New Jersey low contact stress knee replacement system. Ten-year evaluation of meniscal bearings
- PMID: 2922189
New Jersey low contact stress knee replacement system. Ten-year evaluation of meniscal bearings
Erratum in
- Orthop Clin North Am 1989 Oct;20(4):preceding 519
Abstract
A metal-backed, interchangeable, meniscal bearing knee replacement system has been designed to combine low constraint forces with low contact stresses (LCS) to allow nearly normal joint articulation and loading as well as long-term wear resistance of the implants. System components allow unicompartmental, bicompartmental, tricompartmental, and revision knee arthroplasties. In this prospective comparison study, the early series of cases used methyl methacrylate for prosthesis-to-bone fixation, whereas the more recent series of knee replacements used sintered-bead porous coating to allow tissue ingrowth stabilization of all implants. Overall results in the first 149 cemented cases with 2- to 10-year follow-up (mean 7.6 years) were good to excellent in 85.2 per cent, fair in 3.4 per cent, and poor in 11.4 per cent using a strict knee scoring scale. Fair and poor results were seen predominantly in multiply operated and implant revision cases. The best cemented results were noted in primary cases, 95.1 per cent of which had good to excellent results. Overall results in the first 208 cementless cases with 2- to 7-year follow-up (mean 4.4 years) were good to excellent in 91.8 per cent, fair in 2.4 per cent, and poor in 5.8 per cent. Fair and poor results were also seen predominantly in multiply operated and implant revision cases. The best cementless results were noted in primary cases, 98.2 per cent of which had good to excellent results. Of 140 meniscal bearing implants in the entire series, one dislocation occurred (0.7 per cent) which required open bearing replacement. Of 217 rotating platform bearing implants in the entire series, 7 dislocations occurred (3.2 per cent), which required open bearing replacement or revision of components for malposition. These dislocations were seen predominantly in revision cases involving insufficient flexion stability; 6 of the 7 cases were revisions. This study demonstrates the efficacy of mobile bearing elements for use in knee replacement arthroplasty. It is essential that flexion and extension gaps be controlled to maintain contact pressure on such bearings to avoid problems of subluxation or dislocation.
Similar articles
-
A metal-backed, rotating-bearing patellar prosthesis to lower contact stress. An 11-year clinical study.Clin Orthop Relat Res. 1989 Nov;(248):34-49. Clin Orthop Relat Res. 1989. PMID: 2805494
-
Twenty-year evaluation of the New Jersey LCS Rotating Platform Knee Replacement.J Knee Surg. 2002 Spring;15(2):84-9. J Knee Surg. 2002. PMID: 12013078
-
[Uncemented knee prosthesis. Results apropos of 58 cases with a minimum of 5-year follow-up].Rev Chir Orthop Reparatrice Appar Mot. 1995;81(6):505-13. Rev Chir Orthop Reparatrice Appar Mot. 1995. PMID: 8560021 French.
-
Meniscal-bearing unicompartmental knee arthroplasty. An 11-year clinical study.Orthop Rev. 1991 May;20(5):443-8. Orthop Rev. 1991. PMID: 2067879 Review.
-
Mobile-bearing knee arthroplasty: rotation is our salvation!J Arthroplasty. 2004 Jun;19(4 Suppl 1):27-30. doi: 10.1016/j.arth.2004.02.011. J Arthroplasty. 2004. PMID: 15190545 Review.
Cited by
-
Fixed or mobile-bearing total knee arthroplasty.J Orthop Surg Res. 2007 Jan 5;2:1. doi: 10.1186/1749-799X-2-1. J Orthop Surg Res. 2007. PMID: 17204165 Free PMC article.
-
In vivo kinematics comparison of fixed- and mobile-bearing total knee arthroplasty during deep knee bending motion.Knee Surg Sports Traumatol Arthrosc. 2014 Jul;22(7):1612-8. doi: 10.1007/s00167-012-2333-7. Epub 2012 Dec 12. Knee Surg Sports Traumatol Arthrosc. 2014. PMID: 23232786
-
Effect of sagittal femoral component alignment on biomechanics after mobile-bearing total knee arthroplasty.J Orthop Surg Res. 2019 Nov 28;14(1):400. doi: 10.1186/s13018-019-1458-5. J Orthop Surg Res. 2019. Retraction in: J Orthop Surg Res. 2021 Mar 10;16(1):181. doi: 10.1186/s13018-021-02314-8. PMID: 31779650 Free PMC article. Retracted.
-
[From tibiofemoral instability to dislocation in total knee arthroplasty].Orthopade. 2007 Oct;36(10):917-22, 924-7. doi: 10.1007/s00132-007-1142-z. Orthopade. 2007. PMID: 17876569 Review. German.
-
In vivo kinematics of mobile-bearing total knee arthroplasty during deep knee bending under weight-bearing conditions.Knee Surg Sports Traumatol Arthrosc. 2011 Jun;19(6):914-20. doi: 10.1007/s00167-010-1262-6. Epub 2010 Sep 16. Knee Surg Sports Traumatol Arthrosc. 2011. PMID: 20845031
Publication types
MeSH terms
Substances
LinkOut - more resources
Other Literature Sources