Devices for preventing percutaneous exposure injuries caused by needles in healthcare personnel
- PMID: 29190036
- PMCID: PMC6491125
- DOI: 10.1002/14651858.CD009740.pub3
Devices for preventing percutaneous exposure injuries caused by needles in healthcare personnel
Abstract
Background: Percutaneous exposure injuries from devices used for blood collection or for injections expose healthcare workers to the risk of blood borne infections such as hepatitis B and C, and human immunodeficiency virus (HIV). Safety features such as shields or retractable needles can possibly contribute to the prevention of these injuries and it is important to evaluate their effectiveness.
Objectives: To determine the benefits and harms of safety medical devices aiming to prevent percutaneous exposure injuries caused by needles in healthcare personnel versus no intervention or alternative interventions.
Search methods: We searched CENTRAL, MEDLINE, EMBASE, NHSEED, Science Citation Index Expanded, CINAHL, Nioshtic, CISdoc and PsycINFO (until 11 November 2016).
Selection criteria: We included randomised controlled trials (RCT), controlled before and after studies (CBA) and interrupted time-series (ITS) designs of the effect of safety engineered medical devices on percutaneous exposure injuries in healthcare staff.
Data collection and analysis: Two of the authors independently assessed study eligibility and risk of bias and extracted data. We synthesized study results with a fixed-effect or random-effects model meta-analysis where appropriate.
Main results: We included six RCTs with 1838 participants, two cluster-RCTs with 795 participants and 73,454 patient days, five CBAs with approximately 22,000 participants and eleven ITS with an average of 13.8 data points. These studies evaluated safe modifications of blood collection systems, intravenous (IV) systems, injection systems, multiple devices, sharps containers and legislation on the implementation of safe devices. We estimated the needlestick injury (NSI) rate in the control groups to be about one to five NSIs per 1000 person-years. There were only two studies from low- or middle-income countries. The risk of bias was high in 20 of 24 studies. Safe blood collection systems:We found one RCT that found a safety engineered blood gas syringe having no considerable effect on NSIs (Relative Risk (RR) 0.2, 95% Confidence Interval (95% CI) 0.01 to 4.14, 550 patients, very low quality evidence). In one ITS study, safe blood collection systems decreased NSIs immediately after the introduction (effect size (ES) -6.9, 95% CI -9.5 to -4.2) but there was no further decrease over time (ES -1.2, 95% CI -2.5 to 0.1, very low quality evidence). Another ITS study evaluated an outdated recapping shield, which we did not consider further. Safe Intravenous systemsThere was very low quality evidence in two ITS studies that NSIs were reduced with the introduction of safe IV devices, whereas one RCT and one CBA study provided very low quality evidence of no effect. However, there was moderate quality evidence produced by four other RCT studies that these devices increased the number of blood splashes when the safety system had to be engaged actively (relative risk (RR) 1.6, 95% CI 1.08 to 2.36). In contrast there was low quality evidence produced by two RCTs of passive systems that showed no effect on blood splashes. Yet another RCT produced low quality evidence that a different safe active IV system also decreased the incidence of blood leakages. Safe injection devicesThere was very low quality evidence provided by one RCT and one CBA study showing that introduction of safe injection devices did not considerably change the NSI rate. One ITS study produced low quality evidence showing that the introduction of safe passive injection systems had no effect on NSI rate when compared to safe active injection systems. Multiple safe devicesThere was very low quality evidence from one CBA study and two ITS studies. According to the CBA study, the introduction of multiple safe devices resulted in a decrease in NSI,whereas the two ITS studies found no change. Safety containersOne CBA study produced very low quality evidence showing that the introduction of safety containers decreased NSI. However, two ITS studies evaluating the same intervention found inconsistent results. LegislationThere was low to moderate quality evidence in two ITS studies that introduction of legislation on the use of safety-engineered devices reduced the rate of NSIs among healthcare workers. There was also low quality evidence which showed a decrease in the trend over time for NSI rates.Twenty out of 24 studies had a high risk of bias and the lack of evidence of a beneficial effect could be due to both confounding and bias. This does not mean that these devices are not effective.
Authors' conclusions: For safe blood collection systems, we found very low quality evidence of inconsistent effects on NSIs. For safe passive intravenous systems, we found very low quality evidence of a decrease in NSI and a reduction in the incidence of blood leakage events but moderate quality evidence that active systems may increase exposure to blood. For safe injection needles, the introduction of multiple safety devices or the introduction of sharps containers the evidence was inconsistent or there was no clear evidence of a benefit. There was low to moderate quality evidence that introduction of legislation probably reduces NSI rates.More high-quality cluster-randomised controlled studies that include cost-effectiveness measures are needed, especially in countries where both NSIs and blood-borne infections are highly prevalent.
Conflict of interest statement
Viraj Reddy: None known.
Marie‐Claude Lavoie: None known.
Jos Verbeek: None known.
Manisha Pahwa: None known.
Figures





























Update of
-
Devices for preventing percutaneous exposure injuries caused by needles in healthcare personnel.Cochrane Database Syst Rev. 2014 Mar 9;(3):CD009740. doi: 10.1002/14651858.CD009740.pub2. Cochrane Database Syst Rev. 2014. Update in: Cochrane Database Syst Rev. 2017 Nov 14;11:CD009740. doi: 10.1002/14651858.CD009740.pub3. PMID: 24610008 Updated.
References
References to studies included in this review
Asai 1999 active {published data only}
-
- Asai T, Matsumoto S, Matsumoto H, Yamamoto K, Shingu K. Prevention of needlestick injury efficacy of a safeguarded intravenous cannula. Anaesthesia 1999;54(3):258‐61. - PubMed
Asai 2002 active {published data only}
-
- Asai T, Hidaka I, Kawashima A, Miki T, Inada K, Kawachi S. Efficacy of catheter needles with safeguard mechanisms. Anaesthesia 2002;57(6):572‐7. - PubMed
Asai 2002 passive {published data only}
-
- Asai T, Hidaka I, Kawashima A, Miki T, Inada K, Kawachi S. Efficacy of catheter needles with safeguard mechanisms. Anaesthesia 2002;57(6):572‐7. - PubMed
Azar‐Cavanagh 2007 {published data only}
-
- Azar‐Cavanagh M, Burdt P, Green‐McKenzie J. Effect of the introduction of an engineered sharps injury prevention device on the percutaneous injury rate in healthcare workers. Infection Control and Hospital Epidemiology 2007;28(2):165‐70. - PubMed
Baskin 2014 {published data only}
-
- Baskin SB, Oray NC, Yanturali S, Bayram B. The comparison of heparinized insulin syringes and safety‐engineered blood gas syringes used in arterial blood gas sampling in the ED setting (randomized controlled study). American Journal of Emergency Medicine 2014;32(5):432‐7. - PubMed
Chambers 2015 hospitals {published data only}
Chambers 2015 long‐term nursing care {published data only}
Cote 2003 {published data only}
-
- Cote CJ, Roth AG, Wheeler M, ter Rahe C, Rae BR, Dsida RM, et al. Traditional versus new needle retractable i.v. catheters in children: are they really safer, and whom are they protecting?. Anesthesia and Analgesia 2003;96(2):387‐91, table of contents. - PubMed
Edmond 1988 {published data only}
-
- Edmond M, Khakoo R, McTaggart B, Solomon R. Effect of bedside needle disposal units on needle recapping frequency and needlestick injury. Infection Control and Hospital Epidemiology 1988;9(3):114‐6. - PubMed
Gaballah 2012 {published data only}
Goldwater 1989 {published data only}
-
- Goldwater PN, Law R, Nixon AD, Officer JA, Cleland JF. Impact of a recapping device on venepuncture‐related needlestick injury. Infection Control and Hospital Epidemiology 1989;10(1):21‐5. - PubMed
Goris 2015 {published data only}
-
- Goris AL, Gemeinhart N, Babcock HM. Reducing needlestick injuries from active safety devices:a passive safety engineered conversion. American Journal of Infection Control 2015;43:S9‐S10.
Grimmond 2010 {published data only}
-
- Grimmond T, Bylund S, Anglea C, Beeke L, Callahan A, Christiansen E, et al. Sharps injury reduction using a sharps container with enhanced engineering: A 28 hospital non‐randomized intervention and cohort study. American Journal of Infection Control 2010;38(10):799‐805. - PubMed
L'Ecuyer 1996 2wva {published data only}
-
- L'Ecuyer PB, Schwab EO, lademarco E, Barr N, Aton EA, Fraser VJ. Randomized prospective study of the impact of three needleless intravenous systems on needlestick injury rates. Infection Control and Hospital Epidemiology 1996;17:803‐8. - PubMed
L'Ecuyer 1996 mbc {published data only}
-
- L'Ecuyer PB, Schwab EO, Iademarco E, Barr N, Aton EA, Fraser VJ. Randomized prospective study of the impact of three needleless intravenous systems on needlestick injury rates. Infection Control and Hospital Epidemiology 1996;17:803‐8. - PubMed
L'Ecuyer 1996 pbc {published data only}
-
- L'Ecuyer PB, Schwab EO, Iademarco E, Barr N, Aton EA, Fraser VJ. Randomized prospective study of the impact of three needleless intravenous systems on needlestick injury rates. Infection Control and Hospital Epidemiology 1996;17:803‐8. - PubMed
Mendelson 1998 {published data only}
-
- Mendelson MH, Short LJ, Schechter CB, Meyers BR, Rodriguez M, Cohen S, et al. Study of a needleless intermittent intravenous‐access system for peripheral infusions: analysis of staff, patient, and institutional outcomes. Infection Control and Hospital Epidemiology 1998;19(6):401‐6. - PubMed
Phillips 2013 {published data only}
-
- Phillips EK, Conaway M, Parker G, Perry J, Jagger J. Issues in understanding the impact of the Needlestick Safety and Prevention Act on hospital sharps injuries. Infection Control and Hospital Epidemiology 2013;34(9):935‐939. - PubMed
Prunet 2008 active {published data only}
-
- Prunet B, Meaudre E, Montcriol A, Asencio Y, Bordes J, Lacroix G, et al. A prospective randomized trial of two safety peripheral intravenous catheters. Anesthesia and Analgesia 2008;107(1):155. - PubMed
Prunet 2008 passive {published data only}
-
- Prunet B, Meaudre E, Montcriol A, Asencio Y, Bordes J, Lacroix G, et al. A prospective randomized trial of two safety peripheral intravenous catheters. Anesthesia and Analgesia 2008;107(1):155. - PubMed
Reddy 2001 {published data only}
-
- Reddy SG, Emery RJ. Assessing the effect of long‐term availability of engineering controls on needlestick injuries among health care workers: a 3‐year preimplementation and postimplementation comparison. American Journal of Infection Control 2001;29(6):425‐7. - PubMed
Richard 2001 {published data only}
-
- Richard VS, Kenneth J, Ramaprabha P, Kirupakaran H, Chandy GM. Impact of introduction of sharps containers and of education programmes on the pattern of needle stick injuries in a tertiary care centre in India. Journal of Hospital Infection 2001;47(2):163‐5. - PubMed
Rogues 2004 {published data only}
-
- Rogues AM, Verdun‐Esquer C, Buisson‐Valles I, Laville MF, Lashéras A, Sarrat A, et al. Impact of safety devices for preventing percutaneous injuries related to phlebotomy procedures in health care workers. American Journal of Infection Control 2004;32(8):441‐4. - PubMed
Seiberlich 2016 {published data only}
-
- Seiberlich LE, Keay V, Kallos S, Junghans T, Lang E, McRae AD. Clinical performance of a new blood control peripheral intravenous catheter: A prospective, randomized, controlled study. International Emergency Nursing 2016;25:59‐64. - PubMed
Sossai 2010 {published data only}
-
- Sossai D, Puro V, Chiappatoli L, Dagnino G, Odone B, Polimeri A, et al. Using an intravenous catheter system to prevent needlestick injury. Nursing Standard (Royal College of Nursing (Great Britain): 1987) 2010;24(29):42‐6. - PubMed
Valls 2007 {published data only}
-
- Valls V, Lozano MS, Yanez R, Martinez MJ, Pascual F, Lloret J, et al. Use of safety devices and the prevention of percutaneous injuries among healthcare workers. Infection Control and Hospital Epidemiology 2007;28(12):1352‐60. - PubMed
van der Molen 2011 {published data only}
-
- Molen HF, Zwinderman KAH, Sluiter JK, Frings‐Dresen MHW. Better effect of the use of a needle safety device in combination with an interactive workshop to prevent needle stick injuries. Safety Science 2011;49:1180‐6.
Whitby 2008 {published data only}
-
- Whitby M, McLaws ML, Slater K. Needlestick injuries in a major teaching hospital: the worthwhile effect of hospital‐wide replacement of conventional hollow‐bore needles. American Journal of Infection Control 2008;36(3):180‐6. - PubMed
Zakrzewska 2001 {published data only}
-
- Zakrzewska JM, Greenwood I, Jackson J. Cross‐infection control: Introducing safety syringes into a UK dental school ‐ a controlled study. British Dental Journal 2001;190(2):88‐92. - PubMed
References to studies excluded from this review
Beynon 2015 {published data only}
-
- Beynon A. A quality improvement initiative to reduce needlestick injuries. Nursing Standard 2015;29(22):37‐42. - PubMed
Bowden 1993 {published data only}
-
- Bowden FJ, Pollett B, Birrell F, Dax EM. Occupational exposure to the human immunodeficiency virus and other blood‐borne pathogens. A six‐year prospective study. Medical journal of Australia 1993;158(12):810‐2. - PubMed
Buswell 2014 {published data only}
-
- Buswell ML, Hourigan M, Nault A, Bender J. Needlestick injuries in livestock workers and prevention programs. 2013 Agricultural Safety Summit. Journal of Agromedicine 2014;19(2):206‐7.
Carvalho 2016 {published data only}
-
- Carvalho PCF, Reis RK, Pereira FMV, Toffano SEM. Injury rates from peripheral catheters with or without safety devices in a Brazilian public hospital. American Journal of Infection Control 2016;44(7):853‐4. - PubMed
Chaillol 2010 {published data only}
-
- Chaillol I, Ecochard R, Denis MA, Iwaz J, Khoueiry P, Bergeret A. Fast and specific detection of moderate long‐term changes in occupational blood exposures. Occupational and Environmental Medicine 2010;67(11):785‐91. - PubMed
Chakravarthy 2014 {published data only}
-
- Chakravarthy M, Singh S, Arora A, Sengupta S, Munshi N, Rangaswamy S, et al. Epidemiology of sharp injuries ‐ Prospective EPINet data from five tertiary care hospitals in India ‐ Data for 144 cumulated months, 1.5 million inpatient days. Clinical Epidemiology and Global Health 2014;2(3):121‐6.
Cleveland 2007 {published data only}
-
- Cleveland JL, Barker LK, Cuny EJ, Panlilio AL. Preventing percutaneous injuries among dental health care personnel. Journal of the American Dental Association 2007;138(2):169‐78; quiz 247‐8. - PubMed
Cullen 2006 {published data only}
-
- Cullen BL, Genasi F, Symington I, Bagg J, McCreaddie M, Taylor A, et al. Potential for reported needlestick injury prevention among healthcare workers through safety device usage and improvement of guideline adherence: expert panel assessment. The Journal of Hospital Infection 2006;63(4):445‐51. - PubMed
Di Bari 2015 {published data only}
-
- Bari V, Carli G, Puro V. [Prevention of accidental needle sticks before the Directive 2010/32/EU in a sample of Italian hospitals]. La Medicina del Lavoro 2015;106(3):186‐205. - PubMed
Floret 2015 {published data only}
-
- Floret N, Ali‐Brandmeyer O, L'Hériteau F, Bervas C, Barquins‐Guichard S, Pelissier G, et al. Sharp decrease of reported occupational blood and body fluid exposures in French hospitals, 2003‐2012: Results of the French National Network Survey, AES‐RAISIN. Infection Control and Hospital Epidemiology 2015;36(8):963‐8. - PubMed
Ford 2011 {published data only}
-
- Ford J, Phillips P. An evaluation of sharp safety blood evacuation devices. Nursing Standard (Royal College of Nursing (Great Britain): 1987) 2011;25(43):41‐7. - PubMed
Fukuda 2016 {published data only}
-
- Fukuda H, Yamanaka N. Reducing needlestick injuries through safety‐engineered devices: results of a Japanese multi‐centre study. J Hosp Infect 2016;92(2):147‐53. - PubMed
Goossens 2011 {published data only}
-
- Goossens GA, Moons P, Jerome M, Stas M. Prospective clinical evaluation of the Polyperf(R) Safe, a safety Huber needle, in cancer patients. The Journal of Vascular Access 2011;12(3):200‐6. - PubMed
Gramling 2013 {published data only}
-
- Gramling JJ, Nachreiner N. Implementing a sharps injury reduction program at a charity hospital in India. Workplace Health and Safety 2013;61(8):339‐45. - PubMed
Grimmond 2014 {published data only}
-
- Grimmond T. Frequency of use and activation of safety‐engineered sharps devices: A sharps container audit in five Australian capital cities. Healthcare Infection 2014;19(3):95‐100.
Guerlain 2010 {published data only}
Hotaling 2009 {published data only}
-
- Hotaling M. A retractable winged steel (butterfly) needle performance improvement project. Joint Commission Journal on Quality and Patient Safety / Joint Commission Resources 2009;35(2):100‐5, 61. - PubMed
Iinuma 2005 {published data only}
-
- Iinuma Y, Igawa J, Takeshita M, Hashimoto Y, Fujihara N, Saito T, et al. Passive safety devices are more effective at reducing needlestick injuries. The Journal of Hospital Infection 2005;61(4):360‐1. - PubMed
Jagger 2010 {published data only}
-
- Jagger J, Berguer R, Phillips E K, Parker G, Gomaa A E. Increase in sharps injuries in surgical settings versus nonsurgical settings after passage of national needlestick legislation. JOURNAL OF THE AMERICAN COLLEGE OF SURGEONS 2010;210:(4):496‐502. - PubMed
Kanamori 2016 {published data only}
-
- Kanamori H, Weber DJ, DiBiase LM, Pitman KL, Consoli SA, Hill J, et al. Impact of safety‐engineered devices on the Incidence of occupational blood and body fluid exposures among healthcare personnel in an academic facility, 2000‐2014. Infection Control and Hospital Epidemiology 2016;37(5):497‐504. - PubMed
Kim 2015 {published data only}
-
- Kim YG, Jeong IS, Park SM. Sharps injury prevention guidance among health care professionals: A comparison between self‐reported and observed compliance. American Journal of Infection Control 2015;43(9):977‐82. - PubMed
Lamontagne 2007 {published data only}
-
- Lamontagne F, Abiteboul D, Lolom I, Pellissier G, Tarantola A, Descamps JM, et al. Role of safety‐engineered devices in preventing needlestick injuries in 32 French hospitals. Infection Control and Hospital Epidemiology 2007;28(1):18‐23. - PubMed
Laramie 2011 {published data only}
-
- Laramie AK, Pun VC, Fang SC, Kriebel D, Davis L. Sharps injuries among employees of acute care hospitals in Massachusetts, 2002‐2007. Infection Control and Hospital Epidemiology 2011;32(6):538‐44. - PubMed
Lauer 2014 {published data only}
-
- Lauer A‐C, Reddemann A, Meier‐Wronski C‐P, Bias H, G̦decke K, Arendt M, et al. Needlestick and sharps injuries among medical undergraduate students. American Journal of Infection Control 2014;42(3):235‐9. - PubMed
Lipscomb 2010 {published data only}
-
- Lipscomb J, Geiger Brown J, Johnson J, McPhaul K, Trinkoff A, Storr C. Blood exposure and primary prevention in the home care workplace. NIOSH report. Cincinatti: National Institute of Occupational Safety and Health, 2010.
Lu 2015 {published data only}
-
- Lu Y, Senthilselvan A, Joffe A M, Beach J. Effectiveness of safety‐engineered devices in reducing sharp object injuries. Occupational medicine (Oxford, England) 2015;65(1):39‐44. - PubMed
Markkanen 2015 {published data only}
Massachusetts 2011 {published data only}
-
- Massachusetts Department of Public Health. Sharps injuries among hospital workers in Massachusetts. Massachusetts Department of Public Health: Occupational Health Surveillance Program 2011.
McAllister 2014 {published data only}
-
- McAllister L, Anderson J, Werth K, Cho I, Copeland K, Cam Bouveret N, et al. Needle‐free jet injection for administration of influenza vaccine: A randomised non‐inferiority trial. The Lancet 2014;384(9944):674‐81. - PubMed
Menezes 2014 {published data only}
-
- Menezes JA, Bandeira CS, Quintana M, Lima E Silva JC, Calvet GA, Brasil P. Impact of a single safety‐engineered device on the occurrence of percutaneous injuries in a general hospital in Brazil. American Journal of Infection Control 2014;42(2):174‐7. - PubMed
Montella 2014 {published data only}
-
- Montella E, Schiavone D, Apicella L, Silverio P, Gaudiosi M, Ambrosone E, et al. Cost‐benefit evaluation of a preventive intervention on the biological risk in health: the accidental puncture during the administration of insulin in the University Hospital "Federico II" of Naples. Annali di Igiene: Medicina Preventiva e di Comunità 2014;26(3):272‐8. - PubMed
Neo 2016 {published data only}
-
- Neo SHS, Khemlani MH, Sim LK, Seah AST. Winged metal needles versus plastic winged and nonwinged cannulae for subcutaneous infusions in palliative care: A quality improvement project to enhance patient care and medical staff safety in a Singaporean hospital. Journal of Palliative Medicine 2016;19(3):318‐22. - PubMed
Perry 2012a {published data only}
-
- Perry J, Jagger J, Parker G, Phillips E K, Gomaa A. Disposal of sharps medical waste in the United States: impact of recommendations and regulations, 1987‐2007. American Journal of Infection Control 2012;40(4):354‐8. - PubMed
Pigman 1993 {published data only}
-
- Pigman EC, Karch DB, Scott JL. Splatter during jet irrigation cleansing of a wound model: a comparison of three inexpensive devices. Annals of Emergency Medicine 1993;22(10):1563‐7. - PubMed
Rajkumari 2015 {published data only}
Roff 2014 {published data only}
-
- Roff M, Basu S, Adisesh A. Do active safety‐needle devices cause spatter contamination?. Journal of Hospital Infection 2014;86(3):221‐3. - PubMed
Shimatani 2011 {published data only}
-
- Shimatani M, Matsui Y, Yano K. Comparison of the needlstick injuries due to active and passive design safety intravenous catheters. American Journal of Infection Control 2011;39(5):E79.
Sibbitt 2011 {published data only}
-
- Sibbitt WL Jr, Band PA, Kettwich LG, Sibbitt CR, Sibbitt LJ, Bankhurst AD. Safety syringes and anti‐needlestick devices in orthopaedic surgery. The Journal of Bone and Joint Surgery. American volume 2011;93(17):1641‐9. - PubMed
Skolnick 1993 {published data only}
-
- Skolnick R, LaRocca J, Barba D, Paicius L. Evaluation and implementation of a needleless intravenous system: making needlesticks a needless problem. American Journal of Infection Control 1993;21(1):39‐41. - PubMed
Smith 2013 {published data only}
-
- Smith SL, Shames A, Jeannerot F. Usability and validation of a new device for the administration of somatostatin analog therapy: An open‐label, randomized study using a simulated placebo gel. Endocrine Reviews 2013;34(suppl 4):na.
Sossai 2016 {published data only}
Steuten 2010 {published data only}
-
- Steuten L, Buxton M. Economic evaluation of healthcare safety: which attributes of safety do healthcare professionals consider most important in resource allocation decisions?. Quality and Safety in Health Care 2010;19(5):e6. - PubMed
Tosini 2010 {published data only}
-
- Tosini W, Ciotti C, Goyer F, Lolom I, L'Heriteau F, Abiteboul D, et al. Needlestick injury rates according to different types of safety‐engineered devices: results of a French multicenter study. Infection Control and Hospital Epidemiology 2010;31(4):402‐7. - PubMed
Unahalekhaka 2015 {published data only}
-
- Unahalekhaka A, Lueang‐a‐papong S. Prevention of needlestick and sharp injuries among hospitals in Thailand. 42nd annual conference abstracts, APIC 2015, Nashville, TN June 2015. American Journal of Infection Control 2015;43:S44‐5.
References to studies awaiting assessment
Ferrario 2012 {published data only}
-
- Ferrario MM, Landone S, Biasi M, Tagliasacchi R, Riva R, Veronesi G, et al. Time trends of incidence rates of work accident with blood contamination in a North Italian teaching hospital [Analisi dei trend temporali (2004‐11) dei tassi di infortunio biologico in un ospedale universitario del nord Italia. Quali evidenze di efficacia dei sistemi di prelievo a vuoto?]. Giornale Italiano di Medicina del Lavoro ed Ergonomia 2012;34(3 suppl 1):275‐7. - PubMed
Perry 2012 {published data only}
-
- Perry J, Jagger J, Parker G, Philips EK, Gomaa A. Disposal of sharps medical waste in the United States: impact of recommendations and regulations, 1987‐2007. American Journal of Infection Control 2012;40(4):354‐8. - PubMed
Phillips 2010 {published data only}
-
- Phillips EK. Healthcare worker injury risk and the impact of the Needlestick Safety and Prevention Act. 10th Anniversary of the Needlestick Safety and Prevention Act: Mapping Progress, Charting a Future Path, Charlottesville, VA, November 5‐6, 2010. Charlotsville, VA: University of Virginia School of Medicine, 2010 Nov; :1‐14. 2010.
Phillips 2011 {published data only}
-
- Phillips EK, Conaway M, Parker G, Perry J, Jagger J. Changes in sharps injuries among healthcare workers: the effect of HR 5178 (National Needlestick Safety and Prevention Act). NOIRS 2011‐Abstracts of the national occupational injury research symposium, october 18‐20, 2011, Morgantown, West Virginia. National Institute for Occupational Safety and Health, Oct; 130. 2011.
Phillips 2012 {published data only}
-
- Phillips EK, Conaway MR, Jagger JC. Percutaneous injuries before and after the needlestick safety and prevention act. New England Journal of Medicine 2012;366(7):670‐1. - PubMed
Phillips 2012a {published data only}
-
- Phillips E K. Impact of Needlestick Safety & Prevention Act (HR5178) on hospital worker injury. NIOSH report K01‐OH‐009140 2012:1‐33.
Uyen 2014 {published data only}
-
- Uyen Vu. The effective implementation of health and safety regulations: lessons learned from an iwh study on safety‐engineered needles. OOHNA Journal 2014;33(2):24‐6.
Additional references
Akduman 1999
-
- Akduman D, Kim LE, Parks RL, L'Ecuyer PB, Mutha S, Jeffe DB, et al. Use of personal protective equipment and operating room behaviors in four surgical subspecialties: personal protective equipment and behaviors in surgery. Infection Control and Hospital Epidemiology 1999;20(2):110‐4. - PubMed
Ansa 2002
-
- Ansa VO, Udoma EJ, Umoh MS, Anah MU. Occupational risk of infection by human immunodeficiency and hepatitis B viruses among health workers in south‐eastern Nigeria. East African Medical Journal 2002;79(5):254‐6. - PubMed
Ballout 2016
Bryce 1999
-
- Bryce EA, Ford J, Chase L, Taylor C, Scharf S. Sharps injuries: defining prevention priorities. American Journal of Infection Control 1999;27(5):447‐52. - PubMed
Castella 2003
-
- Castella A, Vallino A, Argentero PA, Zotti CM. Preventability of percutaneous injuries in healthcare workers: a year‐long survey in Italy. Journal of Hospital Infection 2003;55(4):290‐4. - PubMed
Cheetham 2016
Chen 2005
Clarke 2002
-
- Clarke SP, Rockett JL, Sloane DM, Aiken LH. Organizational climate, staffing, and safety equipment as predictors of needlestick injuries and near‐misses in hospital nurses. American Journal of Infection Control 2002;30(4):207‐16. - PubMed
De Carli 2003
-
- Carli G, Puro V, Ippolito G, Studio Italiano Rischio Occupazionale da HIV Group. Risk of hepatitis C virus transmission following percutaneous exposure in healthcare workers. Infection 2003;31 Suppl 2:22‐7. - PubMed
Doebbeling 2003
-
- Doebbeling BN, Vaughn TE, McCoy KD, Beekmann SE, Woolson RF, Ferguson KJ, et al. Percutaneous injury, blood exposure, and adherence to standard precautions: are hospital‐based health care providers still at risk?. Clinical Infectious Diseases 2003;37(8):1006‐13. - PubMed
Downs 1998
EPOC 2006
-
- The Cochrane Effective Practice and Organisation of Care Group (EPOC). Including interrupted time‐series (ITS) designs in a EPOC review. http://www.epoc.uottawa.ca/inttime.pdf (accessed 5 September 2006).
Fisman 2002
-
- Fisman DN, Mittleman AM, Sorock GS, Harris AD. Willingness to pay to avoid sharps‐related injuries: a study in injured health care workers. American Journal of Infection Control 2002;30(5):283‐7. - PubMed
Fisman 2007
-
- Fisman DN, Harris AD, Rubin M, Sorock GS, Mittleman AM. Fatigue increases the risk of injury from sharp devices in medical trainees: results from a case‐crossover study. Infection Control and Hospital Epidemiology 2007;28(1):10‐7. - PubMed
GRADEpro 2017 [Computer program]
-
- Brozek J, Oxman A, Schünemann H. GRADEpro. Version 3.2 for Windows. GRADE working group, 2017.
Grimmond 2017
-
- Grimmond T, Good L. Exposure Survey of Trends in Occupational Practice (EXPO‐S.T.O.P.) 2015: A national survey of sharps injuries and mucocutaneous blood exposures among health care workers in US hospitals. American Journal of Infection Control 2017;S0196‐6553(17):30770‐8. - PubMed
Hanrahan 1997
-
- Hanrahan A, Reutter L. A critical review of the literature on sharps injuries: epidemiology, management of exposures and prevention. Journal of Advanced Nursing 1997;25(1):144‐54. - PubMed
Harb 2015
Higgins 2011
-
- Higgins JPT, Green S, editors. Cochrane Handbook for Systematic Reviews of Interventions 5.1.0 [updated March 2011]. The Cochrane Collaboration, 2011. Available from www.cochrane‐handbook.org..
HSE 2012
-
- Beswick A, Robinson E, Evans G, Codling A. An evaluation of the efficacy of safer sharps devices. http://www.hse.gov.uk/research/rrpdf/rr914.pdf. HSE, (accessed May 20th, 2017).
Hutin 2003
Ilhan 2006
-
- Ilhan MN, Durukan E, Aras E, Turkcuoglu S, Aygun R. Long working hours increase the risk of sharp and needlestick injury in nurses: the need for new policy implication. Journal of Advanced Nursing 2006;56(5):563‐8. - PubMed
Jagger 1988
-
- Jagger J, Hunt EH, Brand‐Elnaggar J, Pearson RD. Rates of needle‐stick injury caused by various devices in a university hospital. The New England Journal of Medicine 1988;319(5):284‐8. - PubMed
Jagger 1990
-
- Jagger J, Hunt EH, Pearson RD. Estimated cost of needlestick injuries for six major needled devices. Infection Control and Hospital Epidemiology 1990;11(11):584‐8. - PubMed
Jagger 2013
-
- Jagger J, Perry J. Safety‐engineered devices in 2012: the critical role of healthcare workers in device selection. Infection Control and Hospital Epidemiology 2013;34(6):615‐8. - PubMed
Leigh 2007
-
- Leigh JP, Gillen M, Franks P, Sutherland S, Nguyen HH, Steenland K, et al. Costs of needlestick injuries and subsequent hepatitis and HIV infection. Current Medical Research and Opinion 2007;23(9):2093‐105. - PubMed
MacCannell 2010
-
- MacCannell T, Laramie AK, Gomaa A, Perz JF. Occupational exposure of health care personnel to hepatitis B and hepatitis C: prevention and surveillance strategies. Clinics in Liver Disease 2010;14(1):23‐36, vii. - PubMed
Mast 2004
-
- Mast E, Mahoney F, Kane M, Margolis H. Hepatitis B vaccines. In: Plotkin OW editor(s). Vaccines. Philadelphia: Saunders, 2004:299‐337.
Mischke 2014
Ngatu 2011
-
- Ngatu NR, Phillips EK, Wembonyama OS, Hirota R, Kaunge NJ, Mbutshu LH, et al. Practice of universal precautions and risk of occupational blood‐borne viral infection among Congolese health care workers. American Journal of Infection Control 2012;40(1):68‐70. - PubMed
Oh 2005
-
- Oh HS, Yi SE, Choe KW. Epidemiological characteristics of occupational blood exposures of healthcare workers in a university hospital in South Korea for 10 years. Journal of Hospital Infection 2005;60(3):269‐75. - PubMed
Orji 2002
-
- Orji EO, Fasubaa OB, Onwudiegwu U, Dare FO, Ogunniyi SO. Occupational health hazards among health care workers in an obstetrics and gynaecology unit of a Nigerian teaching hospital. Journal of Obstetrics and Gynaecology 2002;22(1):75‐8. - PubMed
OSHA 2001
-
- Occupational Safety and Health Administration (OSHA). Frequently asked questions needlestick safety and prevention act. https://www.osha.gov/needlesticks/needlefaq.html (accessed 8 August 2017).
Panlilio 2004
-
- Panlilio AL, Orelien JG, Srivastava PU, Jagger J, Cohn RD, Cardo DM, et al. Estimate of the annual number of percutaneous injuries among hospital‐based healthcare workers in the United States, 1997‐1998. Infection Control and Hospital Epidemiology 2004;25(7):556‐62. - PubMed
Parantainen 2011
Pruss‐Ustun 2005
-
- Pruss‐Ustun A, Rapiti E, Hutin Y. Estimation of the global burden of disease attributable to contaminated sharps injuries among healthcare workers. American Journal of Industrial Medicine 2005;48(6):482‐90. - PubMed
Ramsay 2001
-
- Ramsay C. Robust methods for analysis of interrupted time‐series designs for inclusion in systematic reviews. Proceedings of the ninth Annual Cochrane Colloquium; 2001; Lyon. 2001.
Ramsay 2003
-
- Ramsay CR, Matowe L, Grilli R, Grimshaw JM, Thomas RE. Interrupted time‐series designs in health technology assessment: lessons from two systematic reviews of behavior change strategies. International Journal of Technology Assessment in Health Care 2003;19(4):613‐23. - PubMed
Ratner 1994
-
- Ratner SL, Norman SA, Berlin JA. Percuteous injuries on the frontline: a survey of house staff and nurses. American Journal of Preventive Medicine 1994;10:372‐7. - PubMed
RevMan 2014 [Computer program]
-
- The Nordic Cochrane Centre, The Cochrane Collaboration. Review Manager (RevMan). Version 5.3 for Windows. Copenhagen: The Nordic Cochrane Centre, The Cochrane Collaboration, 2014.
Roberts 1999
-
- Roberts RB, Lincoff AS, Frankel RE. Occupational exposure to HIV in an urban university hospital setting. Brazilian Journal of Infectious Diseases 1999;3(2):50‐62. - PubMed
Robinson 2002
-
- Robinson KA, Dickersin K. Development of a highly sensitive search strategy for the retrieval of reports of controlled trials using PubMed. International Journal of Epidemiology 2002;31(1):150‐3. - PubMed
Rogers 2000
-
- Rogers B, Goodno L. Evaluation of interventions to prevent needlestick injuries in health care occupations. American Journal of Preventive Medicine 2000;18(4 Suppl):90‐8. - PubMed
Royle 2003
-
- Royle P, Milne R. Literature searching for randomized controlled trials used in Cochrane reviews: rapid versus exhaustive searches. International Journal of Technology Assessment in Health Care 2003;19(4):591‐603. - PubMed
Saarto 2011
Smith 2006
-
- Smith DR, Choe MA, Jeong JS, Jeon MY, Chae YR, An GJ. Epidemiology of needlestick and sharps injuries among professional Korean nurses. Journal of Professional Nursing 2006;22(6):359‐66. - PubMed
Smith 2006b
-
- Smith DR, Mihashi M, Adachi Y, Nakashima Y, Ishitake T. Epidemiology of needlestick and sharps injuries among nurses in a Japanese teaching hospital. Journal of Hospital Infection 2006;64(1):44‐9. - PubMed
Sohn 2006
-
- Sohn JW, Kim BG, Kim SH, Han C. Mental health of healthcare workers who experience needlestick and sharps injuries. Journal of Occupational Health 2006;48(6):474‐9. - PubMed
Tarigan 2015
-
- Tarigan LH, Cifuentes M, Quinn M, Kriebel D. Prevention of needle‐stick injuries in healthcare facilities: a meta‐analysis. Infection Control and Hospital Epidemiology 2015;36(7):823‐829. - PubMed
Trim 2004
-
- Trim JC. A review of needle‐protective devices to prevent sharps injuries. British Journal of Nursing (Mark Allen Publishing) 2004;13(3):144, 146‐53. [PUBMED: 14997076] - PubMed
Tuma 2006
-
- Tuma S, Sepkowitz KA. Efficacy of safety‐engineered device implementation in the prevention of percutaneous injuries: a review of published studies. Clinical Infectious Diseases 2006;42(8):1159‐70. - PubMed
UNAIDS 2009
-
- UNAIDS. AIDS epidemic update: November 2009. http://www.unaids.org/en/media/unaids/contentassets/dataimport/pub/repor.... UNAIDS, (accessed Dec 15th, 2011).
Verbeek 2005
Waclawski 2004
-
- Waclawski ER. Evaluation of potential reduction in blood and body fluid exposures by use of alternative instruments. Occupational Medicine 2004;54(8):567‐9. - PubMed
Wallis 2007
WHO 2016
-
- WHO. WHO guideline on the use of safety‐engineered syringes for intramuscular, intradermal and subcutaneous injections in health care settings. http://apps.who.int/iris/bitstream/10665/250144/1/9789241549820‐eng.pdf. WHO, (accessed May 10th, 2017). - PubMed
References to other published versions of this review
Lavoie 2012
-
- Lavoie, Marie‐Claude, Verbeek, Jos H, Parantainen, Annika, et al. Devices for preventing percutaneous exposure injuries caused by needles in health care personnel. Cochrane Database of Systematic Reviews 2012, Issue 4. [DOI: 10.1002/14651858.CD009740] - DOI
Publication types