3-Year Clinical Outcomes With Everolimus-Eluting Bioresorbable Coronary Scaffolds: The ABSORB III Trial
- PMID: 29100702
- DOI: 10.1016/j.jacc.2017.10.010
3-Year Clinical Outcomes With Everolimus-Eluting Bioresorbable Coronary Scaffolds: The ABSORB III Trial
Abstract
Background: The Absorb everolimus-eluting poly-L-lactic acid-based bioresorbable vascular scaffold (BVS) provides early drug delivery and mechanical support functions similar to metallic drug-eluting stents (DES), followed by complete bioresorption in approximately 3 years with recovery of vascular structure and function. The ABSORB III trial demonstrated noninferior rates of target lesion failure (cardiac death, target vessel myocardial infarction [TVMI], or ischemia-driven target lesion revascularization) at 1 year in 2,008 patients with coronary artery disease randomized to BVS versus cobalt-chromium everolimus-eluting stents (EES).
Objectives: This study sought to assess clinical outcomes through 3 years following BVS implantation.
Methods: Clinical outcomes from the ABSORB III trial were analyzed by randomized treatment assignment cumulative through 3 years, and between 1 and 3 years.
Results: The primary composite endpoint of target lesion failure through 3 years occurred in 13.4% of BVS patients and 10.4% of EES patients (p = 0.06), and between 1 and 3 years in 7.0% versus 6.0% of patients, respectively (p = 0.39). TVMI through 3 years was increased with BVS (8.6% vs. 5.9%; p = 0.03), as was device thrombosis (2.3% vs. 0.7%; p = 0.01). In BVS-assigned patients, treatment of very small vessels (those with quantitatively determined reference vessel diameter <2.25 mm) was an independent predictor of 3-year TLF and scaffold thrombosis.
Conclusions: In the ABSORB III trial, 3-year adverse event rates were higher with BVS than EES, particularly TVMI and device thrombosis. Longer-term clinical follow-up is required to determine whether bioresorption of the polymeric scaffold will influence patient prognosis. (ABSORB III Randomized Controlled Trial [RCT] [ABSORB-III]; NCT01751906).
Keywords: bioresorbable scaffold; prognosis; randomized trial; stent.
Copyright © 2017 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.
Comment in
-
Can the Vanishing Stent Reappear?: Fix the Technique, or Fix the Device?J Am Coll Cardiol. 2017 Dec 12;70(23):2875-2877. doi: 10.1016/j.jacc.2017.10.009. Epub 2017 Oct 31. J Am Coll Cardiol. 2017. PMID: 29100703 No abstract available.
-
Interventional cardiology: Operator technique is predictive of BVS-related adverse events.Nat Rev Cardiol. 2018 Feb;15(2):67. doi: 10.1038/nrcardio.2017.217. Epub 2017 Dec 21. Nat Rev Cardiol. 2018. PMID: 29263446 No abstract available.
Publication types
MeSH terms
Substances
Associated data
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical