Female gonadal shielding with automatic exposure control increases radiation risks
- PMID: 29046919
- DOI: 10.1007/s00247-017-3996-5
Female gonadal shielding with automatic exposure control increases radiation risks
Abstract
Background: Gonadal shielding remains common, but current estimates of gonadal radiation risk are lower than estimated risks to colon and stomach. A female gonadal shield may attenuate active automatic exposure control (AEC) sensors, resulting in increased dose to colon and stomach as well as to ovaries outside the shielded area.
Objective: We assess changes in dose-area product (DAP) and absorbed organ dose when female gonadal shielding is used with AEC for pelvis radiography.
Materials and methods: We imaged adult and 5-year-old equivalent dosimetry phantoms using pelvis radiograph technique with AEC in the presence and absence of a female gonadal shield. We recorded DAP and mAs and measured organ absorbed dose at six internal sites using film dosimetry.
Results: Female gonadal shielding with AEC increased DAP 63% for the 5-year-old phantom and 147% for the adult phantom. Absorbed organ dose at unshielded locations of colon, stomach and ovaries increased 21-51% in the 5-year-old phantom and 17-100% in the adult phantom. Absorbed organ dose sampled under the shield decreased 67% in the 5-year-old phantom and 16% in the adult phantom.
Conclusion: Female gonadal shielding combined with AEC during pelvic radiography increases absorbed dose to organs with greater radiation sensitivity and to unshielded ovaries. Difficulty in proper use of gonadal shields has been well described, and use of female gonadal shielding may be inadvisable given the risks of increasing radiation.
Keywords: Automatic exposure control; Gonadal shield; Gonads; Ovaries; Radiation dose; Radiography; Shielding.
Similar articles
-
Automatic exposure control in pediatric and adult computed tomography examinations: can we estimate organ and effective dose from mean MAS reduction?Invest Radiol. 2011 Oct;46(10):654-62. doi: 10.1097/RLI.0b013e3182213c55. Invest Radiol. 2011. PMID: 21673583
-
Gonadal Shielding in Radiography: A Best Practice?Radiol Technol. 2016 Nov;88(2):127-134. Radiol Technol. 2016. PMID: 27837123
-
Achievable dose reductions with gonadal shielding for children and adults during abdominal/pelvic radiographic examinations: A Monte Carlo simulation.Med Phys. 2020 Nov;47(11):5514-5522. doi: 10.1002/mp.14495. Epub 2020 Oct 13. Med Phys. 2020. PMID: 32978986
-
Comparison of organ dosimetry methods and effective dose calculation methods for paediatric CT.Australas Phys Eng Sci Med. 2012 Jun;35(2):117-34. doi: 10.1007/s13246-012-0134-4. Epub 2012 Apr 11. Australas Phys Eng Sci Med. 2012. PMID: 22492218 Review.
-
Exploring Past to Present Shielding Guidelines.Radiol Technol. 2023 Nov;95(2):84-93. Radiol Technol. 2023. PMID: 37940176 Review.
Cited by
-
Death of the ALARA Radiation Protection Principle as Used in the Medical Sector.Dose Response. 2020 Apr 29;18(2):1559325820921641. doi: 10.1177/1559325820921641. eCollection 2020 Apr-Jun. Dose Response. 2020. PMID: 32425724 Free PMC article.
-
Unified Open Hardware Platform for Digital X-Ray Devices; its Conceptual Model and First Implementation.IEEE J Transl Eng Health Med. 2020 Jun 4;8:1800311. doi: 10.1109/JTEHM.2020.3000011. eCollection 2020. IEEE J Transl Eng Health Med. 2020. PMID: 32617198 Free PMC article.
-
Gonad shielding in pelvic radiography: modern optimised X-ray systems might allow its discontinuation.Insights Imaging. 2020 Feb 7;11(1):15. doi: 10.1186/s13244-019-0828-1. Insights Imaging. 2020. PMID: 32030539 Free PMC article.
-
European consensus on patient contact shielding.Insights Imaging. 2021 Dec 23;12(1):194. doi: 10.1186/s13244-021-01085-4. Insights Imaging. 2021. PMID: 34939154 Free PMC article.
-
Effect of Silicone Rubber-Lead (SR-Pb) Thickness on Dose Reduction and Image Quality as Gonad Shield.J Biomed Phys Eng. 2020 Dec 1;10(6):699-706. doi: 10.31661/jbpe.v0i0.1912-1007. eCollection 2020 Dec. J Biomed Phys Eng. 2020. PMID: 33364207 Free PMC article.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical