Evaluation of the Implantable Doppler Probe for Free Flap Monitoring in Lower Limb Reconstruction
- PMID: 29179224
- DOI: 10.1055/s-0037-1608628
Evaluation of the Implantable Doppler Probe for Free Flap Monitoring in Lower Limb Reconstruction
Abstract
Background: Timely reexploration and reanastomoses can salvage failing free flaps. The use of the implantable Doppler probe provides direct evidence of vascular impairment of the microvascular anastomoses and allows for postoperative NPWT. The aim of this retrospective study was to compare the Doppler probe to conventional monitoring techniques for free flap monitoring in lower limb reconstruction and to identify risk factors for perfusion disturbance and reexploration.
Methods: All patients receiving free muscle flap reconstruction for lower limb soft tissue defects at our department from 2000 to 2013 were included, and all adverse events, timely detection of perfusion problems, and outcome of revision surgery were assessed by chart analysis.
Results: For lower limb reconstruction, 110 free muscle transfers were performed of which 41 muscle flaps were conventionally monitored and 69 flaps were monitored using the implantable Doppler probe. In 18 cases, the free muscle flaps needed revision because of perfusion disturbances. The salvage rate was 80% with monitoring by the implantable Doppler probe compared with 62.5% using conventional monitoring methods resulting in success rates of 95.7 and 92.7%, respectively.
Conclusion: The use of the implantable Cook-Swartz Doppler probe represents a safe monitoring method for lower limb reconstruction, which allows for the additional use of NPWT. Higher salvage and revision success rates can be attributed to an earlier detection of perfusion impairment. However, a larger patient cohort is necessary to verify superiority over conventional postoperative monitoring.
Thieme Medical Publishers 333 Seventh Avenue, New York, NY 10001, USA.
Conflict of interest statement
None.
Similar articles
-
Postoperative monitoring of lower limb free flaps with the Cook-Swartz implantable Doppler probe: A clinical trial.Microsurgery. 2010 Jul;30(5):354-60. doi: 10.1002/micr.20720. Microsurgery. 2010. PMID: 19967762 Clinical Trial.
-
Rationale for the use of the implantable Doppler probe based on 7 years' experience.Br J Oral Maxillofac Surg. 2014 Jul;52(6):530-4. doi: 10.1016/j.bjoms.2014.03.014. Epub 2014 Apr 8. Br J Oral Maxillofac Surg. 2014. PMID: 24721166
-
The role of the implantable Doppler probe in free flap surgery.Laryngoscope. 2014 Mar;124 Suppl 1:S1-12. doi: 10.1002/lary.24569. Laryngoscope. 2014. PMID: 24375425
-
Comparison of arterial and venous implantable Doppler postoperative monitoring of free flaps: Systematic review and meta-analysis of diagnostic test accuracy.Microsurgery. 2020 May;40(4):501-511. doi: 10.1002/micr.30564. Epub 2020 Feb 7. Microsurgery. 2020. PMID: 32031735
-
Adjunctive technologies in postoperative free-flap monitoring: a systematic review.J Plast Reconstr Aesthet Surg. 2023 Dec;87:147-155. doi: 10.1016/j.bjps.2023.09.015. Epub 2023 Sep 11. J Plast Reconstr Aesthet Surg. 2023. PMID: 37844383
Cited by
-
A Multidisciplinary Approach to End-Stage Limb Salvage in the Highly Comorbid Atraumatic Population: An Observational Study.J Clin Med. 2024 Apr 20;13(8):2406. doi: 10.3390/jcm13082406. J Clin Med. 2024. PMID: 38673679 Free PMC article.
-
A comparative analysis using flowmeter, laser-Doppler |spectrophotometry, and indocyanine green-videoangiography for detection of vascular stenosis in free flaps.Sci Rep. 2020 Jan 22;10(1):939. doi: 10.1038/s41598-020-57777-2. Sci Rep. 2020. PMID: 31969630 Free PMC article.
-
Intraoperative Utility of the Implantable Doppler in Lower Extremity Reconstruction: A Matched Case-control Study.Plast Reconstr Surg Glob Open. 2020 Nov 25;8(11):e3229. doi: 10.1097/GOX.0000000000003229. eCollection 2020 Nov. Plast Reconstr Surg Glob Open. 2020. PMID: 33299699 Free PMC article.
-
In Situ Monitoring and Assessment of Ischemic Skin Flap by High-Frequency Ultrasound and Quantitative Parameters.Sensors (Basel). 2024 Jan 7;24(2):363. doi: 10.3390/s24020363. Sensors (Basel). 2024. PMID: 38257456 Free PMC article.
-
[A systematic approach to plastic surgical foot reconstruction].Unfallchirurg. 2021 Oct;124(10):797-806. doi: 10.1007/s00113-021-01075-7. Epub 2021 Sep 9. Unfallchirurg. 2021. PMID: 34499184 Review. German.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical