. 2018 Jan;221(1):48-53.
doi: 10.1016/j.ijheh.2017.10.002. Epub 2017 Oct 12.

Waterpipe cafés are hazardous to your health: Determination of a waterpipe specific calibration factor

Affiliations

Waterpipe cafés are hazardous to your health: Determination of a waterpipe specific calibration factor

Mark J Travers et al. Int J Hyg Environ Health. 2018 Jan.

Abstract

Introduction: The use of a waterpipe to smoke tobacco has emerged as a popular trend in the United States. Waterpipe smoking establishments have had an increasing presence in the U.S., despite smoke-free air legislation. Dangers of waterpipe smoking have been documented, but less data has been gathered about the waterpipe café itself. This project sought to determine a waterpipe-specific calibration factor (CF) for measuring waterpipe aerosol, and field-test this CF by conducting surveillance on the existing waterpipe cafés of western and central New York.

Methods: Nine laboratory-controlled experiments were conducted to determine a waterpipe-specific CF. In the lab, two TSI SidePak AM510 Personal Aerosol Monitors and two sampling trains for gravimetric PM2.5 sampling were present during waterpipe smoking sessions (lasting 1-3h). Indoor air quality was assessed in 7 waterpipe cafés in three counties of New York, and real-time measurements of particulate matter (PM2.5) and carbon monoxide (CO) were obtained.

Results: Results from the 9 controlled waterpipe experiments determined a calibration factor of 0.38 (SD 0.08), which should be used to convert SidePak measurements to true PM2.5 measurements. When applying the CF to the measurements taken in the 7 public waterpipe venues, the mean PM2.5 concentration was 515μg/m3 micrograms per cubic meter (SD=338.8) while the mean ambient CO was 20.5ppm (SD=18.3). The mean active smoking density was 2.41 waterpipes per 100m3 of air. The PM2.5 levels increased with increasing active smoking density (rho=0.68, p=0.09).

Conclusions: Applying the waterpipe-specific CF for the SidePak, 0.38, allowed for field assessments to be conducted in locations with waterpipe smoke to determine accurate particle exposure concentrations. The concentrations of both particulate matter and carbon monoxide were above established air quality standards and therefore increase the health risks of both patrons and workers of these establishments.

Keywords: Calibration factor; Indoor air quality; PM(2.5); Particulate matter; Waterpipe.

PubMed Disclaimer

Conflict of interest statement

Declaration of Interests: No conflicts of interest for MJT, JAK, LV

Figures

FIGURE 1.
FIGURE 1.
Association between SidePak Aerosol Monitor and Gravimetric PM2.5 in Controlled Lab Experiments with Cigarette Smoke & Hookah Smoke
FIGURE 2.
FIGURE 2.
SidePak AM510 versus Piezobalance (SidePak data multiplied by 0.38 calibration factor)
Figure 3:
Figure 3:
Real-Time Plots of PM2.5 (calibration factor = 0.38) and Carbon Monoxide in 7 Hookah Cafes Sampled.

Similar articles

Cited by

References

    1. Akl EA, Gaddam S, Gunukula SK, Honeine R, Jaoude PA, Irani J, 2010. The effects of waterpipe tobacco smoking on health outcomes: a systematic review. International journal of epidemiology 39, 834–857. - PubMed
    1. Al Rashidi M, Shihadeh A, Saliba NA, 2008. Volatile aldehydes in the mainstream smoke of the narghile waterpipe. Food and chemical toxicology : an international journal published for the British Industrial Biological Research Association 46, 3546–3549. - PMC - PubMed
    1. Center for Disease Control and Prevention, 2004. Indoor air quality in hospitality venues before and after implementation of a clean indoor air law--Western New York, 2003. MMWR Morb Mortal Wkly Rep 53, 1038–1041. - PubMed
    1. Cobb CO, Vansickel AR, Blank MD, Jentink K, Travers MJ, Eissenberg T, 2012. Indoor air quality in Virginia waterpipe cafes. Tob Control. - PMC - PubMed
    1. Cobb CO, Vansickel AR, Blank MD, Jentink K, Travers MJ, Eissenberg T, 2013. Indoor air quality in Virginia waterpipe cafes. Tobacco control 22, 338–343. - PMC - PubMed

Publication types

MeSH terms

Substances