Case Reports
. 2017 Dec:163:53-59.
doi: 10.1016/j.clineuro.2017.10.011. Epub 2017 Oct 16.

Management of major vascular injury during pedicle screw instrumentation of thoracolumbar spine

Affiliations
Case Reports

Management of major vascular injury during pedicle screw instrumentation of thoracolumbar spine

Aleem K Mirza et al. Clin Neurol Neurosurg. 2017 Dec.

Abstract

Objectives: Vascular injury is a rare complication of spinal instrumentation. Presentation can vary from immediate hemorrhage to pseudoaneurysm formation. In the literature, surgical approach to repair has varied based on anatomy, acuity of diagnosis, infection, and available technology. In this manuscript, we aim to describe our institutional experience with vascular injuries in thoraco-lumbar spine surgery.

Patients and methods: We report our institutional experience of three cases of vascular injury secondary to pedicle screw misplacement and their management, as well as a review of the literature.

Results: The first case had a history of previous instrumentation and presented with back pain and fever. The patient was taken for instrumentation exploration via a posterior approach. Aortic violation was discovered at T6 intraoperatively during instrumentation removal and the patient underwent emergent endovascular repair. The second case presented with chronic back pain after multiple prior posterior fusions and CT angiogram showing screw perforation on the aorta at T10. The patient underwent elective endovascular repair with synchronous removal of the instrumentation. The third case presented with radicular leg pain 6 months after L4-S1 posterior lumbar interbody fusion, with CT scan demonstrating the left S1 screw abutting the L5 nerve root and common iliac vein. The patient underwent elective instrumentation revision with intraoperative venography.

Conclusion: Major vascular injury is a known complication of spinal surgery, especially if it involves instrumentation with pedicle screws. Treatment approach has evolved with the advancement of endovascular technology; however, open surgery remains an option when anatomy or infection is prohibitive. In the elective setting, preoperative planning with attention to surgical approach, positioning, and contingencies, should occur in a multidisciplinary fashion. Repair with an aortic stent-graft cuff may minimize unnecessary coverage of the descending thoracic aorta and intercostal arteries.

Keywords: Aorta; Iliac vein; Interbody fusion; Lumbar spine; Multidisciplinary; Pedicle screw; Screw abutment; Screw perforation; Spine surgery; Thoracic spine; Vascular injury; Venography.