Electromagnetic fields in neonatal incubators: the reasons for an alert
- PMID: 28988507
- DOI: 10.1080/14767058.2017.1390559
Electromagnetic fields in neonatal incubators: the reasons for an alert
Abstract
Background: Neonatal incubators are important tools for sick newborns in the first few days of life. Nevertheless, their electric engine, often very close to the newborn's body, emits electromagnetic fields (EMF) to which newborns are exposed. Aim of this paper is to review the available literature on EMF exposure in incubators, and the effects of such exposures on newborns that have been investigated.
Methods: We carried out a systematic review of studies about EMF emissions produced by incubators, using Medline and Embase databases from 1993 to 2017.
Results: We retrieved 15 papers that described the EMF exposure in incubators and their biological effects on babies. EMF levels in incubators appear to be between 2 and 100 mG, depending on the distance of the mattress from the electric engine. In some cases, they exceed this range. These values interfere with melatonin production or with vagal tone. Even caregivers are exposed to high EMF, above 200 mG, when working at close contact with the incubators.
Conclusion: EMF have been described as potentially hazardous for human health, and values reported in this review are an alert to prevent babies' and caregivers' exposure when close to the incubators. A precautionary approach should be adopted in future incubator design, to prevent high exposures of newborns in incubators and of caregivers as well.
Keywords: Electricity; electromagnetic fields; incubator; newborn.
Similar articles
-
Is newborn melatonin production influenced by magnetic fields produced by incubators?Early Hum Dev. 2012 Aug;88(8):707-10. doi: 10.1016/j.earlhumdev.2012.02.015. Epub 2012 Mar 14. Early Hum Dev. 2012. PMID: 22421197 Clinical Trial.
-
Electromagnetic fields produced by incubators influence heart rate variability in newborns.Arch Dis Child Fetal Neonatal Ed. 2008 Jul;93(4):F298-301. doi: 10.1136/adc.2007.132738. Epub 2008 May 1. Arch Dis Child Fetal Neonatal Ed. 2008. PMID: 18450804
-
Reduction of exposure of newborns and caregivers to very high electromagnetic fields produced by incubators.Med Phys. 2005 Jan;32(1):149-52. doi: 10.1118/1.1829404. Med Phys. 2005. PMID: 15719965
-
The infant incubator in the neonatal intensive care unit: unresolved issues and future developments.J Perinat Med. 2009;37(6):587-98. doi: 10.1515/JPM.2009.109. J Perinat Med. 2009. PMID: 19591569 Review.
-
EUROPAEM EMF Guideline 2016 for the prevention, diagnosis and treatment of EMF-related health problems and illnesses.Rev Environ Health. 2016 Sep 1;31(3):363-97. doi: 10.1515/reveh-2016-0011. Rev Environ Health. 2016. PMID: 27454111 Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical