Leadless pacemaker extraction from a single-center perspective

Affiliations

Leadless pacemaker extraction from a single-center perspective

Elkin González Villegas et al. Pacing Clin Electrophysiol. 2018 Feb.

Abstract

Background: Leadless pacemaker can be considered as a technical revolution in cardiac pacing devices, with clear advantages over conventional pacemakers in overcoming all lead-related complications. However, the management of these devices once they reach the end of life (EOL) of the battery is still controversial. In the next years, there will be an increase in the need to define a clear strategy in the management of leadless PM once they reach their EOL. Safe extraction of these devices will define in a great manner this strategy METHODS: We performed the extraction of three functioning Nanostim leadless pacemaker prophylactically in two females and one male patients as part of the Nanostim battery depletion field action recommendation. All patients had a prior transesophageal 3D echocardiography to determine the device intracardiac mobility and the extent of possible endothelialization. For the extractions, we used the Nanostim Retrieval Catheter S1RSIN (St. Jude Medical, St. Paul, MN, USA), which is a proprietary catheter provided by the manufacturing company based on a lasso.

Results: Complete extraction of the devices was achieved in all patients using a relatively short fluoroscopic time (16, 19, and 12 minutes).

Conclusions: The extraction of leadless pacemakers can be considered as a safe and feasible procedure using the tools provided by the manufacturer and designed for the extraction. However, a very low threshold must be maintained to avoid any risk to the patients. Our extraction time ranges are between 983 and 1,070 days, nevertheless it is necessary to gather more long-term data to assess the feasibility and safety of these procedures.

Keywords: extraction; leadless; pacemaker; pacing.