Biomechanical Comparison of Inter-fragmentary Compression Pressures: Lag Screw versus Herbert Screw for Anterior Odontoid Screw Fixation

Affiliations

Biomechanical Comparison of Inter-fragmentary Compression Pressures: Lag Screw versus Herbert Screw for Anterior Odontoid Screw Fixation

Jin-Woo Park et al. J Korean Neurosurg Soc. 2017 Sep.

Abstract

Objective: The purpose of the present study was to compare inter-fragmentary compression pressures after fixation of a simulated type II odontoid fracture with the headless compression Herbert screw and a half threaded cannulated lag screw.

Methods: We compared inter-fragmentary compression pressures between 40- and 45-mm long 4.5-mm Herbert screws (n=8 and n=9, respectively) and 40- and 45-mm long 4.0-mm cannulated lag screws (n=7 and n=10, respectively) after insertion into rigid polyurethane foam test blocks (Sawbones, Vashon, WA, USA). A washer load cell was placed between the two segments of test blocks to measure the compression force. Because the total length of each foam block was 42 mm, the 40-mm screws were embedded in the cancellous foam, while the 45-mm screws penetrated the denser cortical foam at the bottom. This enabled us to compare inter-fragmentary compression pressures as they are affected by the penetration of the apical dens tip by the screws.

Results: The mean compression pressures of the 40- and 45-mm long cannulated lag screws were 50.48±1.20 N and 53.88±1.02 N, respectively, which was not statistically significant (p=0.0551). The mean compression pressures of the 40-mm long Herbert screw was 52.82±2.17 N, and was not statistically significant compared with the 40-mm long cannulated lag screw (p=0.3679). However, 45-mm Herbert screw had significantly higher mean compression pressure (60.68±2.03 N) than both the 45-mm cannulated lag screw and the 40-mm Herbert screw (p=0.0049 and p=0.0246, respectively).

Conclusion: Our results showed that inter-fragmentary compression pressures of the Herbert screw were significantly increased when the screw tip penetrated the opposite dens cortical foam. This can support the generally recommended surgical technique that, in order to facilitate maximal reduction of the fracture gap using anterior odontoid screws, it is essential to penetrate the apical dens tip with the screw.

Keywords: Fracture gap; Herbert screw; Inter-fragmentary compression; Lag screw; Odontoid fracture; Sawbones.

PubMed Disclaimer

Figures

Fig. 1
Fig. 1
Showing the studied implants; the 4.0-mm cannulated cortical screw (left) and the 4.5-mm headless compression Herbert screw (right).
Fig. 2
Fig. 2
Illustration for test setup. PCF: per cubic foot.
Fig. 3
Fig. 3
Showing the test specimens which was mounted onto a Testometric M 500 Universal testing machine.
Fig. 4
Fig. 4
Mean compression pressures of cannulated lag screws and headless compression Herbert screws (n=7–10/groups).

References

    1. Adla DN, Kitsis C, Miles AW. Compression forces generated by Mini bone screws--a comparative study done on bone model. Injury. 2005;36:65–70. - PubMed
    1. Andersson S, Rodrigues M, Olerud C. Odontoid fractures: high complication rate associated with anterior screw fixation in the elderly. Eur Spine J. 2000;9:56–59. - PMC - PubMed
    1. Apfelbaum RI, Lonser RR, Veres R, Casey A. Direct anterior screw fixation for recent and remote odontoid fractures. J Neurosurg. 2000;93(2 Suppl):227–236. - PubMed
    1. Assari S, Darvish K, Ilyas AM. Biomechanical analysis of second-generation headless compression screws. Injury. 2012;43:1159–1165. - PubMed
    1. Beadel GP, Ferreira L, Johnson JA, King GJ. Interfragmentary compression across a simulated scaphoid fracture--analysis of 3 screws. J Hand Surg Am. 2004;29:273–278. - PubMed

LinkOut - more resources