. 2017 Aug;14(2):1626-1632.
doi: 10.3892/etm.2017.4675. Epub 2017 Jun 26.

Application of Corvis ST to evaluate the effect of femtosecond laser-assisted cataract surgery on corneal biomechanics

Affiliations

Application of Corvis ST to evaluate the effect of femtosecond laser-assisted cataract surgery on corneal biomechanics

Yinjuan Wei et al. Exp Ther Med. 2017 Aug.

Abstract

The aim of the present study was to evaluate the effects of femtosecond laser-assisted cataract surgery (FLACS) and phacoemulsification on corneal biomechanics using corneal visualization Scheimpflug technology. The medical records of 50 eyes from 50 patients who received phacoemulsification and intraocular lens implantation because of age-related factors between June 2014 and September 2014 were retrospectively analyzed. FLACS was used in 12 eyes (FLACS group), and conventional phacoemulsification in 38 eyes (PHACO group). The evaluation of corneal biomechanical parameters included the first/second applanation time (A-time1/A-time2), the first/second applanation length (A-length1/A-length2), corneal velocity during the first/second applanation moment (Vin/Vout), highest concavity time, highest concavity-radius (HC-radius), peak distance (PD), deformation amplitude (DA), central corneal thickness (CCT), and intraocular pressure (IOP). The differences in A-length1/A-length2, IOP, CCT, PD, and DA were significant in the PHACO group between those before, 1 week after, and 1 month after surgery. No significant differences in corneal biomechanical parameters were found between those at 1 month after surgery and before surgery. There were significant differences in IOP and CCT in the FLACS group between those before, 1 week after, and 1 month after surgery. There were no significant differences in the other corneal biomechanical parameters. No significant differences were found in corneal biomechanical parameters between those 1 month after surgery and before surgery. There were significant differences in A-length1/A-length2, CCT, PD, and DA between the two groups at 1 week after surgery. There were no significant differences in corneal biomechanical parameters between the two groups at 1 month after surgery. In conclusion, the effect of FLACS on corneal biomechanics is smaller than that of phacoemulsification. The corneal biomechanical parameters are restored to preoperative levels with the healing of the incision, and the reduction of swelling of the tissue near the incision.

Keywords: corneal biomechanics; femtosecond laser; phacoemulsification.

PubMed Disclaimer

Figures

Figure 1.
Figure 1.
The entire process of corneal deformation by airflow used in Corvis ST. Corvis ST, corneal visualization Scheimpflug technology; A-length1/Alength2, first/second applanation length.
Figure 2.
Figure 2.
There were significant differences in A-length1/A-length2, CCT, PD, and DA between the two groups at 1 week after surgery (P=0.041, P=0.025, P=0.016, P=0.025, P=0.047). A-length1/A-length2, first/second applanation length; CCT, central corneal thickness; PD, peak distance; DA, deformation amplitude.

Similar articles

Cited by

References

    1. Nagy Z, Takacs A, Filkorn T, Sarayba M. Initial clinical evaluation of an intraocular femtosecond laser in cataract surgery. J Refract Surg. 2009;25:1053–1060. doi: 10.3928/1081597X-20091117-04. - DOI - PubMed
    1. Masket S, Sarayba M, Ignacio T, Fram N. Femtosecond laser-assisted cataract incisions: architectural stability and reproducibility. J Cataract Refract Surg. 2010;36:1048–1049. doi: 10.1016/j.jcrs.2010.03.027. - DOI - PubMed
    1. Hon Y, Lam AK. Corneal deformation measurement using Scheimpflug noncontact tonometry. Optom Vis Sci. 2013;90:e1–e8. doi: 10.1097/OPX.0b013e318279eb87. - DOI - PubMed
    1. Hayes S, Boote C, Lewis J, Sheppard J, Abahussin M, Quantock AJ, Purslow C, Votruba M, Meek KM. Comparative study of fibrillar collagen arrangement in the corneas of primates and other mammals. Anat Rec (Hoboken) 2007;290:1542–1550. doi: 10.1002/ar.20613. - DOI - PubMed
    1. Boote C, Hayes S, Abahussin M, Meek KM. Mapping collagen organization in the human cornea: left and right eyes are structurally distinct. Invest Ophthalmol Vis Sci. 2006;47:901–908. doi: 10.1167/iovs.05-0893. - DOI - PubMed