. 2018 Feb;48(2):227-234.
doi: 10.1007/s00247-017-3996-5. Epub 2017 Oct 18.

Female gonadal shielding with automatic exposure control increases radiation risks

Affiliations

Female gonadal shielding with automatic exposure control increases radiation risks

Summer L Kaplan et al. Pediatr Radiol. 2018 Feb.

Abstract

Background: Gonadal shielding remains common, but current estimates of gonadal radiation risk are lower than estimated risks to colon and stomach. A female gonadal shield may attenuate active automatic exposure control (AEC) sensors, resulting in increased dose to colon and stomach as well as to ovaries outside the shielded area.

Objective: We assess changes in dose-area product (DAP) and absorbed organ dose when female gonadal shielding is used with AEC for pelvis radiography.

Materials and methods: We imaged adult and 5-year-old equivalent dosimetry phantoms using pelvis radiograph technique with AEC in the presence and absence of a female gonadal shield. We recorded DAP and mAs and measured organ absorbed dose at six internal sites using film dosimetry.

Results: Female gonadal shielding with AEC increased DAP 63% for the 5-year-old phantom and 147% for the adult phantom. Absorbed organ dose at unshielded locations of colon, stomach and ovaries increased 21-51% in the 5-year-old phantom and 17-100% in the adult phantom. Absorbed organ dose sampled under the shield decreased 67% in the 5-year-old phantom and 16% in the adult phantom.

Conclusion: Female gonadal shielding combined with AEC during pelvic radiography increases absorbed dose to organs with greater radiation sensitivity and to unshielded ovaries. Difficulty in proper use of gonadal shields has been well described, and use of female gonadal shielding may be inadvisable given the risks of increasing radiation.

Keywords: Automatic exposure control; Gonadal shield; Gonads; Ovaries; Radiation dose; Radiography; Shielding.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Br J Radiol. 2012 Aug;85(1016):e442-7 - PubMed
    1. Ann ICRP. 2007;37(2-4):1-332 - PubMed
    1. Br J Radiol. 1955 May;28(329):266-73 - PubMed
    1. Ann R Coll Surg Engl. 2003 Nov;85(6):422-5 - PubMed
    1. Orthopedics. 2017 Jul 1;40(4):e623-e627 - PubMed

LinkOut - more resources