Differences between values for plasma and serum in tests performed in the Ektachem 700 XR Analyzer, and evaluation of "plasma separator tubes (PST)"
- PMID: 2910557
Differences between values for plasma and serum in tests performed in the Ektachem 700 XR Analyzer, and evaluation of "plasma separator tubes (PST)"
Abstract
We measured 25 analytes in plasma and serum from the same blood specimen, using the Kodak Ektachem 700 XR Analyzer. For 22 of the analytes, values for plasma were practically the same as those for serum; for the other three, differences between concentrations in plasma and serum were significant, both statistically and medically. Values for the same analytes in plasma from regular heparinized tubes were essentially indistinguishable from those for plasma obtained by use of Plasma Separator Tubes (PST).
Similar articles
-
A comparison of stability of chemical analytes in plasma from the BD Vacutainer® Barricor™ tube with mechanical separator versus tubes containing gel separator.J Clin Lab Anal. 2020 Feb;34(2):e23060. doi: 10.1002/jcla.23060. Epub 2019 Oct 11. J Clin Lab Anal. 2020. PMID: 31605419 Free PMC article.
-
Comparison of Barricor™ vs. lithium heparin tubes for selected routine biochemical analytes and evaluation of post centrifugation stability.Biochem Med (Zagreb). 2018 Jun 15;28(2):020902. doi: 10.11613/BM.2018.020902. Epub 2018 Apr 15. Biochem Med (Zagreb). 2018. PMID: 29666561 Free PMC article.
-
Barricor blood collection tubes are equivalent to PST for a variety of chemistry and immunoassay analytes except for lactate dehydrogenase.Clin Chim Acta. 2019 Sep;496:18-24. doi: 10.1016/j.cca.2019.06.013. Epub 2019 Jun 12. Clin Chim Acta. 2019. PMID: 31201816
-
Evaluation of three different specimen types (serum, plasma lithium heparin and serum gel separator) for analysis of certain analytes: clinical significance of differences in results and efficiency in use.Clin Chem Lab Med. 2006;44(5):662-8. doi: 10.1515/CCLM.2006.099. Clin Chem Lab Med. 2006. PMID: 16681442
-
Stability of Routine Biochemical Analytes in Whole Blood and Plasma From Lithium Heparin Gel Tubes During 6-hr Storage.J Clin Lab Anal. 2016 Sep;30(5):602-9. doi: 10.1002/jcla.21909. Epub 2016 Feb 18. J Clin Lab Anal. 2016. PMID: 26891785 Free PMC article.
Cited by
-
Influence of plasma collection tubes on N-glycome in human blood samples.Pract Lab Med. 2024 Mar 2;39:e00383. doi: 10.1016/j.plabm.2024.e00383. eCollection 2024 Mar. Pract Lab Med. 2024. PMID: 38463195 Free PMC article.
-
Association between periodontitis and uric acid levels in blood and oral fluids: a systematic review and meta-analysis.BMC Oral Health. 2023 Mar 28;23(1):178. doi: 10.1186/s12903-023-02900-8. BMC Oral Health. 2023. PMID: 36973692 Free PMC article.
-
Improving home haemodialysis: Stability evaluation of routine clinical chemistry analytes in blood samples of haemodialysis patients.Biochem Med (Zagreb). 2019 Feb 15;29(1):010709. doi: 10.11613/BM.2019.010709. Biochem Med (Zagreb). 2019. PMID: 30799978 Free PMC article.
-
Comparative study of chemical pathology sample collection tubes at the largest hospital in South Africa.J Med Biochem. 2021 Sep 3;40(4):358-366. doi: 10.5937/jomb0-27216. J Med Biochem. 2021. PMID: 34616225 Free PMC article.
-
Impact of the ovarian cycle and pregnancy on plasma chemistry values in ewes.J Vet Diagn Invest. 2018 Mar;30(2):238-244. doi: 10.1177/1040638717752217. Epub 2018 Jan 2. J Vet Diagn Invest. 2018. PMID: 29291683 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Miscellaneous