Case Reports
. 2017 Jun;17(2):135-138.
doi: 10.17245/jdapm.2017.17.2.135. Epub 2017 Jun 29.

Airway obstruction by dissection of the inner layer of a reinforced endotracheal tube in a patient with Ludwig's angina: A case report

Affiliations
Case Reports

Airway obstruction by dissection of the inner layer of a reinforced endotracheal tube in a patient with Ludwig's angina: A case report

Sung-Min Shim et al. J Dent Anesth Pain Med. 2017 Jun.

Abstract

Intraoperative airway obstruction is perplexing to anesthesiologists because the patient may fall into danger rapidly. A 74-year-old woman underwent an emergency incision and drainage for a deep neck infection of dental origin. She was orally intubated with a 6. 0 mm internal diameter reinforced endotracheal tube by video laryngoscope using volatile induction and maintenance anesthesia (VIMA) with sevoflurane, fentanyl (100 µg), and succinylcholine (75 mg). During surgery, peak inspiratory pressure increased from 22 to 38 cmH2O and plateau pressure increased from 20 to 28 cmH2O. We maintained anesthesia because we were unable to access the airway, which was covered with surgical drapes, and tidal volume was delivered. At the end of surgery, we found a longitudinal fold inside the tube with a fiberoptic bronchoscope. The patient was reintubated with another tube and ventilation immediately improved. We recognized that the tube was obstructed due to dissection of the inner layer.

Keywords: Bronchoscopes; Intubation; Ludwig's Angina; Reinforced Endotracheal Tube.

PubMed Disclaimer

Conflict of interest statement

There are no financial or other issues that might lead to conflict of interest.

Figures

Fig. 1
Fig. 1. Fiberoptic bronchoscopic view of the inside of the endotracheal tube. A longitudinal fold is seen inside the tube.
Fig. 2
Fig. 2. Video laryngoscopic view for reintubation. A new endotracheal tube was prepared at the bedside for entrance of vocal cord.
Fig. 3
Fig. 3. Cut image of the dissected tube. The inner layer was detached from the outer layer and coil.

References

    1. Szekely S, Webb R, Williamson J, Russell W. The Australian Incident Monitoring Study. Problems related to the endotracheal tube: an analysis of 2000 incident reports. Anaesth Intensive Care. 1993;21:611–616. - PubMed
    1. Jeon Y, Kim Y, Joo J, Kang E, In J, Choi J, et al. Partial airway obstruction caused by dissection of a reinforced endotracheal tube. Eur J Anaesthesiol. 2007;24:983–984. - PubMed
    1. Mercanoglu E, Topuz D, Kaya N. The dissection of reinforced endotracheal tube internal wall causing intraoperative airway obstruction under general anesthesia. Case report. Braz J Anesthesiol. 2013;63:372–374. - PubMed
    1. Tose R, Kubota T, Hirota K, Sakai T, Ishihara H, Matsuki A. Obstruction of an reinforced endotracheal tube due to dissection of internal tube wall during total intravenous anesthesia. Masui. 2003;52:1218–1220. - PubMed
    1. Paul M, Dueck M, Kampe S, Petzke F. Failure to detect an unusual obstruction in a reinforced endotracheal tube with fiberoptic examination. Anesth Analg. 2003;97:909–910. - PubMed

Publication types