Reproducibility and acceptability of ultrasound measurements of head-perineum distance
- PMID: 29068541
- DOI: 10.1111/aogs.13251
Reproducibility and acceptability of ultrasound measurements of head-perineum distance
Abstract
Introduction: We aimed to test the reproducibility of head-perineum distance (HPD) measurements using two different ultrasound devices and five examiners, to compare ultrasound measurements and clinical assessments and to study if ultrasound examinations were acceptable for women in labor.
Material and methods: A reproducibility study was performed at Lund University Hospital, Sweden and Landspitali University Hospital, Iceland from February 2015 to February 2017. The study population comprised 40 healthy women in labor. HPD was measured with three replicate measurements from each woman with two different ultrasound devices, and the measurements were compared with clinical assessments. Acceptability was tested with a visual analog scale (VAS), and the mean VAS score from both ultrasound devices was compared with the VAS score from clinical palpation.
Results: The median time interval between start of examinations with devices was 10 min (range 1-26 min). The intra-observer repeatability coefficient was 4.3 mm and the intraclass correlation coefficient was 0.97 (95% CI 0.95-0.98). The intraclass correlation coefficient between the two devices was 0.86 (95% CI 0.74-0.93) and limits of agreement were -9.6 mm to 16.6 mm. However, we observed a significant mean HPD difference between devices (3.5 mm; 95% CI 1.4-5.6 mm). Clinical assessments and the mean measurements of HPD were correlated (r = 0.64, p < 0.01). We found significant differences in acceptability in favor of ultrasound. The mean VAS score for both ultrasound devices was 2.0 vs. 4.1 for clinical examination (p < 0.01).
Conclusion: We found excellent intra-observer repeatability, good correlation but significant difference between devices. Women reported less discomfort with ultrasound than with clinical examinations.
Keywords: Delivery; head-perineum distance; labor; perineum; ultrasound.
© 2017 Nordic Federation of Societies of Obstetrics and Gynecology.
Similar articles
-
Agreement between two- and three-dimensional transperineal ultrasound methods in assessing fetal head descent in the first stage of labor.Ultrasound Obstet Gynecol. 2012 Mar;39(3):310-5. doi: 10.1002/uog.9065. Ultrasound Obstet Gynecol. 2012. PMID: 21630362
-
New technique for automatic sonographic measurement of change in head-perineum distance and angle of progression during active phase of second stage of labor.Ultrasound Obstet Gynecol. 2020 Oct;56(4):597-602. doi: 10.1002/uog.21963. Ultrasound Obstet Gynecol. 2020. PMID: 31909525
-
Agreement between two- and three-dimensional transperineal ultrasound methods for assessment of fetal head-symphysis distance in active labor.Ultrasound Obstet Gynecol. 2014 Feb;43(2):183-8. doi: 10.1002/uog.13204. Epub 2014 Jan 13. Ultrasound Obstet Gynecol. 2014. PMID: 24006290
-
Sonographic prediction of vaginal delivery in prolonged labor: a two-center study.Ultrasound Obstet Gynecol. 2014 Feb;43(2):195-201. doi: 10.1002/uog.13210. Epub 2013 Dec 22. Ultrasound Obstet Gynecol. 2014. PMID: 24105705
-
Agreement between transperineal ultrasound measurements and digital examinations of cervical dilatation during labor.BMC Pregnancy Childbirth. 2015 Oct 24;15:273. doi: 10.1186/s12884-015-0704-z. BMC Pregnancy Childbirth. 2015. PMID: 26496894 Free PMC article.
Cited by
-
The views of pregnant women, midwives, and a women's panel on intrapartum ultrasound research: A pilot study.Australas J Ultrasound Med. 2019 Jun 19;22(3):186-190. doi: 10.1002/ajum.12162. eCollection 2019 Aug. Australas J Ultrasound Med. 2019. PMID: 34760555 Free PMC article.
-
Reproducibility of Breech Progression Angle: Standardization of Transperineal Measurements and Development of Image-Based Checklist for Quality Control.Diagnostics (Basel). 2025 Jul 11;15(14):1757. doi: 10.3390/diagnostics15141757. Diagnostics (Basel). 2025. PMID: 40722507 Free PMC article.
-
Intrapartum Ultrasound Guidance to Make Safer Any Obstetric Intervention: Fetal Head Rotation, Assisted Vaginal Birth, Breech Delivery of the Second Twin.Clin Obstet Gynecol. 2024 Dec 1;67(4):730-738. doi: 10.1097/GRF.0000000000000891. Epub 2024 Oct 18. Clin Obstet Gynecol. 2024. PMID: 39431493 Free PMC article. Review.
-
Time to delivery based on sonographic assessment prior to forceps and vacuum.Australas J Ultrasound Med. 2019 Jan 21;22(2):111-117. doi: 10.1002/ajum.12121. eCollection 2019 May. Australas J Ultrasound Med. 2019. PMID: 34760547 Free PMC article.
-
Autonomous robotic ultrasound scanning system: a key to enhancing image analysis reproducibility and observer consistency in ultrasound imaging.Front Robot AI. 2025 Feb 5;12:1527686. doi: 10.3389/frobt.2025.1527686. eCollection 2025. Front Robot AI. 2025. PMID: 39975565 Free PMC article.