Eight-Year Safety Data for Round and Anatomical Silicone Gel Breast Implants
- PMID: 29040345
- DOI: 10.1093/asj/sjx117
Eight-Year Safety Data for Round and Anatomical Silicone Gel Breast Implants
Abstract
Background: The safety and efficacy of all medical devices, including breast implants, is important and consistent performance is best shown by undertaking long-term clinical and vigilance studies. Local complications such as capsular contracture and rupture are risks often associated with breast implant surgery.
Objectives: The authors investigate and evaluate the safety and performance of Eurosilicone's (Eurosilicone S.A.S, Apt Cedex, France) Cristalline Paragel breast implants at 8 years postimplantation.
Methods: In this prospective clinical study, 995 Eurosilicone textured cohesive Cristalline Paragel mammary implants were implanted in 526 women undergoing augmentation and reconstructive surgery at 17 centers across France. Complications were recorded at 3 months and annually thereafter for 8 years. Descriptive statistics were used and key complications were analysed using the Kaplan-Meier method.
Results: Capsular contracture was reported in 8.5% of implants across all cohorts through 8 years. The Kaplan-Meier risk of capsular contracture (Baker Grade III/IV) per implant was 8.4% in the primary augmentation cohort and 18.0% in the primary reconstruction cohort. Eight implant ruptures were identified by surgeon examination during this follow-up period. The Kaplan-Meier risk of rupture occurring within 8 years postimplantation, across all cohorts, was 1.4% per patient and 0.9% per implant. Actual implant removal rate (explantation/exchange) was 6.0% and 13.8% for primary augmentation and primary reconstruction, respectively. Actual rates of local complications including infection and seroma were low with risk rates of 0.6% and 0.2% by subject.
Conclusions: This multicenter clinical study involving Eurosilicone's silicone gel breast implants in both round and shaped profiles demonstrates an excellent safety and efficacy profile through 8 years.
© 2017 The American Society for Aesthetic Plastic Surgery, Inc. Reprints and permission: journals.permissions@oup.com
Similar articles
-
Ten-Year Safety Data for Eurosilicone's Round and Anatomical Silicone Gel Breast Implants.Aesthet Surg J Open Forum. 2019 Apr 27;1(2):ojz012. doi: 10.1093/asjof/ojz012. eCollection 2019 Jun. Aesthet Surg J Open Forum. 2019. PMID: 33791608 Free PMC article.
-
Five-year Safety Data for Eurosilicone's Round and Anatomical Silicone Gel Breast Implants.Plast Reconstr Surg Glob Open. 2014 May 7;2(4):e138. doi: 10.1097/GOX.0000000000000082. eCollection 2014 Apr. Plast Reconstr Surg Glob Open. 2014. PMID: 25289331 Free PMC article.
-
Safety and Efficacy of the Sientra Silicone Gel Round and Shaped Breast Implants: 6-Year Results of the U.S. Postapproval Study.Plast Reconstr Surg. 2024 Jul 1;154(1):44e-56e. doi: 10.1097/PRS.0000000000010953. Epub 2023 Aug 11. Plast Reconstr Surg. 2024. PMID: 37563756 Free PMC article.
-
Primary breast augmentation clinical trial outcomes stratified by surgical incision, anatomical placement and implant device type.J Plast Reconstr Aesthet Surg. 2013 Sep;66(9):1165-72. doi: 10.1016/j.bjps.2013.04.046. Epub 2013 May 9. J Plast Reconstr Aesthet Surg. 2013. PMID: 23664574 Review.
-
Round Form-Stable Breast Implants: Diagnosis and Management of Complications.Plast Reconstr Surg. 2019 Jul;144(1S Utilizing a Spectrum of Cohesive Implants in Aesthetic and Reconstructive Breast Surgery):73S-81S. doi: 10.1097/PRS.0000000000005953. Plast Reconstr Surg. 2019. PMID: 31246764 Review.
Cited by
-
Letter to the Editor: Discussion of the Article "The Emerging Crisis of Stakeholders in Implant-based Augmentation Mammaplasty in Korea".J Korean Med Sci. 2020 May 11;35(18):e172. doi: 10.3346/jkms.2020.35.e172. J Korean Med Sci. 2020. PMID: 32383367 Free PMC article. No abstract available.
-
A Rationale for Micro-textured Breast Implant Augmentation.Aesthet Surg J Open Forum. 2022 Mar 30;4:ojac020. doi: 10.1093/asjof/ojac020. eCollection 2022. Aesthet Surg J Open Forum. 2022. PMID: 35601234 Free PMC article.
-
Six-Year Prospective Outcomes of Primary Breast Augmentation With Nano Surface Implants.Aesthet Surg J. 2019 Apr 8;39(5):495-508. doi: 10.1093/asj/sjy196. Aesthet Surg J. 2019. PMID: 30423014 Free PMC article.
-
Plastic Surgeons Defend Textured Breast Implants at 2019 U.S. Food and Drug Administration Hearing: Why It Is Time to Reconsider.Plast Reconstr Surg Glob Open. 2019 Aug 30;7(8):e2410. doi: 10.1097/GOX.0000000000002410. eCollection 2019 Aug. Plast Reconstr Surg Glob Open. 2019. PMID: 31592028 Free PMC article.
-
Asiaticoside Combined With Carbon Ion Implantation to Improve the Biocompatibility of Silicone Rubber and to Reduce the Risk of Capsule Contracture.Front Bioeng Biotechnol. 2022 May 12;10:810244. doi: 10.3389/fbioe.2022.810244. eCollection 2022. Front Bioeng Biotechnol. 2022. PMID: 35646845 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical