Polymer coating embolism from intravascular medical devices - a clinical literature review
- PMID: 28759819
- DOI: 10.1016/j.carpath.2017.06.004
Polymer coating embolism from intravascular medical devices - a clinical literature review
Abstract
Over the past three decades, lubricious (hydrophobic and/or hydrophilic) polymer-coated devices have been increasingly adopted by interventional physicians and vascular surgeons to access and treat a wider range of clinical presentations. Recent clinical literature highlights the presence of polymer coating emboli within the anatomy - a result of coating separation from an intravascular device - and associates it with a range of adverse clinical sequelae. The 2015 U.S. Food and Drug Administration safety communication titled "Lubricious Coating Separation from Intravascular Medical Devices" acknowledges these concerns and concludes that it will work with stakeholders to develop nonclinical test methodologies, establish performance criteria, and identify gaps in current national and international device standards for coating integrity performance. Despite this communication and multiple case reports from interventional physicians, pathologists, dermatologists and other involved physician specialties, polymer coating embolism remains clinically underrecognized. This article consolidates the available literature on polymer coating embolism (1986-2016) and highlights the following relevant information for the physician: (a) the history and elusive nature of polymer coating embolism; (b) potential incidence rates of this phenomenon; (c) reported histologic findings and clinical effects of polymer emboli in the anatomy; (d) the importance of the collaborative clinician-pathologist partnership to report polymer embolism findings; and (e) the importance to study particulate release from intravascular devices so as to further understand and potentially evolve coated interventional technologies. Preliminary research on coatings highlights the potential of using iterations of coatings on medical devices that attain the desired therapeutic result and mitigate or eliminate particulates altogether.
Keywords: Catheter-based interventions; Complications; Particulate safety limits; Pathophysiology; Polymer coating embolism; Product quality improvement.
Copyright © 2017 The Authors. Published by Elsevier Inc. All rights reserved.
Similar articles
-
Analysis: Intravascular Devices with a Higher Risk of Polymer Emboli: The Need for Particulate Generation Testing.Biomed Instrum Technol. 2020 Jan-Feb;54(1):37-43. doi: 10.2345/0899-8205-54.1.37. Biomed Instrum Technol. 2020. PMID: 31961742 Free PMC article.
-
Analysis: Particulate Limits for Intravascular Devices: Considerations for Polymer Coating Embolism.Biomed Instrum Technol. 2019 Nov./Dec.;53(6):426-432. doi: 10.2345/0899-8205-53.6.426. Biomed Instrum Technol. 2019. PMID: 31765580
-
Cutaneous microemboli from hydrophilic polymer after endovascular procedures.J Am Acad Dermatol. 2015 Oct;73(4):666-71. doi: 10.1016/j.jaad.2015.07.014. Epub 2015 Aug 12. J Am Acad Dermatol. 2015. PMID: 26277212
-
Hydrophilic Polymer Embolism: Implications for Manufacturing, Regulation, and Postmarket Surveillance of Coated Intravascular Medical Devices.J Patient Saf. 2021 Dec 1;17(8):e1069-e1079. doi: 10.1097/PTS.0000000000000473. J Patient Saf. 2021. PMID: 29557931 Free PMC article. Review.
-
Myocardial hydrophilic polymer emboli following cardiac catheterization: a case report and literature review.Cardiovasc Pathol. 2014 May-Jun;23(3):175-7. doi: 10.1016/j.carpath.2014.01.009. Epub 2014 Feb 4. Cardiovasc Pathol. 2014. PMID: 24666841 Review.
Cited by
-
Thrombotic vasculopathy due to polymer coating emboli after an endovascular procedure.JAAD Case Rep. 2021 Apr 2;11:115-116. doi: 10.1016/j.jdcr.2021.03.034. eCollection 2021 May. JAAD Case Rep. 2021. PMID: 33948474 Free PMC article. No abstract available.
-
Hydrophilic Polymer Embolism: An Underappreciated Complication of Endovascular Procedures that Physicians Should be Aware Of.Cardiovasc Intervent Radiol. 2025 Jun 20. doi: 10.1007/s00270-025-04098-4. Online ahead of print. Cardiovasc Intervent Radiol. 2025. PMID: 40540040 Review.
-
Hydrophilic polymer embolization following flow diversion of cerebral aneurysms.Neuroradiol J. 2021 Aug;34(4):363-369. doi: 10.1177/19714009211004185. Epub 2021 Mar 26. Neuroradiol J. 2021. PMID: 33765885 Free PMC article. Review.
-
Delayed cerebral complications associated with endovascular catheter coil embolization for pulmonary venous shunts mimicking cerebral inflammatory diseases.Heliyon. 2024 Jun 24;10(13):e33541. doi: 10.1016/j.heliyon.2024.e33541. eCollection 2024 Jul 15. Heliyon. 2024. PMID: 39035507 Free PMC article.
-
Hydrophilic polymer coating delamination during neurointerventional treatment after microcatheter withdrawal: particulate identification through attenuated total reflection Fourier-transform infrared spectroscopy.Front Neurol. 2025 Jan 15;15:1479375. doi: 10.3389/fneur.2024.1479375. eCollection 2024. Front Neurol. 2025. PMID: 39882360 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous