Canine central corneal thickness measurements via Pentacam-HR® , optical coherence tomography (Optovue iVue® ), and high-resolution ultrasound biomicroscopy
- PMID: 29034562
- DOI: 10.1111/vop.12518
Canine central corneal thickness measurements via Pentacam-HR® , optical coherence tomography (Optovue iVue® ), and high-resolution ultrasound biomicroscopy
Abstract
Objective: To compare central corneal thickness (CCT) values in canine eyes using Pentacam-HR® Scheimpflug imaging (Pentacam), Optovue® iVue spectral-domain optical coherence tomography (SD-OCT), and high-resolution ultrasound biomicroscopy (UBM) and generate normative canine Pentacam CCT values.
Animals studied: Twenty-four client-owned dogs (37 eyes) with nondiseased cornea(s) presenting to the Colorado State University Veterinary Teaching Hospital.
Procedures: Corneal images were acquired via Pentacam, SD-OCT, and UBM in the listed order. Machine-calculated values of CCT from Pentacam and SD-OCT were compared to operator-measured values from UBM. Bland-Altman analysis was performed to evaluate agreement between instruments.
Results: Mean CCT ± SD measured by Pentacam was 629.73 ± 64.57 μm, by SD-OCT was 610.56 ± 57.48 μm, and by UBM was 689.77 ± 55.93 μm. On average, Pentacam CCT was 19.17 ± 32.90 μm (3%) thicker than SD-OCT and 65.12 ± 44.52 μm (10.3%) thinner than UBM. SD-OCT was on average 82.47 μm (13.5% ) thinner than UBM. The 95% limits of agreement were (-45.31, 83.65), (-152.38, 22.13), and (-126.674, -38.270) for Pentacam vs. SD-OCT, Pentacam vs. UBM, and SD-OCT vs. UBM respectively. All differences were statistically significant (P < 0.01).
Conclusions: Considering there is an average of 7.5% normal diurnal variation in canine CCT, a 3.0% difference between Pentacam and SD-OCT values is likely not clinically relevant. However, Pentacam measurements were both statistically and clinically significantly different from UBM and SD-OCT measurements.
Keywords: Pentacam; Scheimpflug; canine; central corneal thickness; optical coherence tomography; ultrasound biomicroscopy.
© 2017 American College of Veterinary Ophthalmologists.
Similar articles
-
Comparison between Pentacam-HR and optical coherence tomographycentral corneal thickness measurements in healthy feline eyes.Vet Ophthalmol. 2016 Jul;19 Suppl 1:105-14. doi: 10.1111/vop.12397. Vet Ophthalmol. 2016. PMID: 27370363
-
Measurement of central corneal thickness by high-resolution Scheimpflug imaging, Fourier-domain optical coherence tomography and ultrasound pachymetry.Acta Ophthalmol. 2012 Aug;90(5):449-55. doi: 10.1111/j.1755-3768.2010.01947.x. Epub 2010 Jun 18. Acta Ophthalmol. 2012. PMID: 20560892
-
Comparison of central corneal thickness in dogs measured by ultrasound pachymetry and ultrasound biomicroscopy.Vet J. 2018 Feb;232:13-14. doi: 10.1016/j.tvjl.2017.12.003. Epub 2017 Dec 5. Vet J. 2018. PMID: 29428083
-
To Study and Determine the Role of Anterior Segment Optical Coherence Tomography and Ultrasound Biomicroscopy in Corneal and Conjunctival Tumors.J Ophthalmol. 2016;2016:1048760. doi: 10.1155/2016/1048760. Epub 2016 Dec 6. J Ophthalmol. 2016. PMID: 28050274 Free PMC article. Review.
-
The role of imaging technologies for ocular surface tumors.Curr Opin Ophthalmol. 2021 Jul 1;32(4):369-378. doi: 10.1097/ICU.0000000000000771. Curr Opin Ophthalmol. 2021. PMID: 33989235 Free PMC article. Review.
Cited by
-
A Pilot Study to Evaluate the Usefulness of Optical Coherence Tomography for Staging Iris Pigmented Lesions in Cats.Vet Sci. 2024 Jun 7;11(6):261. doi: 10.3390/vetsci11060261. Vet Sci. 2024. PMID: 38922008 Free PMC article.
-
Corneal thickness and anterior chamber depth of the normal adult horse as measured by ultrasound biomicroscopy.Vet Ophthalmol. 2022 May;25 Suppl 1(Suppl 1):17-24. doi: 10.1111/vop.12971. Epub 2022 Jan 27. Vet Ophthalmol. 2022. PMID: 35084084 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources