Pain Free 3 T MRI Scans in Cochlear Implantees
- PMID: 28938274
- DOI: 10.1097/MAO.0000000000001569
Pain Free 3 T MRI Scans in Cochlear Implantees
Abstract
Objective: For cochlear implant recipients, undergoing magnetic resonance imaging (MRI) scans is associated with safety risks and potential side effects. Even following safety guidelines, potential complications (e.g., pain, magnet dislocation, image artifacts) are possible during 1.5 Tesla (T) MRI scans. The stronger static magnetic field of a 3.0 T scanner is associated with further risks of complication, including implant demagnetization. These complications led to the recent development of rotatable internal receiver magnets with a diametrical magnetization.The aim of this study was to evaluate the potential occurrence of pain during 3.0 T MRI scans for cochlear implant recipients with a rotatable, diametrically magnetized implant magnet.
Patients: Five patients implanted with a cochlear implant diametrically magnetized magnet.
Intervention: MRI scanning at 3 T.
Main outcome measure: In the prospective patient study an MRI scan was performed on five implantees and the degree of pain was evaluated by a visual analog scale. Scans were performed initially with a magnet-supporting headband, and depending on the degree of discomfort/pain, repeated without the headband.
Results: In all the patients, all the MRI scans were performed without any pain, even without the use of the supportive headband. Demagnetization was clinically not observed.
Conclusion: 3.0 T MRI scanning can be performed on cochlear implant recipients with a rotatable diametrically magnetized internal magnet without risk of the most frequent cochlear-implant-related MRI complication: pain. This finding enables the expansion of MRI scanning indications up to 3.0 T without complication. Limitations in terms of MRI artifact still persist.
Similar articles
-
Cochlear implants and 1.5 T MRI scans: the effect of diametrically bipolar magnets and screw fixation on pain.J Otolaryngol Head Neck Surg. 2018 Feb 5;47(1):11. doi: 10.1186/s40463-017-0252-9. J Otolaryngol Head Neck Surg. 2018. PMID: 29402322 Free PMC article.
-
Intraindividual Comparison of Image Artifacts of Two Generations of Rotatable Cochlear Implant Magnets.J Int Adv Otol. 2024 Jul 29;20(4):306-311. doi: 10.5152/iao.2024.231466. J Int Adv Otol. 2024. PMID: 39161162 Free PMC article.
-
First MRI With New Cochlear Implant With Rotatable Internal Magnet System and Proposal for Standardization of Reporting Magnet-Related Artifact Size.Otol Neurotol. 2019 Aug;40(7):883-891. doi: 10.1097/MAO.0000000000002269. Otol Neurotol. 2019. PMID: 31219967 Review.
-
A Prospective Study of Pain From Magnetic Resonance Imaging With Cochlear Implant Magnets In Situ.Otol Neurotol. 2018 Feb;39(2):e80-e86. doi: 10.1097/MAO.0000000000001661. Otol Neurotol. 2018. PMID: 29210950
-
Magnetic resonance imaging compatibility and safety of the SOUNDTEC Direct System.Laryngoscope. 2006 Aug;116(8):1321-33. doi: 10.1097/01.mlg.0000230479.39551.4a. Laryngoscope. 2006. PMID: 16885731 Review.
Cited by
-
Risk of magnetic resonance imaging-induced magnet dislocation for different types of cochlear implants: a single-center retrospective study.J Otolaryngol Head Neck Surg. 2023 Apr 21;52(1):28. doi: 10.1186/s40463-023-00633-w. J Otolaryngol Head Neck Surg. 2023. PMID: 37085925 Free PMC article.
-
Magnetic Resonance Imaging with Active Implantable Hearing Devices: Reports from the Daily Radiological Routine in an Outpatient MR Center.J Pers Med. 2023 Jul 31;13(8):1220. doi: 10.3390/jpm13081220. J Pers Med. 2023. PMID: 37623470 Free PMC article.
-
Hearing rehabilitation after subtotal cochleoectomy using a new, perimodiolar malleable cochlear implant electrode array: a preliminary report.Eur Arch Otorhinolaryngol. 2021 Feb;278(2):353-362. doi: 10.1007/s00405-020-06098-1. Epub 2020 Jun 5. Eur Arch Otorhinolaryngol. 2021. PMID: 32504199 Free PMC article.
-
Quality Control after Intracochlear Intralabyrinthine Schwannoma Resection and Cochlear Implantation.Brain Sci. 2021 Sep 16;11(9):1221. doi: 10.3390/brainsci11091221. Brain Sci. 2021. PMID: 34573240 Free PMC article.
-
Dimensions and forms of artefacts in 1.5 T and 3 T MRI caused by cochlear implants.Sci Rep. 2022 Mar 22;12(1):4884. doi: 10.1038/s41598-022-08988-2. Sci Rep. 2022. PMID: 35318407 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical