Evaluating doses of multi-slice CT in brain examinations using various methods
- PMID: 28938858
- DOI: 10.1080/24699322.2017.1379136
Evaluating doses of multi-slice CT in brain examinations using various methods
Abstract
Purpose: The effective dose (HE) and organ or tissue equivalent dose (HT) of a Rando phantom undergoing two brain computed tomography (CT) examination protocols were evaluated using thermoluminescent dosimeters (TLD-100H) and dose length product (DLP) methods.
Materials and methods: TLDs were inserted into the correlated positions of an organ or tissue of Rando phantom, such as thyroid, brain, and salivary gland, using (A) axial scan: scanning the maxillae ranging from external auditory meatus to the parietal bone, and (B) helical scan: scanning from the mandible to the parietal bone. CT examinations were performed on a Philips computer tomography (Brilliance CT) at Lukang Christian Hospital. TLDs were measured using a Harshaw 3500 TLD reader. The HT of organ and tissue during the two protocols was discussed.
Results and conclusion: HE were calculated using ICRP 60 and 103 at 2.67 ± 0.18 and 1.89 ± 0.23 mSv based on an axial scan, and 4.70 ± 0.38 and 4.39 ± 0.37 mSv based on a helical scan, respectively. In the DLP method, HE was estimated from CTDIvol that was recorded directly from the console display of the CT unit and then calculated using AAPM 96. Finally, experimental results are compared with those in literature. Radiologists should choose and adjust protocols to prevent unnecessary radiation to patients and satisfying the as low as reasonably achievable (ALARA) principle. These findings will be valuable to patients, physicians, radiologists, and the public.
Keywords: Thermoluminescent dosimeters (TLD); axial scan; computed tomography (CT); helical scan.
Similar articles
-
Radiation doses with various body weights of phantoms in brain 128-slice MDCT examination.J Radiat Res. 2019 Jul 1;60(4):466-475. doi: 10.1093/jrr/rrz029. J Radiat Res. 2019. PMID: 31197311 Free PMC article.
-
Dose Comparison Using Thermoluminescent Dosimeters During Multislice Computed Tomography With Different Parameters for Simulated Spine Tumor Examination.Health Phys. 2018 Aug;115(2):275-280. doi: 10.1097/HP.0000000000000869. Health Phys. 2018. PMID: 29894329
-
Effective dose to patient measurements for flat-detector computed tomography protocols in acute stroke care.Eur Radiol. 2020 Sep;30(9):5082-5088. doi: 10.1007/s00330-020-06891-w. Epub 2020 Apr 28. Eur Radiol. 2020. PMID: 32346793
-
Verification of organ doses calculated by a dose monitoring software tool based on Monte Carlo Simulation in thoracic CT protocols.Acta Radiol. 2018 Mar;59(3):322-326. doi: 10.1177/0284185117716199. Epub 2017 Jun 15. Acta Radiol. 2018. PMID: 28618854
-
Radiation exposure in perfusion CT of the brain.J Comput Assist Tomogr. 2014 Jan-Feb;38(1):25-8. doi: 10.1097/RCT.0b013e3182a3f9a0. J Comput Assist Tomogr. 2014. PMID: 24378887
Cited by
-
Evaluating environmental radiations of the tomotherapy facility by optimizing full factorial design of the TLD technique.Technol Health Care. 2022;30(S1):481-491. doi: 10.3233/THC-THC228044. Technol Health Care. 2022. PMID: 35124622 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources