Validating Reference Equations for Impulse Oscillometry in Healthy Mexican Children

Affiliations

Validating Reference Equations for Impulse Oscillometry in Healthy Mexican Children

Laura Gochicoa-Rangel et al. Respir Care. 2017 Sep.

Abstract

Background: The impulse oscillometry system (IOS) measures the impedance (Z) of the respiratory system, but proper interpretation of its results requires adequate reference values. The objectives of this work were: (1) to validate the reference equations for the IOS published previously by our group and (2) to compare the adjustment of new available reference equations for the IOS from different countries in a sample of healthy children.

Methods: Subjects were healthy 4-15-y-old children from the metropolitan area of Mexico City, who performed an IOS test. The functional IOS parameters obtained were compared with the predicted values from 12 reference equations determined in studies of different ethnic groups. The validation methods applied were: analysis of the differences between measured and predicted values for each reference equation; correlation and concordance coefficients; adjustment by Z-score values; percentage of predicted value; and the percentage of patients below the lower limit of normality or above the upper limit of normality.

Results: Of the 224 participants, 117 (52.3%) were girls, and the mean age was 8.6 ± 2.3 y. The equations that showed the best adjustment for the different parameters were those from the studies by Nowowiejska et al (2008) and Gochicoa et al (2015). The equations proposed by Frei et al (2005), Hellinckx et al (1998), Kalhoff et al (2011), Klug and Bisgaard (1998), de Assumpção et al (2016), and Dencker et al (2006) overestimated the airway resistance of the children in our sample, whereas the equation of Amra et al (2008) underestimated it. In the analysis of the lower and upper limits of normality, Gochicoa et al equation was the closest, since 5% of subjects were below or above percentiles 5 and 95, respectively. The study found that, in general, all of the equations showed greater error at the extremes of the age distribution.

Conclusions: Because of the robust adjustment of the present study reference equations for the IOS, it can be recommended for both clinical and research purposes in our population. The differential adjustment of other equations underlines the need to obtain local reference values.

Keywords: oscillometry; pulmonary function test; reference values.