Early outcomes of percutaneous pulmonary valve implantation using the Edwards SAPIEN XT transcatheter heart valve system

Affiliations
Observational Study

Early outcomes of percutaneous pulmonary valve implantation using the Edwards SAPIEN XT transcatheter heart valve system

Nikolaus A Haas et al. Int J Cardiol. .

Abstract

Background: Patients with congenital or acquired heart defects affecting the pulmonary valve and right ventricular outflow tract (RVOT) commonly require multiple surgical interventions, resulting in significant morbidity. A less invasive alternative is percutaneous pulmonary valve implantation (PPVI). Though studies have previously reported the safety and efficacy of the early generation transcatheter heart valves (THVs), data on more recent devices are severely lacking.

Methods and results: We performed a multinational, multicentre, retrospective, observational registry analysis of patients who underwent PPVI using the Edwards SAPIEN XT THV. Of the 46 patients that were enrolled, the majority had tetralogy of Fallot as the underlying diagnosis (58.7%), and stentless xenograft as the most common RVOT anatomy (34.8%). Procedural success rate was high (93.5%), with a low frequency of periprocedural complications and adverse events (6.5% and 10.9%, respectively). At 30days post-procedure, NYHA class had improved significantly (90.6% were at NYHA I or II). The rate of moderate/severe pulmonary regurgitation had decreased from 76.1% at baseline to 5.0% at 30days, and the calculated peak systolic gradient had decreased from 45.2 (SD±21.3) mmHg to 16.4 (SD±8.0) mmHg, with these values remaining low up to 2years.

Conclusions: The data suggest the efficacy and safety of the SAPIEN XT THV in PPVI in common anatomies in patients with conduits, as well as those with native pulmonary valves or transannular patches. Continued data collection is necessary to verify long-term findings. CLINICALTRIALS.

Gov identifier: NCT02302131.

Keywords: Congenital heart disease; Percutaneous pulmonary valve implantation; Pulmonary valve; SAPIEN XT.