Multicenter Study
. 2018 Feb 17;38(2):151-161.
doi: 10.1093/asj/sjx117.

Eight-Year Safety Data for Round and Anatomical Silicone Gel Breast Implants

Affiliations
Multicenter Study

Eight-Year Safety Data for Round and Anatomical Silicone Gel Breast Implants

Franck Duteille et al. Aesthet Surg J. .

Abstract

Background: The safety and efficacy of all medical devices, including breast implants, is important and consistent performance is best shown by undertaking long-term clinical and vigilance studies. Local complications such as capsular contracture and rupture are risks often associated with breast implant surgery.

Objectives: The authors investigate and evaluate the safety and performance of Eurosilicone's (Eurosilicone S.A.S, Apt Cedex, France) Cristalline Paragel breast implants at 8 years postimplantation.

Methods: In this prospective clinical study, 995 Eurosilicone textured cohesive Cristalline Paragel mammary implants were implanted in 526 women undergoing augmentation and reconstructive surgery at 17 centers across France. Complications were recorded at 3 months and annually thereafter for 8 years. Descriptive statistics were used and key complications were analysed using the Kaplan-Meier method.

Results: Capsular contracture was reported in 8.5% of implants across all cohorts through 8 years. The Kaplan-Meier risk of capsular contracture (Baker Grade III/IV) per implant was 8.4% in the primary augmentation cohort and 18.0% in the primary reconstruction cohort. Eight implant ruptures were identified by surgeon examination during this follow-up period. The Kaplan-Meier risk of rupture occurring within 8 years postimplantation, across all cohorts, was 1.4% per patient and 0.9% per implant. Actual implant removal rate (explantation/exchange) was 6.0% and 13.8% for primary augmentation and primary reconstruction, respectively. Actual rates of local complications including infection and seroma were low with risk rates of 0.6% and 0.2% by subject.

Conclusions: This multicenter clinical study involving Eurosilicone's silicone gel breast implants in both round and shaped profiles demonstrates an excellent safety and efficacy profile through 8 years.

PubMed Disclaimer

Similar articles

Cited by

Publication types

MeSH terms