Effect of international sensitivity index (ISI) of thromboplastins on precision of international normalised ratios (INR)
- PMID: 2921349
- PMCID: PMC1141799
- DOI: 10.1136/jcp.42.1.92
Effect of international sensitivity index (ISI) of thromboplastins on precision of international normalised ratios (INR)
Abstract
The contribution of the thromboplastin international sensitivity index (ISI) to the interlaboratory coefficient of variation (CV) of the international normalised ratio (INR) with individual reagents was assessed. In theory the precision of the INR should increase with lower ISI values. An empirical relation has been established between the ISI, the INR, and its CV for two rabbit thromboplastins used in sufficient numbers for analysis in the United Kingdom. This was based on the cumulative data from the United Kingdom National External Quality Assessment Scheme (NEQAS) surveys over two years beginning in 1986. The actual precision achieved in NEQAS for the two reagents depends on the ISI value of the thromboplastin and it agreed closely with the figure predicted by the empirical model. The findings show that the ISI value of a thromboplastin strongly influences the interlaboratory variability of the INR obtained with it. The CV of the INR approximates to the CV of the prothrombin ratio multiplied by the ISI. Manufacturers of thromboplastin should therefore be encouraged to produce reagents showing good precision of results reported as simple prothrombin ratios and a low ISI value to avoid impairment in precision when ratio results are transformed to INR.
Comment in
-
ISI value of thromboplastin.J Clin Pathol. 1989 Oct;42(10):1118-9. doi: 10.1136/jcp.42.10.1118-b. J Clin Pathol. 1989. PMID: 2584421 Free PMC article. No abstract available.
Similar articles
-
Effect of thromboplastin and coagulometer interaction on the precision of the International Normalised Ratio.J Clin Pathol. 1995 Jan;48(1):13-7. doi: 10.1136/jcp.48.1.13. J Clin Pathol. 1995. PMID: 7706513 Free PMC article.
-
Quality control of the prothrombin time and international normalized ratios. National and international schemes.Ric Clin Lab. 1990 Jan-Mar;20(1):59-69. doi: 10.1007/BF02910149. Ric Clin Lab. 1990. PMID: 2356406
-
The prothrombin time/international normalized ratio (PT/INR) Line: derivation of local INR with commercial thromboplastins and coagulometers--two independent studies.J Thromb Haemost. 2011 Jan;9(1):140-8. doi: 10.1111/j.1538-7836.2010.04109.x. J Thromb Haemost. 2011. PMID: 20942851
-
[Prothrombin time and its standardization].Rinsho Byori. 2002 Aug;50(8):779-85. Rinsho Byori. 2002. PMID: 12373814 Review. Japanese.
-
International standardization of laboratory control of oral anticoagulant therapy: a survey of thromboplastin reagents used for prothrombin time testing.J Heart Valve Dis. 1993 Jan;2(1):42-52. J Heart Valve Dis. 1993. PMID: 8269109 Review.
Cited by
-
Accuracy, precision, and quality control for point-of-care testing of oral anticoagulation.J Thromb Thrombolysis. 2001 Sep;12(1):35-40. doi: 10.1023/a:1012734426811. J Thromb Thrombolysis. 2001. PMID: 11711687 Review.
-
United Kingdom external quality assessment in blood coagulation: the first 21 years.J Clin Pathol. 1989 Jan;42(1):1-3. doi: 10.1136/jcp.42.1.1. J Clin Pathol. 1989. PMID: 2921336 Free PMC article. No abstract available.
-
Effect of thromboplastin and coagulometer interaction on the precision of the International Normalised Ratio.J Clin Pathol. 1995 Jan;48(1):13-7. doi: 10.1136/jcp.48.1.13. J Clin Pathol. 1995. PMID: 7706513 Free PMC article.
-
Investigation of the Molecular Mechanism of Coagulopathy in Severe and Critical Patients With COVID-19.Front Immunol. 2021 Dec 16;12:762782. doi: 10.3389/fimmu.2021.762782. eCollection 2021. Front Immunol. 2021. PMID: 34975853 Free PMC article.
-
Evaluation of the Ciba Corning Biotrack 512 coagulation monitor for the control of oral anticoagulation.J Clin Pathol. 1991 Nov;44(11):950-3. doi: 10.1136/jcp.44.11.950. J Clin Pathol. 1991. PMID: 1752987 Free PMC article.
References
MeSH terms
Substances
LinkOut - more resources
Full Text Sources