. 2018 Apr;27(4):921-930.
doi: 10.1007/s00586-017-5333-y. Epub 2017 Oct 14.

Robot-assisted and conventional freehand pedicle screw placement: a systematic review and meta-analysis of randomized controlled trials

Affiliations
Meta-Analysis

Robot-assisted and conventional freehand pedicle screw placement: a systematic review and meta-analysis of randomized controlled trials

Shutao Gao et al. Eur Spine J. 2018 Apr.

Abstract

Purpose: Several studies have revealed that robot-assisted technique might improve the pedicle screw insertion accuracy, but owing to the limited sample sizes in the individual study reported up to now, whether or not robot-assisted technique is superior to conventional freehand technique is indefinite. Thus, we performed this systematic review and meta-analysis based on randomized controlled trials to assess which approach is better.

Methods: Electronic databases including PubMed, EMBASE, CENTRAL, ISI Web of Science, CNKI and WanFang were systematically searched to identify potentially eligible articles. Main endpoints containing the accuracy of pedicle screw implantation and proximal facet joint violation were evaluated as risk ratio (RR) and the associated 95% confidence intervals (95% CIs), while radiation exposure and surgical duration were presented as mean difference (MD) or standard mean difference (SMD). Meta-analyses were performed using RevMan 5.3 software.

Results: Six studies involving 158 patients (688 pedicle screws) in robot-assisted group and 148 patients (672 pedicle screws) in freehand group were identified matching our study. The Grade A accuracy rate in robot-assisted group was superior to freehand group (RR 1.03, 95% CI 1.00, 1.06; P = 0.04), but the Grade A + B accuracy rate did not differ between the two groups (RR 1.01, 95% CI 0.99, 1.02; P = 0.29). With regard to proximal facet joint violation, the combined results suggested that robot-assisted group was associated with significantly fewer proximal facet joint violation than freehand group (RR 0.07, 95% CI 0.01, 0.55; P = 0.01). As was the radiation exposure, our findings suggested that robot-assisted technique could significantly reduce the intraoperative radiation time (MD - 12.38, 95% CI - 17.95, - 6.80; P < 0.0001) and radiation dosage (SMD - 0.64, 95% CI - 0.85, - 0.43; P < 0.00001). But the overall surgical duration was longer in robot-assisted group than conventional freehand group (MD 20.53, 95% CI 5.17, 35.90; P = 0.009).

Conclusions: The robot-assisted technique was associated with equivalent accuracy rate of pedicle screw implantation, fewer proximal facet joint violation, less intraoperative radiation exposure but longer surgical duration than freehand technique. Powerful evidence relies on more randomized controlled trials with high quality and larger sample size in the future.

Keywords: Meta-analysis; Pedicle screw; Robot assisted; Spine surgery; Systematic review.

PubMed Disclaimer

Comment in

References

    1. J Neurosurg Spine. 2014 Feb;20(2):196-203 - PubMed
    1. Eur Spine J. 2014 Feb;23(2):291-7 - PubMed
    1. J Robot Surg. 2016 Jun;10 (2):145-50 - PubMed
    1. Spine (Phila Pa 1976). 1992 Mar;17(3):299-306 - PubMed
    1. Occup Med (Lond). 2005 Sep;55(6):498-500 - PubMed

MeSH terms

LinkOut - more resources