Pencil Beam Algorithms Are Unsuitable for Proton Dose Calculations in Lung

Affiliations
Comparative Study

Pencil Beam Algorithms Are Unsuitable for Proton Dose Calculations in Lung

Paige A Taylor et al. Int J Radiat Oncol Biol Phys. .

Abstract

Purpose: To compare analytic and Monte Carlo-based algorithms for proton dose calculations in the lung, benchmarked against anthropomorphic lung phantom measurements.

Methods and materials: A heterogeneous anthropomorphic moving lung phantom has been irradiated at numerous proton therapy centers. At 5 centers the treatment plan could be calculated with both an analytic and Monte Carlo algorithm. The doses calculated in the treatment plans were compared with the doses delivered to the phantoms, which were measured using thermoluminescent dosimeters and film. Point doses were compared, as were planar doses using a gamma analysis.

Results: The analytic algorithms overestimated the dose to the center of the target by an average of 7.2%, whereas the Monte Carlo algorithms were within 1.6% of the physical measurements on average. In some regions of the target volume, the analytic algorithm calculations differed from the measurement by up to 31% in the internal gross target volume (iGTV) (46% in the planning target volume), over-predicting the dose. All comparisons showed a region of at least 15% dose discrepancy within the iGTV between the analytic calculation and the measured dose. The Monte Carlo algorithm recalculations showed dramatically improved agreement with the measured doses, showing mean agreement within 4% for all cases and a maximum difference of 12% within the iGTV.

Conclusions: Analytic algorithms often do a poor job predicting proton dose in lung tumors, over-predicting the dose to the target by up to 46%, and should not be used unless extensive validation counters the consistent results of the present study. Monte Carlo algorithms showed dramatically improved agreement with physical measurements and should be implemented to better reflect actual delivered dose distributions.

PubMed Disclaimer

Conflict of interest statement

Conflicts of interest: The authors have no potential conflicts of interest to disclose.

Figures

Figure 1
Figure 1
The anthropomorphic left lung phantom used for proton therapy quality assurance.
Figure 2
Figure 2
Axial and sagittal view of the proton lung phantom with gantry angle orientation indicated. A: Anterior; L: Left; P: Posterior; R: Right; Sup: Superior; Inf: Inferior.
Figure 3
Figure 3
Dose (relative biological effectiveness [RBE]) profile through the center of the planning target volume (PTV) in the left-right, posterior-anterior, and inferior-superior directions for institution 2 in the sample study. Film measurements are shown in blue, the analytic pencil beam algorithm in green, and the Monte Carlo recalculation in red.
Figure 4
Figure 4
Dose (relative biological effectiveness [RBE]) profiles through the center of the planning target volume (PTV) in the left-right direction of the film measurements (blue), analytic pencil beam (PB) algorithm (green), and Monte Carlo (MC) recalculation (red). See figure 2 for Institution 2 profile.

References

    1. Bush DA, Cheek G, Zaheer S, et al. High-dose hypofractionated proton beam radiation therapy is safe and effective for central and peripheral early-stage non-small cell lung cancer: results of a 12-year experience at Loma Linda University Medical Center. Int J Radiat Oncol Biol Phys. 2013;86:964–968. - PubMed
    1. Nguyen QN, Ly NB, Komaki R, et al. Long-term outcomes after proton therapy, with concurrent chemotherapy, for stage II–III inoperable non-small cell lung cancer. Radiother Oncol. 2015;115:367–372. - PMC - PubMed
    1. Hoppe BS, Henderson R, Pham D, et al. A Phase 2 Trial of Concurrent Chemotherapy and Proton Therapy for Stage III Non-Small Cell Lung Cancer: Results and Reflections Following Early Closure of a Single-Institution Study. Int J Radiat Oncol Biol Phys. 2016;95:517–522. - PubMed
    1. Institute NC; Center IHQ, editor. Guidelines for the Use of Proton Radiation Therapy in NCI-Sponsored Cooperative Group Clinical Trials. 2012 irochouston.mdanderson.org.
    1. Taylor PA, Kry SF, Alvarez P, et al. Results From the Imaging and Radiation Oncolocy Core Houston’s Anthropomorphic Phantoms Used for Proton Therapy Clinical Trial Credentialing. International Journal of Radiation Oncology Biology Physics. 2016;95:7. - PMC - PubMed

Publication types

LinkOut - more resources