Comparative Study
. 2017 Sep;18(5):364-368.
doi: 10.1002/acm2.12150. Epub 2017 Aug 2.

Technical note: No increase in effective dose from half compared to full rotation pelvis cone beam CT

Affiliations
Comparative Study

Technical note: No increase in effective dose from half compared to full rotation pelvis cone beam CT

Pascal Hauri et al. J Appl Clin Med Phys. 2017 Sep.

Abstract

Purpose: To image the abdomen of a patient with a gantry mounted imaging system of a linear accelerator, different cone beam computed tomography (CBCT) protocols are available. The whole-body dose of a full rotation abdomen CBCT and a half rotation CBCT was compared. In our clinic, both CBCT protocols are used in daily routine work.

Methods: With an adult anthropomorphic Alderson phantom, the whole-body dose per CBCT scan was measured with thermoluminescence dosimeters. The half rotation CBCT was applied such that the gantry mounted X-ray source rotated around the right side of the phantom. The 183 measurement locations covered all ICRP recommended critical organs (except the gonads). The effective dose was calculated with the mean organ dose and the corresponding tissue weighting factors. A point-by-point dose comparison of both protocols was conducted.

Results: The effective dose was 5.4 mSv ±5% and 5.0 mSv ±5% (estimated type B 1σ) for the full and the half rotation CBCT respectively. There was no significant difference (α = 0.05) in the effective dose within the precision of the measurement (1σ = 5%). The half rotation CBCT displayed an inhomogeneous dose distribution in a transversal phantom slice in contrast with the full rotation CBCT. In the imaging region, the mean dose was (20.5 ± 3.4) mGy and (19.2 ± 7.4) mGy (measured type A 1σ) for the full and the half rotation CBCT respectively.

Conclusion: The half compared to the full rotation CBCT displays a smaller field-of-view in a transversal slice and no significant difference in the effective dose. Hence, the full rotation CBCT is favorable compared to the half rotation CBCT. However, by using the half rotation protocol, critical volumes in the patient can be spared compared to the full rotation protocol.

Keywords: Cone Beam CT; abdomen; effective dose; radiation therapy.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Measured point doses per CBCT scan in the Alderson phantom. Each cross corresponds to a unique measurement location for the pelvis and the pelvis spotlight protocol. For all points below the 45° line the dose of the pelvis spotlight was lower compared to the pelvis protocol.
Figure 2
Figure 2
The measured point doses (mGy) per scan in a transversal slice of the Alderson phantom for (a) the pelvis CBCT and (b) the pelvis spotlight CBCT. The arrow in the right upper corner of the CBCT slice indicates the rotation of X‐ray source around the phantom. The cross in the milled of the slice represents the image isocenter. (c) The dose profiles along the RL axis of the full and the half rotation CBCT for the same transversal slice from (a) and (b). The error bars represent the standard deviation of the dose measurements.

References

    1. Murphy MJ, Balter J, Balter S, et al. The management of imaging dose during image‐guided radiotherapy: report of the AAPM Task Group 75. Med Phys. 2007;34:4041–4063. - PubMed
    1. Montanari D, Scolari E, Silvestri C, et al. Comprehensive evaluations of cone‐beam CT dose in image‐guided radiation therapy via GPU‐based Monte Carlo simulations. Phys Med Biol. 2014;59:1239–1253. - PubMed
    1. Kim S, Alaei P. Implementation of full/half bowtie filter models in a commercial treatment planning system for kilovoltage cone‐beam CT dose estimations. J Appl Clin Med Phys. 2016;17:5988. - PMC - PubMed
    1. Gardner SJ, Studenski MT, Giaddui TK, et al. Investigation into image quality and dose for different patient geometries with multiple cone‐beam CT systems. Med Phys. 2014;41:031908. - PMC - PubMed
    1. Poirier Y, Tambasco M. Experimental validation of a kV source model and dose computation method for CBCT imaging in an anthropomorphic phantom. J Appl Clin Med Phys. 2016;17:6021. - PMC - PubMed