Reduction of artifacts caused by orthopedic hardware in the spine in spectral detector CT examinations using virtual monoenergetic image reconstructions and metal-artifact-reduction algorithms
- PMID: 28932962
- DOI: 10.1007/s00256-017-2776-5
Reduction of artifacts caused by orthopedic hardware in the spine in spectral detector CT examinations using virtual monoenergetic image reconstructions and metal-artifact-reduction algorithms
Abstract
Objective: Aim of this study was to assess the artifact reduction in patients with orthopedic hardware in the spine as provided by (1) metal-artifact-reduction algorithms (O-MAR) and (2) virtual monoenergetic images (MonoE) as provided by spectral detector CT (SDCT) compared to conventional iterative reconstruction (CI).
Methods: In all, 28 consecutive patients with orthopedic hardware in the spine who underwent SDCT-examinations were included. CI, O-MAR and MonoE (40-200 keV) images were reconstructed. Attenuation (HU) and noise (SD) were measured in order to calculate signal-to-noise ratio (SNR) of paravertebral muscle and spinal canal. Subjective image quality was assessed by two radiologists in terms of image quality and extent of artifact reduction.
Results: O-MAR and high-keV MonoE showed significant decrease of hypodense artifacts in terms of higher attenuation as compared to CI (CI vs O-MAR, 200 keV MonoE: -396.5HU vs. -115.2HU, -48.1HU; both p ≤ 0.001). Further, artifacts as depicted by noise were reduced in O-MAR and high-keV MonoE as compared to CI in (1) paravertebral muscle and (2) spinal canal-CI vs. O-MAR/200 keV: (1) 34.7 ± 19.0 HU vs. 26.4 ± 14.4 HU, p ≤ 0.05/27.4 ± 16.1, n.s.; (2) 103.4 ± 61.3 HU vs. 72.6 ± 62.6 HU/60.9 ± 40.1 HU, both p ≤ 0.001. Subjectively both O-MAR and high-keV images yielded an artifact reduction in up to 24/28 patients.
Conclusion: Both, O-MAR and high-keV MonoE reconstructions as provided by SDCT lead to objective and subjective artifact reduction, thus the combination of O-MAR and MonoE seems promising for further reduction.
Keywords: Metal artifact reduction; Orthopedic hardware; Spectral detector computed tomography; Spinal instrumentation; Virtual monoenergetic images.
Similar articles
-
Reducing artifacts from total hip replacements in dual layer detector CT: Combination of virtual monoenergetic images and orthopedic metal artifact reduction.Eur J Radiol. 2019 Feb;111:14-20. doi: 10.1016/j.ejrad.2018.12.008. Epub 2018 Dec 7. Eur J Radiol. 2019. PMID: 30691660
-
Assessment of 70-keV virtual monoenergetic spectral images in abdominal CT imaging: A comparison study to conventional polychromatic 120-kVp images.Abdom Radiol (NY). 2017 Oct;42(10):2579-2586. doi: 10.1007/s00261-017-1151-2. Abdom Radiol (NY). 2017. PMID: 28421243
-
Reduction of artifacts from contrast media in spectral-detector CT by combined use of virtual monoenergetic images and orthopedic metal-artifact-reduction algorithm.Medicine (Baltimore). 2025 May 16;104(20):e42370. doi: 10.1097/MD.0000000000042370. Medicine (Baltimore). 2025. PMID: 40388764 Free PMC article.
-
CT artifacts from port systems: Virtual monoenergetic reconstructions from spectral-detector CT reduce artifacts and improve depiction of surrounding tissue.Eur J Radiol. 2019 Dec;121:108733. doi: 10.1016/j.ejrad.2019.108733. Epub 2019 Nov 6. Eur J Radiol. 2019. PMID: 31739270 Review.
-
Current and Novel Techniques for Metal Artifact Reduction at CT: Practical Guide for Radiologists.Radiographics. 2018 Mar-Apr;38(2):450-461. doi: 10.1148/rg.2018170102. Radiographics. 2018. PMID: 29528826 Review.
Cited by
-
Reduction of Metal Artifacts Caused by Titanium Peduncular Screws in the Spine by Means of Monoenergetic Images and the Metal Artifact Reduction Software in Dual-Energy Computed Tomography.J Med Phys. 2022 Apr-Jun;47(2):152-158. doi: 10.4103/jmp.jmp_121_21. Epub 2022 Aug 5. J Med Phys. 2022. PMID: 36212200 Free PMC article.
-
Computed Tomography- (CT-) Based Virtual Surgery Planning for Spinal Intervertebral Foraminal Assisted Clinical Treatment.J Healthc Eng. 2021 Mar 6;2021:5521916. doi: 10.1155/2021/5521916. eCollection 2021. J Healthc Eng. 2021. Retraction in: J Healthc Eng. 2023 Oct 11;2023:9834843. doi: 10.1155/2023/9834843. PMID: 33747415 Free PMC article. Retracted.
-
Photon-Counting Detector CT Virtual Monoenergetic Images in Cervical Trauma Imaging-Optimization of Dental Metal Artifacts and Image Quality.Diagnostics (Basel). 2024 Mar 15;14(6):626. doi: 10.3390/diagnostics14060626. Diagnostics (Basel). 2024. PMID: 38535045 Free PMC article.
-
Comparison of Metal Artifact Reduction Algorithms in Patients with Hip Prostheses: Virtual Monoenergetic Images vs. Orthopedic Metal Artifact Reduction.J Korean Soc Radiol. 2022 Nov;83(6):1286-1297. doi: 10.3348/jksr.2021.0130. Epub 2022 Apr 20. J Korean Soc Radiol. 2022. PMID: 36545406 Free PMC article.
-
Optimal combination periprosthetic vasculature visualization and metal artifact reduction by spectral computed tomography using virtual monoenergetic images in total hip arthroplasty.Insights Imaging. 2023 Oct 26;14(1):181. doi: 10.1186/s13244-023-01533-3. Insights Imaging. 2023. PMID: 37880460 Free PMC article.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous