Investigating steam penetration using thermometric methods in dental handpieces with narrow internal lumens during sterilizing processes with non-vacuum or vacuum processes
- PMID: 28778810
- DOI: 10.1016/j.jhin.2017.07.033
Investigating steam penetration using thermometric methods in dental handpieces with narrow internal lumens during sterilizing processes with non-vacuum or vacuum processes
Abstract
Background: Dental handpieces are required to be sterilized between patient use. Vacuum steam sterilization processes with fractionated pre/post-vacuum phases or unique cycles for specified medical devices are required for hollow instruments with internal lumens to assure successful air removal. Entrapped air will compromise achievement of required sterilization conditions. Many countries and professional organizations still advocate non-vacuum sterilization processes for these devices.
Aim: To investigate non-vacuum downward/gravity displacement, type-N steam sterilization of dental handpieces, using thermometric methods to measure time to achieve sterilization temperature at different handpiece locations.
Methods: Measurements at different positions within air turbines were undertaken with thermocouples and data loggers. Two examples of widely used UK benchtop steam sterilizers were tested: a non-vacuum benchtop sterilizer (Little Sister 3; Eschmann, Lancing, UK) and a vacuum benchtop sterilizer (Lisa; W&H, Bürmoos, Austria). Each sterilizer cycle was completed with three handpieces and each cycle in triplicate.
Findings: A total of 140 measurements inside dental handpiece lumens were recorded. The non-vacuum process failed (time range: 0-150 s) to reliably achieve sterilization temperatures within the time limit specified by the international standard (15 s equilibration time). The measurement point at the base of the handpiece failed in all test runs (N = 9) to meet the standard. No failures were detected with the vacuum steam sterilization type B process with fractionated pre-vacuum and post-vacuum phases.
Conclusion: Non-vacuum downward/gravity displacement, type-N steam sterilization processes are unreliable in achieving sterilization conditions inside dental handpieces, and the base of the handpiece is the site most likely to fail.
Keywords: Dental handpieces; Dental infection control; Non-vacuum sterilizers; Steam sterilization; Thermocouples; Vacuum sterilizers.
Copyright © 2017 The Healthcare Infection Society. Published by Elsevier Ltd. All rights reserved.
Similar articles
-
Failure of non-vacuum steam sterilization processes for dental handpieces.J Hosp Infect. 2017 Dec;97(4):343-347. doi: 10.1016/j.jhin.2017.09.004. Epub 2017 Sep 10. J Hosp Infect. 2017. PMID: 28903057
-
[Sterilization effect analysis of B-class pulsation table top vacuum sterilizer to dental handpieces].Hua Xi Kou Qiang Yi Xue Za Zhi. 2007 Jun;25(3):269-71. Hua Xi Kou Qiang Yi Xue Za Zhi. 2007. PMID: 17629204 Chinese.
-
Steam sterilization of air turbine dental handpieces.Acta Odontol Scand. 1983 Dec;41(6):321-6. doi: 10.3109/00016358309162342. Acta Odontol Scand. 1983. PMID: 6362319
-
Autoclave sterilization of dental handpieces: A literature review.J Prosthodont Res. 2020 Jul;64(3):239-242. doi: 10.1016/j.jpor.2019.07.013. Epub 2019 Aug 12. J Prosthodont Res. 2020. PMID: 31416709 Review.
-
A Tale of Two Sterilizers.PDA J Pharm Sci Technol. 2020 Jan-Feb;74(1):162-169. doi: 10.5731/pdajpst.2019.009993. Epub 2019 Feb 15. PDA J Pharm Sci Technol. 2020. PMID: 30770480 Review.
Cited by
-
Effectiveness of steam sterilization of reusable medical devices in primary and secondary care public hospitals in Nepal and factors associated with ineffective sterilization: A nation-wide cross-sectional study.PLoS One. 2019 Nov 21;14(11):e0225595. doi: 10.1371/journal.pone.0225595. eCollection 2019. PLoS One. 2019. PMID: 31751421 Free PMC article.
-
Air-source heat pump heating based water vapor compression for localized steam sterilization applications during the COVID-19 pandemic.Renew Sustain Energy Rev. 2021 Jul;145:111026. doi: 10.1016/j.rser.2021.111026. Epub 2021 Apr 16. Renew Sustain Energy Rev. 2021. PMID: 36569373 Free PMC article.
-
Microbiological-Chemical Sourced Chondroitin Sulfates Protect Neuroblastoma SH-SY5Y Cells against Oxidative Stress and Are Suitable for Hydrogel-Based Controlled Release.Antioxidants (Basel). 2021 Nov 16;10(11):1816. doi: 10.3390/antiox10111816. Antioxidants (Basel). 2021. PMID: 34829687 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources