105 specimens were tested on each analyser over a 10 day period and handled as normal clinical specimens. After the initial HIV serology test, the specimens were sent through the same analyser and tested for a second time.
Fig 2
Fig 2. Determining sero-positivity at first and second testing episodes.
The number of negative results decreased and the number of positive results increased on second testing on the Roche analyser at UN laboratory.
Fig 3
Fig 3. S/CO values of the positive test in S+/R- and S-/R+ discordant samples were generally close to the cut-off (weakly reactive).
(a) S-/R+ samples had S/CO values between 1.0 and 99.1. (The positive range of Roche assay is 1->1000); (b) S+/R- samples had S/CO values between 1.1 and 5.8 (the positive range of Siemens assay is 1->12).
Fig 4
Fig 4. Western blots of selected specimens: negative, positive and low positive controls are labelled N, P and LP respectively.
Specimens of interest are labelled A to G. Specimens A, B and C were concordantly weakly positive on both serology assays and HIV viral loads were high. Their western blots were scored as positive (A) and indeterminate (B, C). Specimens D, E, F and G were also WB indeterminate, but RNA negative. A follow up specimen on D was antibody negative.

Similar articles

Cited by

References

    1. Simbayi, L. C., Shisana, O., Rehle, T., Onoya, D., Jooste, S., Zungu, N., et al. South African National HIV Prevalence, Incidence and Behaviour Survey, 2012. 2014. Available from: http://www.hsrc.ac.za/en/research-data/view/6871 - PubMed
    1. Wolpaw BJ, Mathews C, Chopra M, Hardie D, de Azevedo V, Jennings K, et al. The failure of routine rapid HIV testing: a case study of improving low sensitivity in the field. BMC Health Serv Res [Internet]. 2010;10:73 Available from: http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=2851712&tool=p... doi: 10.1186/1472-6963-10-73 - DOI - PMC - PubMed
    1. Begg K, Young T, Stone P. Analysis of POCT / VCT performed at South African primary health care clinics. 2012;1–44. Avaiable from: http://www.sead.co.za/downloads/POCT-clinics-2011.pdf
    1. World Health Organization. Recommendations for the selection and use of HIV antibody tests. Wkly Epidemiol Rec. 1997;72(12):81–8. - PubMed
    1. Alexander TS. Human Immunodeficiency Virus Diagnostic Testing: 30 Years of Evolution. Clin Vaccine Immunol [Internet]. 2016;23(4):249–53. Available from: doi: http://cvi.asm.org/lookup/doi/10.1128/CVI.00053-16 - DOI - PMC - PubMed

Substances