Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2017 Nov;10(11):e006530.
doi: 10.1161/CIRCIMAGING.117.006530.

Head-to-Head Comparison of Global and Regional Two-Dimensional Speckle Tracking Strain Versus Cardiac Magnetic Resonance Tagging in a Multicenter Validation Study

Affiliations
Comparative Study

Head-to-Head Comparison of Global and Regional Two-Dimensional Speckle Tracking Strain Versus Cardiac Magnetic Resonance Tagging in a Multicenter Validation Study

Mihaela Silvia Amzulescu et al. Circ Cardiovasc Imaging. 2017 Nov.

Abstract

Background: Despite widespread use to characterize and refine prognosis, validation data of two-dimensional (2D) speckle tracking (2DST) echocardiography myocardial strain measurement remain scarce.

Methods and results: Global and regional subendocardial peak-systolic Lagrangian longitudinal (LS) and circumferential strain (CS) by 2DST and 2D-tagged (2DTagg) cardiac magnetic resonance imaging were compared against sonomicrometry in a dynamic heart phantom and among each other in 136 patients included prospectively at 2 centers. The ability of regional LS and CS 2DST and 2DTagg to identify late gadolinium enhancement was compared using receiver operating characteristics curves. In vitro, both LS-2DST and 2DTagg highly agreed with sonomicrometry (intraclass correlation coefficient [ICC], 0.89 and ICC, 0.90, both P<0.001 with -3±2.8% and 0.34±4.35% bias, respectively). In patients, both global LS and global CS 2DST agreed well with 2DTagg (ICC, 0.89 and ICC, 0.80; P<0.001); however, they provided systematically greater values (relative bias of -37±27% and -25±37% for global LS and global CS, respectively). On regional basis, however, ICC (from 0.17 to 0.81) and relative bias (from -9 to -98%) between 2DST and 2DTagg varied strongly among segments. Ability to discriminate infarcted versus noninfarcted segments by late gadolinium enhancement was similarly good for regional LS 2DTagg and 2DST (area under the curve, 0.66 versus 0.59; P=0.08), while it was lower for CS 2DST than 2DTagg (area under the curve, 0.61 versus 0.75; P<0.001).

Conclusions: The high accuracy against sonomicrometry and good agreement of global LS and global CS by 2DST and 2DTagg confirm the overall validity of 2DST strain measurement. Yet, higher intertechnique segmental variability and lower ability for detecting infarct suggest that 2DST strain estimates may be less performant on regional than on global basis.

Keywords: 2-dimensional; echocardiography; magnetic resonance imaging; software validation; speckle tracking; strain.

PubMed Disclaimer

Comment in

  • Myocardial Strain: Still a Long Way to Go.
    Reichek N. Reichek N. Circ Cardiovasc Imaging. 2017 Nov;10(11):e007145. doi: 10.1161/CIRCIMAGING.117.007145. Circ Cardiovasc Imaging. 2017. PMID: 29138231 No abstract available.