Randomized Controlled Trial
. 2017 Dec 12;70(23):2852-2862.
doi: 10.1016/j.jacc.2017.10.010. Epub 2017 Oct 31.

3-Year Clinical Outcomes With Everolimus-Eluting Bioresorbable Coronary Scaffolds: The ABSORB III Trial

Affiliations
Free article
Randomized Controlled Trial

3-Year Clinical Outcomes With Everolimus-Eluting Bioresorbable Coronary Scaffolds: The ABSORB III Trial

Dean J Kereiakes et al. J Am Coll Cardiol. .
Free article

Abstract

Background: The Absorb everolimus-eluting poly-L-lactic acid-based bioresorbable vascular scaffold (BVS) provides early drug delivery and mechanical support functions similar to metallic drug-eluting stents (DES), followed by complete bioresorption in approximately 3 years with recovery of vascular structure and function. The ABSORB III trial demonstrated noninferior rates of target lesion failure (cardiac death, target vessel myocardial infarction [TVMI], or ischemia-driven target lesion revascularization) at 1 year in 2,008 patients with coronary artery disease randomized to BVS versus cobalt-chromium everolimus-eluting stents (EES).

Objectives: This study sought to assess clinical outcomes through 3 years following BVS implantation.

Methods: Clinical outcomes from the ABSORB III trial were analyzed by randomized treatment assignment cumulative through 3 years, and between 1 and 3 years.

Results: The primary composite endpoint of target lesion failure through 3 years occurred in 13.4% of BVS patients and 10.4% of EES patients (p = 0.06), and between 1 and 3 years in 7.0% versus 6.0% of patients, respectively (p = 0.39). TVMI through 3 years was increased with BVS (8.6% vs. 5.9%; p = 0.03), as was device thrombosis (2.3% vs. 0.7%; p = 0.01). In BVS-assigned patients, treatment of very small vessels (those with quantitatively determined reference vessel diameter <2.25 mm) was an independent predictor of 3-year TLF and scaffold thrombosis.

Conclusions: In the ABSORB III trial, 3-year adverse event rates were higher with BVS than EES, particularly TVMI and device thrombosis. Longer-term clinical follow-up is required to determine whether bioresorption of the polymeric scaffold will influence patient prognosis. (ABSORB III Randomized Controlled Trial [RCT] [ABSORB-III]; NCT01751906).

Keywords: bioresorbable scaffold; prognosis; randomized trial; stent.

PubMed Disclaimer

Comment in

Publication types

Associated data