Kinematically aligned total knee arthroplasty limits high tibial forces, differences in tibial forces between compartments, and abnormal tibial contact kinematics during passive flexion

Affiliations

Kinematically aligned total knee arthroplasty limits high tibial forces, differences in tibial forces between compartments, and abnormal tibial contact kinematics during passive flexion

Joshua D Roth et al. Knee Surg Sports Traumatol Arthrosc. 2018 Jun.

Abstract

Purpose: Following total knee arthroplasty (TKA), high tibial forces, large differences in tibial forces between the medial and lateral compartments, and anterior translation of the contact locations of the femoral component on the tibial component during passive flexion indicate abnormal knee function. Because the goal of kinematically aligned TKA is to restore native knee function without soft tissue release, the objectives were to determine how well kinematically aligned TKA limits high tibial forces, differences in tibial forces between compartments, and anterior translation of the contact locations of the femoral component on the tibial component during passive flexion.

Methods: Using cruciate retaining components, kinematically aligned TKA was performed on thirteen human cadaveric knee specimens with use of manual instruments without soft tissue release. The tibial forces and tibial contact locations were measured in both the medial and lateral compartments from 0° to 120° of passive flexion using a custom tibial force sensor.

Results: The average total tibial force (i.e. sum of medial + lateral) ranged from 5 to 116 N. The only significant average differences in tibial force between compartments occurred at 0° of flexion (29 N, p = 0.0008). The contact locations in both compartments translated posteriorly in all thirteen kinematically aligned TKAs by an average of 14 mm (p < 0.0001) and 18 mm (p < 0.0001) in the medial and lateral compartments, respectively, from 0° to 120° of flexion.

Conclusions: After kinematically aligned TKA, average total tibial forces due to the soft tissue restraints were limited to 116 N, average differences in tibial forces between compartments were limited to 29 N, and a net posterior translation of the tibial contact locations was observed in all kinematically aligned TKAs during passive flexion from 0° to 120°, which are similar to what has been measured previously in native knees. While confirmation in vivo is warranted, these findings give surgeons who perform kinematically aligned TKA confidence that the alignment method and surgical technique limit high tibial forces, differences in tibial forces between compartments, and anterior translation of the tibial contact locations during passive flexion.

Keywords: Anterior sliding; Contact force imbalance; Intra-articular load; Kinematic alignment; Soft tissue balancing; Tibial force sensor; Tibiofemoral joint.

PubMed Disclaimer

Similar articles

Cited by

References

    1. J Arthroplasty. 2016 Oct;31(10 ):2181-7 - PubMed
    1. J Biomech. 2017 Jun 14;58:195-202 - PubMed
    1. Knee Surg Sports Traumatol Arthrosc. 2013 Oct;21(10 ):2271-80 - PubMed
    1. J Arthroplasty. 2016 Sep;31(9 Suppl):102-5 - PubMed
    1. J Biomech. 2000 Apr;33(4):465-73 - PubMed

LinkOut - more resources